R & D of LAr TPC @ KEK
Download
Report
Transcript R & D of LAr TPC @ KEK
Development of Liquid Ar TPC @ KEK
Takuya Hasegawa, Takashi Kobayashi,
Takasumi Maruyama, Koichiro Nishikawa,
Masashi Tanaka
2/12
Next Generation n and Nucleon Decay Physics
CP violation in nm→ne oscillation.
beyond q13
Larger far detector will be needed
p0 reduction will become more important
Proton decay: p→nK+
SUSY models predict large branching fraction
Charged Kaon identification is important
Liquid Argon TPC is an excellent detector for these
particular physics modes
100 kt LAr detector has comparable sensitivity to
500 kt water cherenkov detector
3/12
Liquid Argon TPC
Ionization
Electron
E
Ionization Electron
105 e / MeV
Slow drift: m/ms@ kV/cm
Small diffusion: 3 mm after
20 m drift
Cherenkov Light
Drift electron lifetime:
O2,H2O contamination
No amplification in liquid
νμ Charged current
Charged
Particles
Scintillation Light
νe Charged current
A. Bueno, et.al.,, hep-ph/0701101
4/12
Large Size Detector
REadout
GEM
GAS Argon
5kV/cm
GAr
20 m
Liquid Argon
1 kV/cm
E-field
E ≈ 3 kV/cm
LAr
E≈ 1 kV/cm
R&D items:
Recover attenuation
after long drift
→double phase readout
λ=0.5m/ppb
80 m
Ionization electron readout device/electronics
Large vessel、Argon purification、cryostat、HV
Analysis technique etc...
5/12
KEK R&D Status
Last summer, KEK neutrino group has started R&D
effort of the liquid Argon TPC
Still under establishing basic techniques
Helps from KEK
cryogenic/accelerator division
Detector Technology Project (測定器開発室)
Liquid Xenon R&D group
Collaborating with LAr R&D project at ETHZ
Built small test stand to study;
Readout device and electronics
Liquid Argon purity
Short term goal
Establish double phase TPC
6/12
KEK Test Stand
NIM A241 (1985), 62
Refer ’85 C. Rubbia’s paper
20cmfx25cm test chamber
inside an open bath of LAr
Liquefy Gas Argon inside
test chamber
Purify Gas Argon
Commercially available filter
Oxygen (cupper reductant)
Water (molecular sieve)
Specification: <10 ppb
Hydrosorb
Oxysorb
Vacuum
pump
Open bath
Liquid Ar
Gas Ar
Test
Chamber
7/12
Test Chamber
Open bath
30cmfx45cm
Test chamber: CF253 flange
vacuum vessel
Port: signal/HV/Gas input/Vacuum
super insulator
8/12
Ion chamber
Simple ion chamber
10cmx10cm Cu plates
5 cm distance
-2 kV HV (cathode)
Trigger counter
select events pass through
the detector
Oscilloscope readout
no amplification
1 MW coupling
readout t ~ 400 ms (>>
drift time ~50 ms)
Observed signal
Blue: Trigger signal
Yellow: Anode signal
Took data for ~20 hors
~650 trigger events
9/12
Signal Charge
Integrate the oscilloscope waveform
to estimate signal charge
Integration cutoff 200—600 ms
40% pedestal event (trigger
coverage is larger than the
detector)
Divide data into 4 time periods
No major degradation
t ~ 50 – 100 hours
200 ms
400 ms
600 ms
w/o pedestal
with pedestal
10/12
Drift Electron Attenuation
Signal charge is sensitive to
the drift electron attenuation
Simple toy MC to estimate the
signal charge for different
attenuation length, then
compare with the data
Signal charge distribution
Cosmic data (red)
MC prediction (black)
Data and MC shows the best
agreement with no attenuation
This analysis is an estimation
of absolute signal yield
large systematic
uncertainty (~50%)
>5 cm (10 ppb) after
considering the systematics
5 cm
11/12
R&D Strategy at KEK Test Stand
purification and
liquefaction
Purity
monitor
Observe Cosmic muon
(simple ion chamber)
2D wire chamber
Establish 1 phase
TPC
Establish double
phase TPC
TPC readout
electronics
Extraction Grid
Strip/Pad
readout
GEM basic
properties
completed
ongoing
major milestone
12/12
Summary
As a candidate of next generation neutrino/nucleon
decay detector, KEK neutrino group is interested in
Liquid Argon TPC.
