Lecture X Gas Chromatography

Download Report

Transcript Lecture X Gas Chromatography

Lecture X

Gas Chromatography

Outline

GC Theory

 What are the separations?

Instrumentation

Applications

Conclusions

Brief request for next week’s lecture FT-IR.

Gas Chromatography (GC)

This method depends upon the solubility and boiling points of organic liquids in order to separate them from a mixture. It is both a qualitative (identity) and quantitative (how much of each) tool.

 

GC Theory

An inert gas such as helium is passed through the column as a port which is much hotter than the column and is vaporized. The gaseous sample mixes with the helium gas and begins to travel with the carrier gas through the column. As the different compounds in the sample have varying solubility in the column liquid and as these compounds cool a bit, they are deposited on the column support. However, the column is still hot enough to vaporize the compounds and they will do so but at different rates since they have different boiling points. The process is repeated many, many times along the column. Eventually the components of the injected sample are separated and come off of the column at different times (called "retention times").

There is a detector at the end of the column which signals the change in the nature of the gas flowing out of the column. Recall that helium is the carrier gas and will have a specific thermal conductivity, for example. Other compounds have their own thermal conductivities. The elution of a compound other than helium will cause a change in conductivity and that change is converted to an electrical signal. The detector, in turn, sends a signal to a strip chart recorder or to a computer. Detectors come in several varieties, for example, thermal detectors, flame-ionization and electron capture detectors.

Additional Information

From http://www.shu.ac.uk/schools/sci/che m/tutorials/chrom/gaschrm.htm

Theory

In order to understand GC, need to focus on the general principles of separations which has its roots in solvent extraction. The theory of solvent extraction are used to explain all forms of chromatography i.e. HPLC, LC, GC, EP, and TLC.

Consider two solvents S1 and S2 and solute X that is in S1. The partition between the two phases.

X(S1)

X(S2);

GC Theory

Partition coefficient K is an equilibrium constant is K = [X]S 1 /[X]S 2 . Suppose that solute X in V 1 (water) is extracted with V 2 (CCl 4 ). Let m be the moles of X in the system and let q be the fraction of X remaining in phase 1 at equilibrium. The molarity in phase 1 (water) is therefore q m /V 1 .

GC Theory

The fraction of total solute transferred to phase 2 (CCl 4 ) is (1-q) and the molarity in phase 2 is

(1-q) m / V 2 . Then

K = ((1-q) m / V 2 )/ q m V 1 .

For GC:

K = C s /C m where C s is the concentration in the stationary phase (column) and C m is the concentration in the mobile phase (gas).

Suppose I

2

in H

2

O is 1 and I

2

in CCl

4

is 0.

After shaking and letting settle, I 0.0014 and I

q n 2 in CCl 4 2 in H 2 O is

K = 7 x 10 2 . Fraction (q) remaining after the 1st extraction is:

q = V 1 /(V 1 +KV 2 ). Fraction remaining after n = (V 1 /(V 1 +KV 2 )) n . It is more efficient to do extraction can be used to describe GC where the liquid is the mobile phase and the column is the stationary phase. Each extraction is a theoretical plate.

Start

Theory of separation

Water + I 2 Water + I 2 CCl 4 Water + I 2 CCl 4 + I 2 CCl 4 + I 2 Shake 1 Shake 2

Theory GC

As the gas moves the solute (analyte) through and over the stationary phase, the solute will be in equilibrium with the gas and the solid phase. Since there is a mobile phase, the separation will appear as a chromatogram showing the separation of the analytes.

Diffusion in and out of column pores

Solute in and out of packing by diffusion.

GC Theory

Analyte Column + packing To detector Time

Separations To detector Time 1 Time 2 Time 3

Capacity Factor (k’)

• While inside the column, a retained component spends part of its time on the stationary phase and part time in the mobile phase

   

• When in the mobile phase, solutes move at the same speed as the mobile phase • this means that all solutes spend the same amount of time in the mobile phase (t o ) • the amount of time the solute spends on the stationary phase is equal to t R - t o (adjusted retention time, t’ R ) • the ratio t’ R / t o is the capacity of the column to retain the solute (k’)

t R t o t’ R .

