IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 16.2: Optimization of protection in fluoroscopy Practical.
Download ReportTranscript IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 16.2: Optimization of protection in fluoroscopy Practical.
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 16.2: Optimization of protection in fluoroscopy Practical exercise IAEA International Atomic Energy Agency Overview / Objectives • To become familiar with quality control tests in fluoroscopy • To measure the maximum dose rate to the patient IAEA 16.2: Optimization of protection in fluoroscopy 2 IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 16.2: Optimization of protection in fluoroscopy Maximum dose rate IAEA International Atomic Energy Agency Fluoroscopy - Maximum Dose Rate Purpose : • to know the maximum dose rate to the patient Method : • place ~2mm lead over the image intensifier • place a dosimeter ~ 30 cm from the intensifier • operate unit at maximum kVp and mA • measure dose rate • X ray tube below the patient table IAEA 16.2: Optimization of protection in fluoroscopy 4 Set-up for measurement of maximum dose rate. IAEA 16.2: Optimization of protection in fluoroscopy 5 The ionization chamber should be protected to avoid pressure and possible malfunctioning. It should be in contact with the acrylic to measure also backscattering. IAEA 16.2: Optimization of protection in fluoroscopy 6 The intensifier should be as far as possible from the table with the x-ray tube as close as possible to the patient position. These are the worst conditions in terms of maximum radiation dose to the skin patient IAEA 16.2: Optimization of protection in fluoroscopy 7 Note: the X-ray tube is placed very close to the table. IAEA 16.2: Optimization of protection in fluoroscopy 8 A radio-opaque marker (a key) can be used to center the chamber in the image field IAEA 16.2: Optimization of protection in fluoroscopy 9 The key shows the motion of the C-arm or the table. IAEA 16.2: Optimization of protection in fluoroscopy 10 The key should be removed after having centred the chamber. IAEA 16.2: Optimization of protection in fluoroscopy 11 A folded lead apron is used to measure the maximum dose rate (use enough layers to approximate 2 mm of lead). The lead apron is placed over the protected chamber. IAEA 16.2: Optimization of protection in fluoroscopy 12 Note: This geometry is the worst in terms of patient radiation dose— the tube close to the table and the intensifier far from the patient. It also increases the scattered radiation dose to the physician and staff. IAEA 16.2: Optimization of protection in fluoroscopy 13 The dose rate is 79.5 mGy/min for this situation IAEA 16.2: Optimization of protection in fluoroscopy 14 The image quality is poor. The chamber edge is hardly visible. IAEA 16.2: Optimization of protection in fluoroscopy 15 The generator selects the maximum possible kV (110 kV for this equipment) and 3.6 mA. IAEA 16.2: Optimization of protection in fluoroscopy 16 In the high contrast mode with the same geometry, the maximum of 110 kV and 7.34 mA is selected by the generator. IAEA 16.2: Optimization of protection in fluoroscopy 17 The image quality is still poor although slightly improved. IAEA 16.2: Optimization of protection in fluoroscopy 18 In the high contrast mode the maximum dose rate is 186.7 mGy/min. IAEA 16.2: Optimization of protection in fluoroscopy 19 Fluoroscopy - Maximum Dose Rate Analysis : • FDA limits: • normal-dose fluoroscopy: maximum air-kerma rate 88 mGy/min (30 cm from the image receptor) • High-dose-rate fluoroscopy: maximum air-kerma rate 176 mGy/min (30 cm from image receptor) Frequency : • acceptance, 6-monthly • tube change, generator repair IAEA 16.2: Optimization of protection in fluoroscopy 20 Importance of quality assurance (QA) • It is very important to have a QA programmes established in your practice and not to base dose controls on specification provided by the manufacturer or on the additional checks performed by the manufacturer. IAEA 16.2: Optimization of protection in fluoroscopy 21 Where to Get More Information Quality Control in Diagnostic Imaging, Gray JE, Winkler NT, Stears J, Frank ED. Available at no cost. http://www.diquad.com/QC%20Book.html IAEA 15.3: Optimization of protection in radiography 22