Vital Signs Monitor
UConn BME 4900
Vital Signs Monitor
Purpose
As the population ages, many
people are required by their
doctors to take vital signs on a
daily basis.
Developed for the average
person to use at home.
Vital signs transmitted to a
computer and ultimately to a
doctor.
Measurements
What can it measure?
Electrocardiogram
(ECG)
Three leads (Right
arm, left arm, right
leg)
What is an ECG?
Records a time waveform of heart electrical
activity.
Used to diagnose heart problems
Arrhythmia
Myocardial infarction
Conduction blocks (bundle branch block)
ECG Trace
Cardiac Electrical Activity
The ECG Waveform
Measuring Electrical Activity
3 leads
2 arm
1 leg
As cardiac muscle depolarizes it creates
a potential on the skin
Potential measured as a vector
Difference between right and left
electrode measured with a differential
amplifier.
Measuring Electrical Activity
Circuit consists of:
3 op amp differential amplifier
Low pass filter (Cutoff: 0.05 Hz)
High pass filter (Cutoff: 160 Hz)
Gain Amplifier
Differential Amplifier
Feedback
Amplifier
Differential
Amplifier
Buffer
What does the buffer do?
Provides a high input impedance which means low current flow
What is the purpose of C2 and C3?
Remove DC offset
Offset would be amplified by 1000x
10 mV at 1000x gain = ~10V
Differential Amplifier
V1 & V2 are in the
microvolt range
We need it to be in the
milivolt range so Rf/R1 is
set to ~1,000
1st Order LPF
The ECG is known to be a
low frequency signal
Usually less than 100Hz
A LPF can be used to
remove the high frequency
noise in the signal
High Pass Filter
ECG frequency range ~0.05 Hz – 150 Hz
Why do we need to eliminate frequencies <0.05
Hz?
Avoid distortion of the ST wave
Microprocessor
Provides Analog to Digital (A/D) conversion
Waveform y(t) sampled at a fixed rate Δt
Voltage read every Δt and converted to a number
If resolution is 8 bits
8
Gives us 2 = 256 counts over 5V range
~19.5 mV per count
Patient Isolation
Bluetooth is wireless technology
Bluetooth provides separation of the patient from
the equipment and line source
Very important to protect patient from electrical
shock
Labview Software
Labview Requirements
Receive serial data from bluetooth
Display raw and filtered EKG
Calculate heart rate