Unretained Solute k’ = (t r - t 0 ) / t 0 k’ = (t’ r ) / t 0

GC Process

Column Efficiency (N)

Solutes are placed on an GC column in a narrow band

• Each solute band spreads as it moves through the column due to diffusion and mass transfer affects

• The later eluting bands will spread more

• Peak shape follow a Gaussian distribution

t o t 1 t 2 Band spreading eventually causes peaks to merge into the baseline. We want to minimize band spreading as much as possible.

Chromatogram nomenclature

GC Peak parameters

Typical separation

GC Peak fitting

GC General

The chromatogram shows the order of column), the time of elution (retention time), and the relative amounts of the elution is related to the boiling points and polarities of the substances in the mixture. increasing boiling point but occasionally the relative polarity of a compound will cause it your sample.

Elution Order

       

Compound pentane hexane cyclohexane isooctane toluene Boiling Point ( 36 69 80 99 110 4-methyl-2-pentanone 117 octane 126 0 C)

Example Chromatogram

The observed elution pattern appears below. Notice the reversed elution of toluene and 4-methyl-2-pentanone.

GC Chromatogram

Column performance

Plate height H = L / N where L is the length of the packed column and N is the number of theoretical plates.

For example: A solute with a retention time of 407 s has a width at the base of 13 s on a 12.2 m long column.

N = 16 * 407 2 / 13 2 = 1.57 X 104.

H = 12.2 / 1.57 X 104 = 0.78 mm;

Van Deemter equation:

H = A + B /

x + C

x . All three terms A, B, and

A is for multiple paths, B / longitudinal diffusion,

x

C

x is due to is due to equilibration time.

Peak performance

Longitudinal diffusion is important because diffusion takes place along the axis of the column and contributes to peak broadening. The faster the linear flow, the less time spent in the column and the less diffusional broadening occurs.

Separation efficiency

Efficiency of separations will depend on:

1. Elution times of the peaks,

2. Broadness of the peaks.

 

2 = 2D m t = 2D mL /

x .

Plate height due to long diff:

H D = 2D length.

m /

x = B /

x where D m is the diffusion coefficient of the solute in the mobile phase and t is time and L is column

Column efficiency

C

x comes from the finite time required mobile phase and the stationary phase. Although some solute is stuck in the mobile phase, most of it will move on and elute. The plate height for finite equilibration time i.e. mass transfer is:

H s

 (

k

'

k

'

d

2  1 ) 2

D S

x

C

x

More column stuff

D s = is the diffusion coefficient of the solute in the stationary phase, d is thickness of the stationary phase, and k’ is the capacity factor. Effected by temperature and thickness of stationary phase.

GC Instrumentation

GC Instrumentation

Carrier gas The carrier gas must be chemically inert. Commonly used gases include nitrogen, helium, argon, and carbon dioxide. The choice of carrier gas is often dependant upon the type of detector which is used. The carrier gas system also contains a molecular sieve to remove water and other impurities.

 

Sample injection port For optimum column efficiency, the sample should not be too large, and should be introduced onto the column as a "plug" of vapour slow injection of large samples causes band broadening and loss of resolution. The most common injection method is where a microsyringe is used to inject sample through a rubber septum into a flash vapouriser port at the head of the column. The temperature of the sample port is usually about 50°C higher than the boiling point of the least volatile component of the sample. For packed columns, sample size ranges from tenths of a microliter up to 20 microliters. Capillary columns, on the other hand, need much less sample, typically around 10-3 mL. For capillary GC, split/splitless injector;

Instrumentation

 

Detectors There are many detectors which can be used in gas chromatography. Different detectors will give different types of selectivity. A non-selective detector responds to all compounds except the carrier gas, a selective detector responds to a range of compounds with a common physical or chemical property and a specific detector responds to a single chemical compound. Detectors can also be grouped into concentration dependant detectors and mass flow dependant detectors. The signal from a concentration dependant detector is related to the concentration of solute in the detector, and does not usually destroy the sample Dilution of with make-up gas will lower the detectors response. Mass flow dependant detectors usually destroy the sample, and the signal is related to the rate at which solute molecules enter the detector. The response of a mass flow dependant detector is unaffected by make-up gas.

