IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 16.2: Optimization of protection in fluoroscopy Practical.

Download Report

Transcript IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 16.2: Optimization of protection in fluoroscopy Practical.

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
RADIATION PROTECTION IN
DIAGNOSTIC AND
INTERVENTIONAL RADIOLOGY
Part 16.2: Optimization of protection in
fluoroscopy
Practical exercise
IAEA
International Atomic Energy Agency
Overview / Objectives
• To become familiar with quality control tests
in fluoroscopy
• To measure the maximum dose rate to the
patient
IAEA
16.2: Optimization of protection in fluoroscopy
2
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 16.2: Optimization of protection in
fluoroscopy
Maximum dose rate
IAEA
International Atomic Energy Agency
Fluoroscopy - Maximum Dose Rate
Purpose :
• to know the maximum dose rate to the patient
Method :
• place ~2mm lead over the image intensifier
• place a dosimeter ~ 30 cm from the intensifier
• operate unit at maximum kVp and mA
• measure dose rate
• X ray tube below the patient table
IAEA
16.2: Optimization of protection in fluoroscopy
4
Set-up for
measurement of
maximum dose
rate.
IAEA
16.2: Optimization of protection in fluoroscopy
5
The ionization chamber should be protected to
avoid pressure and possible malfunctioning. It
should be in contact with the acrylic to measure
also backscattering.
IAEA
16.2: Optimization of protection in fluoroscopy
6
The intensifier
should be as far as
possible from the
table with the x-ray
tube as close as
possible to the
patient position.
These are the worst
conditions in terms
of maximum
radiation dose to the
skin patient
IAEA
16.2: Optimization of protection in fluoroscopy
7
Note: the X-ray tube is placed very
close to the table.
IAEA
16.2: Optimization of protection in fluoroscopy
8
A radio-opaque marker (a key) can be used to
center the chamber in the image field
IAEA
16.2: Optimization of protection in fluoroscopy
9
The key shows the motion of the C-arm
or the table.
IAEA
16.2: Optimization of protection in fluoroscopy
10
The key should be removed after having
centred the chamber.
IAEA
16.2: Optimization of protection in fluoroscopy
11
A folded lead apron
is used to measure
the maximum dose
rate (use enough
layers to
approximate 2 mm
of lead). The lead
apron is placed over
the protected
chamber.
IAEA
16.2: Optimization of protection in fluoroscopy
12
Note: This geometry is the worst in terms of patient
radiation dose— the tube close to the table and the
intensifier far from the patient. It also increases the
scattered radiation dose to the physician and staff.
IAEA
16.2: Optimization of protection in fluoroscopy
13
The dose rate is
79.5 mGy/min
for this situation
IAEA
16.2: Optimization of protection in fluoroscopy
14
The image quality is poor. The chamber
edge is hardly visible.
IAEA
16.2: Optimization of protection in fluoroscopy
15
The generator selects the maximum
possible kV (110 kV for this equipment)
and 3.6 mA.
IAEA
16.2: Optimization of protection in fluoroscopy
16
In the high contrast mode with the same
geometry, the maximum of 110 kV and
7.34 mA is selected by the generator.
IAEA
16.2: Optimization of protection in fluoroscopy
17
The image quality is still poor although
slightly improved.
IAEA
16.2: Optimization of protection in fluoroscopy
18
In the high contrast mode the maximum
dose rate is 186.7 mGy/min.
IAEA
16.2: Optimization of protection in fluoroscopy
19
Fluoroscopy - Maximum Dose Rate
Analysis :
• FDA limits:
• normal-dose fluoroscopy: maximum air-kerma rate 88
mGy/min (30 cm from the image receptor)
• High-dose-rate fluoroscopy: maximum air-kerma rate
176 mGy/min (30 cm from image receptor)
Frequency :
• acceptance, 6-monthly
• tube change, generator repair
IAEA
16.2: Optimization of protection in fluoroscopy
20
Importance of quality assurance (QA)
• It is very important to have a QA
programmes established in your practice
and not to base dose controls on
specification provided by the manufacturer
or on the additional checks performed by the
manufacturer.
IAEA
16.2: Optimization of protection in fluoroscopy
21
Where to Get More Information
Quality Control in Diagnostic Imaging, Gray JE,
Winkler NT, Stears J, Frank ED. Available at no
cost. http://www.diquad.com/QC%20Book.html
IAEA
15.3: Optimization of protection in radiography
22