We have built a 10L test chamber
Successfully observed cosmic muons
Took data for ~20 hours without major
degradation of the signal
Comparison of the signal charge between data
and MC prediction shows we have already
obtained good liquid Argon purity (<10 ppb)
Short term plan
Establish double phase TPC with 10L chamber
13/12
Backup
14/12
液化能力テスト
まずは、安価な窒素を用いて液化能力
の検証を行った。
テスト容器を液体窒素で満たした浴槽
に固定する
気体窒素をテスト容器内に送り込む
容器内の圧力は1.4気圧程度
容器内に挿入した液面計で液化速度を
測定。
使用した気体量から液体量を推測する
ことも可能。
1時間当たりに
4 cm、1.3L の液体を生成
20 kgf/cm2、1m3の気体を使用
当面の実験には十分な液化速度が得ら
れた
液化速度を改善する方法
液化時の容器内圧力を上げる
浴槽中の液体とテスト容器の熱交
換の効率を上げる
15/12
油回転ポンプ+ターボ分子ポンプ
油回転ポンプはオイルの逆流が
問題となるので、ドライスクロール
ポンプを購入予定
ターボ分子ポンプをできるかぎり
容器に近づけることにより、達成
真空度は飛躍的に向上する。
リボンヒーターにより容器を100℃まで
ベーキング
到達真空度(20℃): 6x10-4 Pa
アウトガス速度: 0.2 Pa/hour
テスト容器を液体アルゴンに完全に浸
した場合、真空が大きく破れる
ガスケット(銅)とフランジ(SUS)の
熱膨張率の違いから生じた隙間
に、液体アルゴンが入り込み容器
内で気化していると思われる。
液体アルゴンの液面をフランジよ
りも下に保って実験を行うことに
する
真空度(Pa)
真空テスト
液面がフランジ
に到達する
時間
16/12
平行板検出器
10cmx10cmの二枚の銅板を平行に
間隔5cmで並べた単純な検出器
下側銅板(カソード)に高電圧(-2kV)
上側銅板(アノード)より信号読み出し
期待される信号
dに生じた電荷Qが速度vで銅板間を
動く場合
d
D
電荷Q
速度v
Qd/D
d/v
-HV
宇宙線が検出器を突き抜けた場合
Q=5(cm)x3(MeV/cm)/24(e
V/電子)=6.3x105電子相当
読み出される信号電荷量: Q/2
Q/2
D
-HV
Qv/D
Qv/D
D/v
ドリフト電子の減衰がある場合、信号
電荷は少なくなる
17/12
読み出しエレクトロニクス
増幅/ADC/バッファリング/トリ
ガー/データ書き出し 一連の読
み出し
実際に宇宙線の飛跡を再構成す
る
CAENの液体アルゴンに特化し
たシステムの使用を検討してい
る (もしくはAMPTEK)
CAEN SY2791液体アルゴン用
読み出しシステム(256 ch)
二相読み出しの研究
アルゴンの液相から気相にドリ
フト電子をとりだす。
どのように電場を形成するか
取り出しグリッドの最適化
液面の高さの制御
hep-ph/0811.3384
18/12
循環・純化装置
GEM
ゲッタ
ポンプ
カーボン製のGEM(林栄製)
→ 放電しにくい特徴がある。
→ 高圧印加後の時間安定性なども課
題。
まずは必要かどうか判断
左図 (循環純化装置の例)
→ ゲッターと呼ばれるフィルタ
→ 気体循環ポンプ
右図
→ 浴槽を使うスキームを使う場合、冷
却能力を上げるための工夫