Components of GC:

Column, oven, injector, and detector. These parameters (HETP, etc) are affected by the various components of the instrumentation. Perhaps the column is the most important component of the GC. With it, different separations can be accomplished.

See Figure 27-1 Pg 703 of text for instrumentation.

GC Instrumentation

Column temperature

For precise work, column temperature must be controlled to within tenths of a degree. The optimum column temperature is dependant upon the boiling point of the sample. As a rule of thumb, a temperature slightly above the average boiling point of the sample results in an elution time of 2 - 30 minutes. Minimal temperatures give good resolution, but increase elution times. If a sample has a wide boiling range, then temperature programming can be useful. The column temperature is increased (either continuously or in steps) as separation proceeds.

Mobile Phase (gas)

GC Under the hood

GC Column and Oven

Typical GC (dual column)

Sample Injections

Next, the sample injection system. Here it is column as a plug and of a suitable size. Also, the injector should provide consistent and reproducible injections. See Figure 27-3, Pg 704. The micro syringe is used to load the sample onto the column. The syringe should be clean and accurate and gas tight. The syringe is injected through a rubber septum. The septum should be replaced after many injections to insure gas tightness onto the column. An auto sampler can be used to inject the samples. Typical volumes range from 0.2 to 20

a splitter. (See Figure.) A suitable solvent is also necessary for the proper separations and injections.

GC Injector

Carrier Gas

This is the mobile phase and should be pure gas so as not to react with the column or analyte. Gas is usually He, Ar, N 2 , or H 2 . used. He and H2 give better resolution 2 . Pressure is comes with regulators. Can you find where in GC equations that are dependent on pressure?

Columns

The column is the most important component of GC. Here is where the separations take place. All the various equations we discussed above are dependent on properties of the column. There are four types of columns: wall-coated open tubular (WCOT), support coated open tubular (SCOT), micropacked, fused silica open tubular (FSOT), and packed column. The FSOT column is the most flexible. Open tubular is also capillary. Particle size is important because the efficiency of GC column increases rapidly with decreasing particle size of the packing material.

Column

The column sits in a temperature controlled environment that is

0.50. Temperature is very important in GC. Can you remember what equations are affected by temperature? See page 706 Fig. 27-5 for temperature effects on separations. Normally, one does a temperature program to get the various analytes off the column for better separations (resolutions).

Columns

  

There are two general types of column, packed and capillary (also known as open tubular). Packed columns contain a finely divided, inert, solid support material (commonly based on diatomaceous earth) coated with liquid stationary phase. Most packed columns are 1.5 - 10m in length and have an internal diameter of 2 - 4mm.

Capillary columns have an internal diameter of a few tenths of a millimeter. They can be one of two types; wall-coated open tubular (WCOT) or support-coated open tubular (SCOT). Wall-coated columns consist of a capillary tube whose walls are coated with liquid stationary phase. In support-coated columns, the inner wall of the capillary is lined with a thin layer of support material such as diatomaceous earth, onto which the stationary phase has been adsorbed. SCOT columns are generally less efficient than WCOT columns. Both types of capillary column are more efficient than packed columns.

In 1979, a new type of WCOT column was devised - the Fused Silica Open Tubular (FSOT) column;

FSOT column

Capillary column

Detectors

How is the analyte detected? Several detectors

FID (flame ionization detector) is the most widely used detector. See figure 27-6, Pg 707. Based on the production of ions when compounds are burned then detecting the current produced from the ionization. What compounds can not be detected with this detector?

