Year 2000 Marketing Outlook

Download Report

Transcript Year 2000 Marketing Outlook

Using the
MiniRAE 3000
MiniRAE 3000 & ppbRAE 3000
MiniRAE 3000 (Model Number PGM7320)
Lamp
Range
Resolution
10.6 eV
0.1 ppm to 15,000
ppm
0.1 ppm
9.8 eV
0.1 ppm to 5,000
ppm
0.1 ppm
11.7 eV
0.1 ppm to 2,000
ppm
0.1 ppm
ppbRAE 3000 (Model Number PGM7340)
Lamp
Range
Resolution
10.6 eV
1 ppb to 10,000 ppm
1 ppb
9.8 eV
0.01 ppm to 5,000
ppm
1 ppb
11.7 eV
0.01 ppm to 2,000
ppm
1 ppb
Agenda
•
•
•
•
•
•
•
•
Self-cleaning
User Interface
Turning on the MiniRAE 3000
Recommended Daily Start-up procedure
Hygiene mode & displays
Alarm signals
Programming mode
Charging station
MiniRAE 3000: Self-Cleaning
Patented RAE Systems
Breakthrough!
• Lamp runs for 4 hours during charging
• Generates small amounts of ozone to scrub
sensor and lamp clean
• Without probe user will see lamp glow purple
during charging
• Increases lamp life
• Drastically increases PID stability and reduces
requirement for cleaning (Refer to TN-165)
MiniRAE 3000: Self-Cleaning duty-cycle
Normal operating
pump
Pump during duty-cycle
• Normal operation, pump shows inflow and outflow
• Duty-cycle, pump shows inflow and pump stopped
• With pump stopped, lamp generates small amounts of ozone which
helps to scrub sensor and lamp clean
• Pump starts, dirt is removed from lamp
• Drastically increases PID stability and reduces requirement for
cleaning (Refer to TN-165)
User Interface
LEDs and flashlight
• User Interface:
– Y/+: tests alarms
– N/– MODE (on)
– Flashlight on/off key
Display
MODE key
Flashlight
on/off key
Control Buttons
• Y/+ always controls left-hand response & increases
numerical values
• N/- controls right-hand response & moves to next digit
• MODE controls center & returns to previous menu
Start-up: Turning On
• Press and hold
• Release Mode key
• Self-test
• Final display
Start-up: Programmable Zero calibration
•
•
•
•
Unit performs Zero calibration after warm-up
“apply zero gas…”
If in fresh air, the M3K can be calibrated,
The ppbRAE 3000 needs Zero grade air, or a
Zeroing kit
Start-up: Lamp Alarm
• “Lamp” alarm may occur on start-up
• This indicates that PID lamp has failed to light
– Wait 1-2 minutes
• If “Lamp” message remains, turn off MiniRAE 3000 and restart
– “Lamp” alarm clears and MiniRAE 3000 is ready for use
• If after restart, “Lamp” message remains, the PID needs service
Start-up: Humidity Check
•
It is important to check
if the instrument
responds to moisture
1. Cup hand over inlet or
breathe into inlet for 10-20
seconds
2. Do not block flow
•
If M2K reads >2 ppm or
ppbRAE reads >500 ppb,
then the sensor needs
cleaning
Hygiene/Search Mode
Two main operating modes:
• Hygiene Mode: The factory default
mode. After warm-up the meter samples
continuously
• Search Mode: After warm-up the pump
shuts off and “Ready…” is displayed.
Hygiene Mode/Basic User
*Dashed (- - -) line is automatic return
Hygiene Mode/Advanced User
*Dashed (- - -) line is automatic return
Hygiene Mode/Advanced User
• Change reference gas without entering
Programming Mode
• Change measurement gas when reference gas
is Isobutylene
Hygiene Mode
•
•
•
•
•
•
•
•
Reading in ppm
Gas information
Calibration needed
Wireless status
Battery status
Datalog
Pump Status
Press N/- to move to next screen
TWA/STEL/PEAK
•
•
•
•
•
TWA reading
STEL reading
Peak reading
Press Y/+ to clear
Press N/- to move to next screen
Date/Time/Temp
•
•
•
•
Date
Time
Internal temperature
Press N/- to move to next screen
Gas Display
•
•
•
•
Calibration Gas
Measurement Gas
Correction factor
Press N/- to move to next screen
PC Communication
• Enter PC communication and stop sampling
• Press Y/+ to enter PC Comm mode
– Press Y/+ to exit after communication is complete
• Press N/- to move to next screen
Alarm signals
Search Mode
• A discrete sampling mode can easily
start/stop datalogging for many points
• Perfect for drum or headspace sampling
• Optional graphic screen
Optional Graphic Screen
• Using ProRAE Studio,
set display to graphic
display
Or
• Leave display in ppm
reading
Search Mode/Basic User
*Dashed (- - -) line is automatic return
Search Mode/Advanced User
• Change reference gas without entering
Programming Mode
• Change measurement gas when reference gas
is Isobutylene
Programming Mode
• Press and Hold N/- and MODE to enter
Programming
• Input password “0000”
• Or Press MODE
Programming Mode
• Navigate using interface of
“Select,” “Back,” and
“Next”
• Press “MODE” to exit
Programming
• Y/+ to increase numerical
values or confirm a
question
• N/- to decrease numerical
values or deny a question
Two-Point Calibration
*Dashed (- - -) line is automatic return
Zero calibration
• Press Y/+ to select
Zero Calibration
• Press MODE to return
to main display
• Press N/- to select
Span Calibration
Zero calibration
• If in fresh air, the M3K can be calibrated,
• The ppbRAE 3000 needs Zero grade air, or a
Zeroing kit
• Press Y/+ to start calibration
Span Calibration
• The menu will return to calibration selection with
Span Calibration highlighted
– Using a Matched-flow regulator, tedlar bag, or 500
cc/min regulator, attach 100ppm Isobutylene for the
MiniRAE 3000 or 10ppm Isobutylene for the ppbRAE
3000
• Press Y/+ to enter Span calibration.
• Press N/- to skip Span calibration and return to
Zero calibration.
• Press MODE to exit and return to the top
calibration menu.
Span Calibration
•
•
•
Turn on your span calibration gas.
Press Y/+ to initiate calibration.
Press N/- to abort calibration
Span Calibration
• Calibration is complete when instrument reads:
Span 1 is done!
Reading = xxx. ppm
• Instrument then returns to Zero Calibration menu
• Select MODE to return to regular Operations
• Instrument will then update settings
Power Off
• Hold
for full 3 seconds
• Audible alarm will beep and unit will countdown
5 seconds
• Release
when you see “Unit Off…”
• Unit will stay on if you release MODE key before
countdown completes
Charging unit
• To Charge unit
– Push instrument down
on charger
– Set back of unit
against back wall of
charger
• To remove unit from
charger
– Push down on
instrument
– Pull unit towards you
Removing battery
• Rubber boot stays on
monitor
• First push down on
tab
• Then pull battery from
instrument
Charging battery
• Use back of charging
station to charge
battery externally
– Push into contacts
– Lay battery flat
• To remove battery
– Push battery towards
station
– Lift battery off of
charging station
Maintenance
• Clean PID Lamp & Sensor
– When display creeps upwards after good zero
– When PID responds to moisture
– When movement of PID results in response on
display
Clean Sensor
Bias Electrode
Sensing Electrode
No dirt build-up to
foster a decrease in
airspace resistance
Dirty Sensor
Bias Electrode
Sensing Electrode
Dirt build-up absorbs water and
breaks down airspace resistance
leading to sensor “leakage” or
moisture response