48 Hour Tool Procedure
Download
Report
Transcript 48 Hour Tool Procedure
48 hour tool
Step by Step
Purpose of 48-hour tool
Available for NC and RC
Removal of placed implant from patient
Substitute for transfer part after transfer part failure
Implant needs to be discharged after removal
Can be done up to 48 hours after placement,
afterwards osseointegration is too far advanced!
Do not use for height correction
STRAUMANN
Education
2
Situation
Implant stuck & transfer
piece breaks
STRAUMANN
Education
3
Situation
Implant stuck & transfer
piece breaks
Transfer part has
predetermined breakage
point:
RC: 100 Ncm
NC: 80 Ncm
Up until this point, the
internal connection will not
be damaged
The breakage point is
located at the head of the
transfer screw
STRAUMANN
Education
4
Removing the Implant
Remove adapter with
screw head
STRAUMANN
Education
5
Removing the Implant
Remove transfer nut from
implant (no unscrewing
needed)
The remaining screw
remains secure within the
implant
CAUTION
Make sure the transfer nut
is not aspirated
STRAUMANN
Education
6
Removing the Implant
Remove remaining screw
with tweezers or pliers by
counter clockwise rotations
CAUTION
Make sure the remaining
screw is not aspirated
STRAUMANN
Education
7
Removing the Implant
Mount 48-hour tool like an
abutment
Secure tool by handtightening the mounting
screw
STRAUMANN
Education
8
Removing the Implant
Mount ratchet and slowly
unscrew implant
Remove and sacrifice the
implant
STRAUMANN
Education
9