TCD (thermal conductivity detector). Operates on the changes in the thermal conductivity of the gas stream brought about by the presence of analyte molecules. See Figure 27-7 on page 708. He is the carrier gas most often used with this detector because it has a high thermal conductivity.

Detection

The effluent from the column is mixed with hydrogen and air, and ignited. Organic compounds burning in the flame produce ions and electrons which can conduct electricity through the flame. A large electrical potential is applied at the burner tip, and a collector electrode is located above the flame. The current resulting from the pyrolysis of any organic compounds is measured. FIDs are mass sensitive rather than concentration sensitive; this gives the advantage that changes in mobile phase flow rate do not affect the detector's response. The FID is a useful general detector for the analysis of organic compounds; it has high sensitivity, a large linear response range, and low noise. It is also robust and easy to use, but unfortunately, it destroys the sample.

Flame Ionization Detector

Detectors continued

ECD (electron capture detector). Uses Ni-63 as a radiation source to cause ionization of the substance using N as the carrier gas. Detector is sensitive to functional groups containing electronegative species such as halogens, quinones, peroxides, and nitro groups. Hence, very good detector for environmental analysis where pesticides need to be measured. See Figure 27-8 page 709.

AED (atomic emission detector). It is a AA unit using MIP that accepts the output from the GC. See Figure 27-9, Page 709.

Mass spectrometer

Different GC detectors

PDD (pulsed discharge detector)

In the electron capture mode, the PDD is a selective detector for monitoring high compounds. For this type of compound, the minimum detectable quantity (MDQ) is at The PDD in helium photoionization mode is an excellent replacement for flame ionization detectors in petrochemical or refinery environments, where the flame and use of for

PDD detector

Applications

Pesticides

Analysis of pesticide residues in soil, water, and food is crucial for maintaining safe levels in the environment. The PDD in the ECD mode is highly selective for monitoring electron capturing compounds such as chlorinated pesticides and other halogens. This chromatogram illustrates the sensitivity of the non-radioactive PDECD for such compounds.

Sample: Pesticide calibration mix Detector mode: Electron capture Detector temp: 330°C Column: 25 m x 0.32 mm x 25 µm, HP-5 Column temp: 150°C to 300°C at 10°C/min Sample volume: 1 µL, 10:1 split Discharge gas: Helium, 30 mL/min Dopant gas: 5% methane in helium, 2.4 mL/min Attenuation: 1

Pesticide separations

Retention time (sec)

Headspace gas chromatography analysis

Headspace GC (HSGC) analysis employs a specialized sampling and sample introduction technique, making use of the equilibrium established between the volatile components of a liquid or solid phase and the gaseous / vapor phase in a sealed sample container. Aliquots of the gaseous phase are sampled for analysis.

Headspace sampling

Headspace sampling

HSGC

 Examples of HSGC are the forensic analysis of blood and urine alcohol levels, quality and production control of diesel fuel and beer constituents. Aromatic flavors and trace volatiles in foods and soft-drinks are also readily analyzed. and HSGC analysis of volatile free fatty acids produced by bacteria, particularly anaerobic bacteria, enables a fingerprint of the particular microorganisms to be obtained, which assists in the identification of the bacteria.

Food analysis

 Analysis of foods is concerned with the assay of lipids, proteins, carbohydrates, preservatives, flavours, colorants and texture modifiers, and also vitamins, steroids, drugs and pesticide residues and trace elements. Most of the components are non volatile and although HPLC is now used routinely for much food analysis, GC is still frequently used. For examples, derivatization of lipids and fatty acid to their methyl esters(FAMEs), of proteins by acid hydrolysis followed by esterification (N-propyl esters) and of carbohydrates by silylation to produce volatile samples suitable for GC analysis.

GC Food

 

GC quality control analysis of food products can confirm the presence and quantities of the analytes For example, fruits, fruit derived foodstuffs, vegetables and soft drinks, tea and coffee, were analyzed for their polybasic and hydroxy acid contents (citric, maleic acids) as TMS derivatives.

