Transcript HPLC 10

Sequence Automation
In This Section, We Will Discuss:
 How to set up a sequence.
 How to perform automated recalibrations.
 How to do bracketing.
 How data files are named.
 How to use Sequence Summary.
 How to run the sequence.
 The sequence container (folder).
2
Building a Sequence for Automated Analysis
1. Go to the Method and Run Control view.
2. Go to View > Preferences and set up sequence folder
naming convention.
3. From the Sequence menu, select New Sequence.
3. Edit the Sequence Parameters.
4. Edit the Sequence Table.
5. Select a Sequence Output.
6. Save the Sequence
7. Run the Sequence form the RunControl menu or
the Sequence Table.
Note: Hypersequence is a feature for the Agilent 1100/1200
Series Well plate Handler only.
3
Preferences
•Specifies naming convention for sequence
data containers.
4
Sequence Parameters
5
Explicit Calibration
Calib
Calib
Calib
Samples
Samples
Standards
Standards
Standards
Update
Calibration
Update
Calibration
Calculate
Sample
Amounts
After Each
Injection
Update
Calibration
Calculate
Sample
Amounts
After Each
Injection
6
Sequence Table Explicit Recalibration
Additional
Columns in
Sequence Tables:
•Sample Amount
•ISTD Amount
•Multiplier
•Dilution
•Data file
•Inj Volume
•Lims ID
•AutoBalance
7
Customize Sequence Table
8
Explicit Plus Cyclic Recalibration
9
Bracketing Calibration
Injection
Calib
Calib
Calib
Samples
Samples
Standards
Standards
Standards
Update
Calibration
Update
Calibration
Calculate
Sample
Amounts
For Each
Injection
Calculate
Sample
Amounts
For Each
Injection
10
Bracketed Sequence Operation
11
Partial Sequence Preview of Bracketing
Will show you the order of injections.
12
Fill in Sequence Table Using Sequence Table Fill
Down Wizard
13
Sequence Output
A Sequence Summary Report
allows you to select from a
simple one page Sample
Summary to a comprehensive
report documenting all aspects
of the analysis including crosssample statistics.
14
Edit Sequence Summary
Statistics calculated in
Standard or Extended
format.
15
Extended Statistics
16
Save the Sequence File
Save Current Sequence
17
Sequence Sampling Diagram
Shows sequence tray set-up and
sequence processing status. Click to
set-up sequence. Select Start to
begin sequence.
Select from the View menu.
18
Start a Sequence
•RunControl menu
•Start button
•F6
•Sequence Table
•Partial Sequence Table
•When you start a sequence, a sequence folder is created in the DATA
directory.
•The sequence folder is named according to the subdirectory specified in
Sequence Parameters, and the naming convention specified in Preferences.
•A copy of the master methods (Chem32\1\Methods\names.M) and sequence
template file (Chem32\1\Sequences\name.S) are copied to the folder.
The sequence acquisition is run from these copies.
19
Sequence Folder Data Structure
Sequence Container (Folder)
Subdirectory sequencetemplatename year-month-day hour-minute-second
C:\Chem32\1\DATA\03-15-06\CLASSSEQUENCE 2006-03-15 10-46-09
Acquired Data
Copy of Master Method
Batch file for Batch Review
Copy of Sequence Template
Sequence Logbook
20
Sequence Data Files
Format for Automatic Injection: 003-0101
003 Vial Number
01 Sequence Line
01 Replicate
Prefix
Prefix000001
Prefix000002, etc. for a total of 15 characters
Format for a Cyclic Calibration Standard: C1-02001
C Indicates a cyclic calibration standard was used
1 Calib Line number
02 Sequence Line
001 Replicate
21
Recalibration Settings
22
Calibration Report Options
23