AC Resistive Wall Wake Field Measurement, Theory

Download Report

Transcript AC Resistive Wall Wake Field Measurement, Theory

Resistive Wall Wakes in the Undulator
Beam Pipe – Theory, Measurement
Karl Bane
FAC Meeting
April 7, 2005
• Work done with G. Stupakov, Jiufeng Tu
see:
SLAC-PUB-10707/LCLS-TN-04-11 Revised, Oct. 2004
SLAC-LCLS-TN-05-6, Feb. 2005
April 7-8, 2005
AC Wake--Measurement, Theory
1
Karl Bane
[email protected]
Outline
• Review ac resistive wall wake calculations given last time
•Analyze reflectivity measurements performed by J. Tu at Brookhaven
•Discuss implications
April 7-8, 2005
AC Wake--Measurement, Theory
2
Karl Bane
[email protected]
Introduction
• in the LCLS, the relative energy variation (within the bunch) induced
within the undulator must be kept to a few times the Pierce parameter;
if it becomes larger, part of the beam will not reach saturation
• the largest contributor to energy change in the undulator is the
(longitudinal) resistive wall wakefield
• earlier calculations included only the so-called dc conductivity of the
metal (beam pipe) wall; we showed last time that:
--if the ac conductivity is included the wake effect is larger
--that the anomalous skin effect is negligible, and can be ignored
--that the wake effect can be ameliorated by going to aluminum,
and to a flat chamber.
April 7-8, 2005
AC Wake--Measurement, Theory
3
Karl Bane
[email protected]
•the free-electron model of conductivity is expected to be valid at
low frequencies; it has two parameters: dc conductivity , and
relaxation time 
• at 295K: Cu--= 5.261017/s, = 2.5210-14 s;
Al --= 3.351017/s, = 0.7510-14 s
•if it is valid, the wake in a round beam pipe can be approximated:
with the plasma frequency
April 7-8, 2005
AC Wake--Measurement, Theory
4
Karl Bane
[email protected]
Point charge wake
April 7-8, 2005
AC Wake--Measurement, Theory
5
Karl Bane
[email protected]
charge—1 nC,
energy—14 GeV,
tube radius—2.5 mm,
tube length—130 m
induced energy deviation
for round chamber
April 7-8, 2005
AC Wake--Measurement, Theory
6
induced energy deviation
for flat chamber
Karl Bane
[email protected]
Table I: Figure of merit, E (minimum total energy variation over 30
m stretch of beam), for different assumptions of beam pipe
shape and material. Nominally, vertical aperture is 2a= 5 mm.
Note that Cu-dc results are not physically realizable.
--figure of merit only gives rough idea of effect in LCLS; better to use
analytic model of Z. Huang and G. Stupakov or simulations
April 7-8, 2005
AC Wake--Measurement, Theory
7
Karl Bane
[email protected]
FTIR Reflectivity Measurements
(J. Tu talk)
~  ~( )  n  ik
n
4~ ( )
41
~
 ( ) 
i  1 
i


~
Eref
n~  1
i ( )
~
~   ( )e
n 1
Einc
Kramers-Kronig relation: causality
0
 ( 0 ) 



ln[ R( )]  ln[ R( 0 )]
0
April 7-8, 2005
AC Wake--Measurement, Theory
 02   2
8
We can measure
~ 2 ~ 2
Eref
n 1
R ~
 ~
n 1
Einc
Karl Bane
[email protected]
Measured conductivity compared with calculation
Kramers Kronig analysis of
measurements
Calculation using Ashcroft-Mermin , 
--1(0) is dc conductivity
--Kramer’s Kronig analysis of measurements gives dc conductivity factor
1/2.5 of expected value
April 7-8, 2005
AC Wake--Measurement, Theory
9
Karl Bane
[email protected]
Behavior of R for free-electron model
Frequency k vs. reflectivity R for a metallic conductor, assuming the free
electron model (solid line). For this example =1.21016 /s and =5.410-15
s. Analytic guideposts are also given (dashes).
--for LCLS bunch interested in : [10,100] m, or k: [0.06,0.6]m-1
April 7-8, 2005
AC Wake--Measurement, Theory
10
Karl Bane
[email protected]
Brookhaven reflectivity data for Cu and evaporated Al
range of interest
Measurement results from Brookhaven: reflectivity R vs.
frequency k for copper and evaporated aluminum.
--Cu absorption band at 10 m-1 also seen in plot in Ashcroft-Mermin
April 7-8, 2005
AC Wake--Measurement, Theory
11
Karl Bane
[email protected]
Al survey of measurements, Shiles, et al, 1980
range of interest
1evk= 5 m-1
--also slope in R(), but only ½ as steep
--clear absorption resonance, only hinted at in Brookhaven results
April 7-8, 2005
AC Wake--Measurement, Theory
12
Karl Bane
[email protected]
Al fit
region of interest
Aluminum reflectivity: comparison of measurements (blue) with
calculations using nominal ,  (green); and fitted values: 0.63 nominal ,
0.78 nominal  (red). The position of k= 1/c for the fit is also shown.
--in literature no data below 0.2 m-1
April 7-8, 2005
AC Wake--Measurement, Theory
13
Karl Bane
[email protected]
Cu fit
region of interest
Copper reflectivity: comparison of measurements (blue) with calculations
using nominal ,  (green); and fitted values: 0.66 nominal , 0.67 nominal
 (red). The position of k= 1/c for the fit is also shown.
--in literature no data below 0.3 m-1
April 7-8, 2005
AC Wake--Measurement, Theory
14
Karl Bane
[email protected]
For Al, how does the fit taken affect
E for the LCLS?
Consider (,):
nom.– (3.351017/s, 0.7510-14 s)
fit– (2.121017/s, 0.5810-14 s)
model2– (1.681017/s, 0.4810-14 s)
April 7-8, 2005
AC Wake--Measurement, Theory
15
Karl Bane
[email protected]
Conclusion and Discussion
--The resistive wall wakefield in the beam pipe of the LCLS undulator will
induce a significant energy variation within the bunch; it seems that this will
inhibit much of the beam from lasing.
--This effect can be ameliorated by going from a pipe made of Cu to one of
Al (by ~2), and from one with a round to a flat cross-section (by ~30%).
--To really quantify this, one should go to analytical calculation of Z. Huang
and G. Stupakov, or simulations (S. Reichle and W. Fawley)
April 7-8, 2005
AC Wake--Measurement, Theory
16
Karl Bane
[email protected]
Brookhaven Measurements
--fits (/nom, /nom) are approx. (0.65, 0.80) for Al, and (0.65, 0.65) for
Cu; fit is reasonably good, but does not cover desired frequency (k) range
for Cu (only 40%)
--At higher frequencies the measured R varies linearly with frequency,
unlike the constant dependence expected by the free-electron model.
This point should be understood before giving too much credence to our
results
--The deviation in R for Cu above 0.3 m-1 is small and may not have
much effect. However, it is difficult to know precisely the implication for
the wakefield of a copper pipe: reflectivity at normal incidence alone,
without a model, is not enough information to make such a calculation.
--Measurements of more Al samples are in the works.
April 7-8, 2005
AC Wake--Measurement, Theory
17
Karl Bane
[email protected]