All food and beverage products on sale today must be carefully assayed for contamination with pesticides, herbicides and many other materials that are considered a health risk. The analysis of food involves separating and identifying very complex mixtures, the components of which are present at very low concentrations. GC is the ideal technique for use in food and beverage assays and tests. Furthermore, the origin of many herbs and spices can often be identified from the peak pattern of the chromatograms from their head space analysis.

Food and Cancer

Chemicals that can cause cancer have a wide variety of molecular structures and include hydrocarbons, amines, certain drugs, some metals and even some substances occurring naturally in plants and molds. In this way, many nitrosamines have carcinogenic properties and these are produced in a number of ways such as cigarette smoke. GC can be used to identify these nitro-compounds in trace quantities.

Drugs

 

There are still numerous GC applications involving both quantitative and qualitative identification of the active components and possible contaminants, adulterants or characteristic features which may indicate the source of the particular sample. Forensic analysis frequently users GC to characterize drugs of abuse, in some cases the characteristic chromatographic fingerprint gives an indication of the source of manufacture of the sample or worldwide source of a vegetable material such as cannabis.

Analytical procedures, chromatographic methods and retention data are published for over 600 drugs, poisons and metabolites. These data are extremely useful for forensic work and in hospital pathology laboratories to assist the identification of drugs.

Pyrolysis gas chromatography

 Pyrolysis GC (PGC) is used principally for the identification of non-volatile materials, such as plastics, natural and synthetic polymers, drugs and some microbiological materials. The thermal dissociation and fragmentation of the sample produces a chromatogram which is a fingerprint for that sample. The small molecules produced in the pyrolysis reaction are frequently identified using a GC-MS system and information on molecular structure for identification is also obtained.

Metal chelates and inorganic materials

Although inorganic compounds are generally non-volatile, GC analysis can be achieved by converting the metal species into volatile derivatives. Only some metal hydrides and chlorides have sufficient volatility for GC. Organometallics other than chelates, which can be analyzed directly, include boranes, silanes, germanes, organotin and lead compounds.

Environmental analysis

 Environmental pollution is an age-old trademark of progressed, populations have increased and standards of living have improved. So the demands on the environment have increased, with all the attendant problems for the ecosystems. Combustion of fossil fuel, disposal of waste materials and products, treatment of crops with pesticides and herbicides have all contributed to the problems. Technological developments have enabled man to study these problems and realize that even trace quantities of pollutants can gave detrimental effects on health and on the stability of the environment. There is a vast amount of literature on the use of GC for studying a wide variety of these problems.

GC application

Every year many new substances are synthesized that differ radically from the natural products that exist in biosystems. The Environmental Protection Agency is empowered to control water pollution and the production, use and disposal of toxic chemicals.

It follows that detailed studies must be made of their effect on the environment and their method of movement through the ecosystem. Many of the compounds are not biodegradable and will thus progressively pollute the environment. There are a number of tragic examples of which DDT (dichlorodiphenyltrichloroethane) and the PCBs (polychlorinated biphenyls) are well known instances. The materials of interest are present in environmental samples at very low concentrations and are often to be found among a myriad of other compounds from which they must be separated and identified. It follows that GC, with its inherent high sensitivity and high separating power, is one of the more commonly used techniques in the analysis of environmental samples.

Various configurations:

Column: capillary

Autosampler

Computer controlled for data acquisition and analysis of peaks. Some come with their own compressor for the air supply. Sensitivity is very important because of detection limits. More expensive system will use a MS for the detector to do GC-MS. Can do headspace analysis for volatiles in water etc.

GC Companies

ThermoQuest

Agilent (Hewlett Packard)

Perkin-Elmer

Buck Scientific

Thermo Nicolet

Applications

Go to PDF files

Summary

GC separations are based on a partitioning between the analyte with the mobile phase (gas) and the stationary phase (column).

Remember your equilibrium constants?