IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 16.1: Optimization of protection in fluoroscopy Practical.

Download Report

Transcript IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 16.1: Optimization of protection in fluoroscopy Practical.

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
RADIATION PROTECTION IN
DIAGNOSTIC AND
INTERVENTIONAL RADIOLOGY
Part 16.1: Optimization of protection in
fluoroscopy
Practical exercise
IAEA
International Atomic Energy Agency
Overview
• To become familiar with quality control tests
in fluoroscopy.
• To measure the standard entrance dose rate
to the patient
• To assess the patient thickness variation
effect on scattered radiation
IAEA
16.1: Optimization of protection in fluoroscopy
2
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 16.1: Optimization of protection in
fluoroscopy
Measurement of standard entrance dose rate
IAEA
International Atomic Energy Agency
Fluoroscopy - Standard Dose Rate
Purpose :
• Measurement of dose rate at the entrance of
patient for different thickness
• Effect on scattered radiation
Method :
• Use different water equivalent absorber (acrylic, 20 cm
for a standard patient) or copper sheets (2 mm for a
standard patient
• Place dosimeter on input (x-ray tube side) of absorber
IAEA
16.1: Optimization of protection in fluoroscopy
4
Set-up for
measurement of
standard dose
rate.
IAEA
16.2: Optimization of protection in fluoroscopy
5
The ionization chamber should be protected
pressure and possible malfunctioning. It should
be in contact with the acrylic to include
backscatter in the measurement.
IAEA
16.2: Optimization of protection in fluoroscopy
6
Use 10 cm thickness of acrylic to simulate a thin
patient. The table to intensifier distance is 35
cm (this distance will be kept constant for the
different patient thicknesses)
IAEA
16.1: Optimization of protection in fluoroscopy
7
The chamber is easily centred using the
fluoroscopic image
IAEA
16.1: Optimization of protection in fluoroscopy
8
The entrance
dose rate is
1.78 mGy/min.
IAEA
16.1: Optimization of protection in fluoroscopy
9
The phantom
thickness is
now 20 cm.
IAEA
16.1: Optimization of protection in fluoroscopy
10
The chamber now reads 8.85 mGy/min.
IAEA
16.1: Optimization of protection in fluoroscopy
11
If the distance
between the
input screen of
the intensifier
and the entrance
of the patient is
increased, i.e.,
extra 20 cm.
The patient
entrance dose
rate will
increase.
IAEA
16.1: Optimization of protection in fluoroscopy
12
Note: the chamber looks magnified (the
intensifier is further away from the patient).
IAEA
16.1: Optimization of protection in fluoroscopy
13
Before (intensifier-table
distance: 35 cm)
Now (intensifier-table
distance: 55 cm)
IAEA
16.1: Optimization of protection in fluoroscopy
14
The entrance dose
rate with the
intensifier at 55 cm
from the table is
17.9 mGy/min (to
be compared with
the previous value
of 8.85 mGy/min).
IAEA
16.1: Optimization of protection in fluoroscopy
15
Now the
phantom
thickness is
increased up to
30 cm of acrylic.
IAEA
16.1: Optimization of protection in fluoroscopy
16
The patient
entrance dose
rate at the
surface of the
phantom
increases to
24.8 mGy/min
(the reading was
8.85 with 20 cm
thickness).
IAEA
16.1: Optimization of protection in fluoroscopy
17
The dose rate due to scatter radiation also
increases with the patient thickness. For 30 cm
acrylic, 3 mGy/h is measured close to the
phantom.
IAEA
16.1: Optimization of protection in fluoroscopy
18
With the high contrast mode the dose rate for
scatter radiation (30 cm acrylic), increases to
7 mGy/h. The patient entrance dose rate is
now 59.6 mGy/min.
IAEA
16.1: Optimization of protection in fluoroscopy
19
With 10 cm
acrylic the
dose rate due
to scattered
radiation is
0.2 mGy/h.
IAEA
16.1: Optimization of protection in fluoroscopy
20
With 20 cm
acrylic the
dose rate due
to scattered
radiation is
1 mGy/h.
IAEA
16.1: Optimization of protection in fluoroscopy
21
With 30 cm
acrylic the
dose rate due
to scattered
radiation is
3 mGy/h.
IAEA
16.1: Optimization of protection in fluoroscopy
22
Note that with the extra thickness the image
quality is extremely poor (the border of the
chamber is hardly visible)
IAEA
16.1: Optimization of protection in fluoroscopy
23
Fluoroscopy - Standard Dose Rate
Analysis :
• should be < 25 mGy/min
Frequency :
• acceptance, tube change
• generator repair
• intensifier repair
• 6 monthly
IAEA
16.1: Optimization of protection in fluoroscopy
24
Where to Get More Information
Quality Control in Diagnostic Imaging, Gray JE,
Winkler NT, Stears J, Frank ED. Available at no
cost. http://www.diquad.com/QC%20Book.html
IAEA
15.3: Optimization of protection in radiography
25