Automated Hematology Dr.P.Dahim Health Reference Laboratory • • • • Principles of Automation Calibration Quality Control Flagging Hematology Automation • Two General Principles Electronic resistance ( impedance) Light scattering
Download ReportTranscript Automated Hematology Dr.P.Dahim Health Reference Laboratory • • • • Principles of Automation Calibration Quality Control Flagging Hematology Automation • Two General Principles Electronic resistance ( impedance) Light scattering
Automated Hematology Dr.P.Dahim Health Reference Laboratory 1 • • • • Principles of Automation Calibration Quality Control Flagging 2 Hematology Automation • Two General Principles Electronic resistance ( impedance) Light scattering 3 Electronic Resistance (Impedance) • Utilizes non-conductive properties of blood cells as blood cell passes through orifice of aperture it displaces its own volume increased resistance between electrodes results in an electrical pulse RBCs and Platelets counted together, separated by pulse heights hydrodynamic focusing forces cells to pass single file through sensing zone 4 Cell counting 5 Light scattering • Cells counted as passed through focused beam of light( LASER) • Sum of diffraction(bending around corners), refraction (bending due to change in speed) and reflection (light rays turned back by obstruction). • Multi angle polarized scatter separation (M.A.P.S.S) 0° : indicator of cell size 10° : indicator of cell structure and complexity 90° polarized: indicates nuclear lobularity 90° depolarized: differentiate eosinophils 6 Principles of counting &sizing • Counting and sizing red cells ,WBCs , Plts by • • • • • counting the number of pulses and height of pulses Measuring Hb by modified HiCN method Measuring MCV by computing the mean of height of pulses Hct (PCV ) derived from MCV & RBC MCH derived from Hb & RBC MCHC derived from Hb , RBC & MCV 7 • RDW : SD or CV of individual measurement of red cell volume • HDW :The degree of variation in red cell hemoglobinization,CV of measurements of hemoglobin concentration of individual cells • NRBC : included in the WBC counting 8 WBC Differential Count : • Impedance technology with current of various frequencies • Light scattering & light absorbance 9 Reticulocyte counting : • Flourescence – based methods 10 Histograms • RBC, PLT, and WBC plotted on histogram • X-Axis Cell size in femtoliters (fL) • Y-Axis # of cells 11 Histogram WBC: Distribution with three individual peaks and valleys at specific regions representing the lymphocytes, monocytes, and granulocytes. All curves normally start and end at baseline 12 مشخصات شناسنامه دستگاههای شمارشگر سلولی خودکار • • • • • • -1تاريخ خريد -2تاريخ نصب ،کاليبراسيون و نام فرد راه اندازی کننده -3تاريخ شروع بکار -3کارخانه و کشور سازنده -4مدل وشماره سلاير دستگاه -5شرايط دستگاه هنگام خريد (نو ،مستعمل ،بازسازی شده ) 13 • • • • • 14 -6نام وآدرس شرکت پشتيبان -7نام کاربر يا کاربران دستگاه -8مشخصات و نحوه آموزش کاربر دستگاه توسط شرکت پشتيبان -9ميزان عدم دقت و عدم دقت دستگاه در هنگام نصب ساير مستندات مشخصات محلولها و مواد مصرفی تاريخ آخرين سرويس دوره ای و خدمات پشتيبانی تاريخ آخرين کاليبراسيون تاريخ آخرين تعمير با ذکر قطعه تعويض شده -تاريخ آخرين مشکل و راه حل آن 15 اصول کار كار با دستگاه سل كانتر مي بايست بطور كامل مطابقكاتالوگ دستگاه يا آموزش كارشناسان شركت پشتيبان صورت گيرد . 16 اصول کار درصورت تعويض کاربر دستگاه ،می بايست از آموزشوی در مورد چگونگی کار با دستگاه و نحوه نگهداری و اصول کنترل کيفی اطمينان حاصل نمود . 17 اصول کار کليه موارد مربوط به نگهداری دستگاه ،از قبيل تاريخانجام شستشوهای الزم ،تعمير ،سرويس ويا تعويض محلولها می بايست ثبت ونگهداری شوند. 18 اصول کار هر روز شمارش زمينه يا Back groundدستگاه ارزيابي و در صورت امکان ثبت و نگهداری شود .بطور معمول شمارش زمينه هر دستگاه در کاتالوگ مربوطه آمده است ولی مقادير قابل قبول سازمان جهانی بهداشت به شرح زير می باشند : • RBC: <0/03 x 1012/l • WBC: <0/04 x 109 / l • Haemoglobin: <0/2g/dl • Platelets: <5 x 109 /l 19 Calibration • Calibration provides the most accurate results possible. • For best performance, calibrate all the CBC parameters. • The WBC differential is calibrated at the factory. They do not require calibration in the laboratory. 