Fundamental Measurements
Download
Report
Transcript Fundamental Measurements
Fundamental Techniques and
Measurements
Mass
Measurements
Volume Measurements
Preparation of a solution of known
concentration
UV-Visible Spectrophotometer
Electronic Balance
What
does an electronic balance measure?
_____
force
If you took an electronic balance with a
capacity of 100 g to the moon what would
its range be? _____
600 g
Mass: Electronic Balance
Accuracy
4 to 6 significant
digits
Calibration
Use known mass
Check weekly or
when balance is
moved
Sources of error
Balance must be calibrated and
maintained in same orientation in
_________
field
gravity
hydroscopic chemicals: dry to
constant mass first (will increase in
mass rapidly as they reabsorb water
on the balance!)
When preparing a solution of a given
concentration it may be difficult to get
the exact mass desired
evaporation of wet samples
Electronic Balance
M o d el
D I-100
D I-800
D I-5000
For
Cap acity Reso lutio n
100 g
0.0001 g
800 g
0.01 g
5000 g
0.1 g
maximum accuracy use balance with
lowest
_______
capacity possible!
Don’t forget to clean the balance if you spill
any chemicals!!!!!!
Volume
Volumetric
accuracy
Graduated
accuracy
flask
0.16 mL
of ______/100
mL
cylinder
0.6 mL
of ______/100
mL
Beaker
accuracy
5 mL
of _____/100
mL
What will accuracy of
solution be if you use
pipette, volumetric
flask, and electronic
1%
balance? ________
What controls the
Pipette
accuracy? _______
Pipette
accuracy
0.6% for 100-1000 µL
of ± _____
0.8% for 10-100 µl
accuracy of ± _____
Digital Pipettes
Air
displacement
Do not directly contact fluid volume
avoids
contamination of pipette
avoids sample carryover
Require
body
air tight connection between tip and
Pipette Workings
piston
cylinder
Pipette tip
Pipettes: Sources of Error
Jetting
Incorrect
transfer technique (getting too
much sample) Wipe tip on container to remove droplets
Contamination from previous samples
Viscous fluids
Hot or cold fluids
Fluids with high vapor pressure
Preparation of Solutions
Example: Prepare 100 mL of a 30 mM solution of
methylene blue.
The molecular weight of methylene blue
(C16H18N3SCl) is 319.87 g.
CV M
30 x 10-3 mole MB 319.87 g MB
mole MB
L
concentration
conversion
100 x 10-3 L = 0.9596 g MB
volume
mass
Preparation of Dilutions
Prepare
100 mL of a 300 µM solution from
the 30 mM solution
Conservation
mass
of _____
M dilute M concentrat e
CdiluteVdilute Cconcentrat eVconcentrat e
Vconcentrate
Vconcentrate
CdiluteVdilute
Cconcentrate
300 M fa
100 mLf
a
a30 mMf
= 1 mL
UV-Visible Spectrophotometer
Theory
Instrument
Sample
requirements
Software
Light Attenuation by an Aqueous
Solution
P0
dP
P
kP
dx
P0
dP
P
dx
P is light intensity
(photons/s)
P
dP
P0
p
x
kdx
0
P
ln kx
P
0
k C
Theory: Light Attenuation = f(?)
For
a given excitation process, a molecule
absorbs only one discrete amount of energy:
expect very narrow absorption lines.
Different vibrational and rotational states
yield _______
broad absorption lines.
Exponential decay with distance
A log
Po
P
= bc
A=bc
Po
- _________
incident light intensity
P light intensity after passing through
sample
length
b - path
______________
c - ______________
concentration
- ___________
extinction coefficient (function of
wavelength and molecule)
Absorption Spectra
Absorption Spectra for Methylene Blue
Broad peaks
blue
red looks ______
Absorbs _____,
Instrument Light Path
Diode Array
Spectrograph Lens
Grat ing
Slit
Sam ple Cell
Shut ter
Source Lens
Deut erium Lamp
Absorbance Measurement
Limitations
lamp
is a function of the _____.
If absorbance is high what is P? ______
small
Suppose A = 3, what is Po/P? _____
1000
Suppose I create samples of higher and
higher concentration. What will happen to
the absorbance measurements?
Po
minimum (non zero) P that
There is a _________
A log = bc can be measured by an instrument.
P
A _______
doesn’t keep increasing! Amax 3
Po
Sample Requirements
Sipper
cell
peristaltic
pump draws sample into sipper cell
requires a few mL to displace previous cell
contents
sample
Light source
pump
detector
Software
Reference (single sample)
Standards (multiple samples)
subtracts absorbance of sample
cell and reference solution
usually distilled water or reagent
blank
calibration
used to create a __________
curve
Samples (multiple samples)
after sampling standards can be
used to estimate the
concentration of samples
Maximum Absorbance
Max absorbance f()
lamp intensity
________________
detector sensitivity
________________
cell absorbance
________________
reference absorbance
________________
acceptable error
________________
absorbance readings
that exceed this value
will not be used in
analysis
Standards
your name
general description
rinse time
sample time
sample concentrations
select number of
samples by moving
this control
Samples
enter sample
descriptions here
select number of samples
by moving this control