Experiment 5. Spectrophotometric Determination of an
Equilibrium Constant
In this experiment, spectra are obtained for different starting mixtures of I2
and mesitylene in n-heptane. The spectra are recorded on a Shimadzu UV-2101
spectrophotometer (below), which is equipped with a temperature-controlled cell
compartment. The cell holder accommodates 6 sample cells and the reference
cell. Light traverses these cells rightleft. Take care to orient the cells
accordingly when you place them in the holder.
2
It is important that you come to lab having
already planned how you will prepare the several
mixtures you will study. You will use volumetric
glassware —pipettes, graduated cylinders and
volumetric flasks (six 5-mL, one 10-mL, two 25mL). Make sure you understand pipetting
techniques. For example, the pipette bulb is not
“installed” on the end of the pipette, rather is just
brought into gentle contact to achieve an
adequate seal. Also, pipettes are just allowed to
drain, are not “blown” empty. If no-one in your
team is acquainted with such methods, ask the TA
for a demonstration.
3
Cleanliness is extremely important in this experiment. Oil from fingerprints on the cuvettes can give measurable absorbance. Also, the slightest
contamination by acetone in the reaction mixes can ruin the experiment. One
reason we start recording spectra at the long-wavelength limit of 800 nm is to
check against such effects. The absorbance in this region should be very slight
(< 0.01 au), so if at any time your spectra show greater absorbance than this for
l  800 nm, check with the instructor.
At the end of the period, make sure that you put all remaining reagents in
the waste bottle provided. All glassware must be left spotlessly clean and dry
for the next team. (Wet with acetone won’t do, for reasons mentioned above.)