HbA1c Euro 2014

Download Report

Transcript HbA1c Euro 2014

Kwaliteit van Medische Laboratoria
GLOBAAL RAPPORT EKE HbA1c - EURO 2014.12
Target HbA1c = 62.8 mmol/mol (Referentiemethode NGSP/DCCT)
04 Augustus 2014
INSTRUMENT
RESULT
(Mean)
N
SD
CV%
Bias
Bias%
Capillarys 2 flex – piercing IFCC mmol/mol
63.0
7
2.5
4.0
0.2
0.3
D-10_A1c_Program IFCC mmol/mol
66.2
9
3.1
4.7
3.4
5.4
Menarini HA-8140 DM IFCC mmol/mol
66.0
1
****
****
3.2
5.1
Menarini HA-8160 TP IFCC mmol/mol
64.6
13
2.1
3.2
1.8
2.8
Menarini HA-8160 VP IFCC mmol/mol
64.5
45
1.7
2.7
1.7
2.7
Menarini HA-8180V IFCC mmol/mol
63.7
9
2.4
3.8
0.8
1.3
Minicap flex-piercing mmol/mol
61.7
6
1.2
2.0
-1.2
-1.8
62.0
2
1.4
2.3
-0.8
-1.3
64.1
5
1.4
2.2
1.3
2.0
TOSOH G7 IFCC mmol/mol
64.0
4
3.6
5.6
1.2
1.9
TOSOH G8 IFCC mmol/mol
64.4
37
1.6
2.4
1.6
2.5
Vantage (DCA 2000) IFCC mmol/mol
67.0
1
****
****
4.2
6.7
Variant II_HbA1c IFCC mmol/mol
62.5
2
3.5
5.7
-0.3
-0.5
Variant-II_Turbo_A1c_Program IFCC
mmol/mol
64.3
4
0.5
0.8
1.4
2.3
Tina-quant, Whole blood, Gen.2, IFCC
mmol/mol
Tina-quant, Whole blood, Gen.3, IFCC
mmol/mol
Uw individueel rapport kunt u opvragen via de site https://biocli.wiv-isp.be/toolkit/login/login.cfm
De enquêtecoördinatoren
Master Y. Lenga
Dr. K. Vernelen
Rue Juliette Wytsmanstraat 14
1050 Brussels | Belgium
T + 32 2 642 51 11 | F + 32 2 642 50 01
[email protected] | www.wiv-isp.be