20 When to Calibrate • At installation • After the replacement of any component that involves dilution characteristics or the primary measurements (such as the apertures) • When advised to do so by your service representative 21 Calibration Normal fresh blood samples Calibrators 22 Accuracy • • 23 ×100ميانگين روش دستگاهي -ميانگين روش دستي ميانگين روش دستگاهی = CF پايداری کاليبراسيون • /n / n-1 (∑d)2 )Sd =√ ∑(d2_ • tn = d √n • sd 24 Precision check • Purpose: check reproducibility of instrument • Precision quick method Run 5 replicates of one patient’s sample free of elevated bilirubin, lipemia, and hemoloysis • Calculate % CV or SD • Refer to the methodology in procedure manual for expected % CV or SD 25 Quality Control • Purpose of QC Assure proper functionality of instrumentation Monitoring the Integrity of the Calibration 26 Quality Control Methods • Assayed stabilized material (Commercial) • Previously analyzed patient samples Easily obtained Cost effective Results and samples readily available 27 Quality control • • • • • • • Replicate tests (control samples) Control charts Duplicate & Check tests Correlation system Delta check Daily means of MCV,MCH,MCHC Comparison the results with reference methods 28 Replicate tests • بررسي عدم دقت دستگاه 29 Control charts • استفاده از خون كنترل • تفسير با استفاده از قوانين سازمان جهاني بهداشت يا وستگارد 30 Quality Control • OUT OF CONTROL!!! Repeat the assay ( One time occurrence ) Check integrity of material Troubleshoot Verify instrumentation 31 Duplicate test درهرسري کاري 3-2نمونه به صورت مضاعف((Duplicate آزمايش ميشوند تا با بررسي اختالف خوانده ها از طريق محاسبات آماري از وجود خطاهاي تصادفي آگاه گرديد .افزايش اختالف بين دو خوانده بيش از 2SDمحاسبه شده ،احتمال وجود خطاي تصادفي را مطرح مي نمايد. • SD= √∑d2 • 2n 32 Check test • در ابتدای سری کاری يا صبح2 -3نمونه آزمايش شده وهمان نمونه ها در انتهاي سري كاري يا بعد از ظهرمجددا آزمايش مي شوند. نتايج حاصل از اين دو آزمايش با استفاده از فرمول Duplicate testقابل مقايسه می باشند .اختالف نتايج در محدوده 2SDقابل قبول می باشد .در صورت نگهداری نمونه ها در شرايط مناسب ، هرگونه تغيير در نتايج خارج از اين محدوده نشاندهنده اشکال درعملکرد دستگاه و يا معرفها مي باشد. • اين آزمايش براي بررسي هموگلوبين و گلبولهای قرمز مناسب بوده و به ميزان کمتر براي شمارش گلبولهاي قرمز و پالکت کاربرد دارد ولي براي هماتوکريت بخصوص اگر فاصله زماني آزمايش نمونه ها بيش از 6ساعت باشد کارايي ندارد. 33 Delta check • • • • • • Hb 2 g/l PCV 0.05 L/L MCV >6 fL MCH > 5 pg WBC Normal to abnormal Platelets Reduced or increased by more than 50% 34 Daily means of MCV,MCH,MCHC • مراکز آزمايشگاهي با پذيرش بيمار زياد (حداقل روزی 100بيمار • تفاوت مقادير بيش از ٪3از ميانگين ،نشان دهنده بروز خطا و لزوم بررسی کاليبراسيون دستگاه می باشد. 35 Comparison the results with reference methods • انجام محاسبات آماري براي بررسي هم خواني نتايج الزامي است . 36 ERRORS INACCURACY 37 Reliability of Electronic Counters • High Precision • Accuracy? - Recirculation of cells - Red cell agglutination - Lipid droplets, Microorganisms, - Extraneous particles - Faulty maintenance - Incorrectly calibration 38 Interfering substances • • • • • • • Cold Agglutinins Cryoglobulins Lipemia Platelet Clumps RBC fragments NRBCs Clot 39 WBC Count Interferences Unusual RBC abnormalities that resist lysing malarial parasites giant platelets, platelet clumps, NRBCs fragmented white cells, agglutinated white cells, lyse-resistant red cells, cryoglobulin, some extremely elevate proteins, unlysed particles greater than 35 fL in size 40 RBC Count Interferences • Very high WBC count, • high concentration of very large platelets, • Auto-agglutination 41 Hgb • Very high WBC count, • Severe lipemia, • certain unusual RBC abnormalities that resist lysing 42 MCV • Very high WBC count, • high concentration of very large platelets, • Auto-agglutination 43 Flagging • WBC Suspect flags Blasts Imm Grans Variant lymphs 44 More Flagging • RBC Suspect flags NRBCs Macrocytic RBCs Dimorphic RBC population Micro RBCs/RBC fragments RBC agglutination Check S lide 45 Troubleshooting Flagged Results: • Refer to the Instrument Manual to troubleshoot codes, flags, and messages displayed with patient results 46 47