Intrinsic Safety Overview Rick Ales 8/20/2003 © Swagelok Company, 2003 Evolution of NeSSI End User Value Gen III Circa 200? Gen II Circa 2005 Micro Analytical (enabled by Technology Advances) By.

Download Report

Transcript Intrinsic Safety Overview Rick Ales 8/20/2003 © Swagelok Company, 2003 Evolution of NeSSI End User Value Gen III Circa 200? Gen II Circa 2005 Micro Analytical (enabled by Technology Advances) By.

Intrinsic Safety Overview
Rick Ales 8/20/2003
© Swagelok Company, 2003
Evolution of NeSSI
End User
Value
Gen III
Circa 200?
Gen II
Circa 2005
Micro Analytical
(enabled by Technology Advances)
By Line
(enabled by Intrinsic Safety)
PnP Smart Devices
(enabled by Communication Standard)
We are Here
Gen I
Now
Rick Ales 8/20/2003
Miniature/Modular
(enabled by Mechanical Standard)
© Swagelok Company, 2003
Plug-n-Play Top Hat Concept
Presented at IFPAC 2002 by Rob Dubois, Dow Chemical.
Rick Ales 8/20/2003
© Swagelok Company, 2003
Hazardous Area Protection Methods
Purge
Mount electric devices in a cabinet
to keep explosive atmosphere out.
Explosion Proof Mount electric devices in a cabinet
to keep explosion in.
Intrinsic Safety
Rick Ales 8/20/2003
Prevent the explosion ignition by
limiting electrical energy.
© Swagelok Company, 2003
FISCO Power Budget
• Fieldbus Intrinsically Safe COncept model
• Developed by PTB1 in Braunschweig, Germany
• Doubles the number of devices per segment.
Max Voltage
17.8 Vdc
Max Current IIC Group Gas
183 ma @ 14 Vdc
Max Current IIB Group Gas
380 ma @ 14 Vdc
Max Power
5.24 W
Max Co / Lo
5 nF / 10 uH per Device
1. Physikalisch-Technische Bundesanstalt (PTB), http://www.ptb.de/en/index.html
Johannsmeyer, "Investigation into the intrinsic safety of fieldbus systems (FISCO)". PTB, report W53, ISBN 3-89429-310-1.
Physikalische Technische Bundesanstalt, "Investigations into the intrinsic safety of fieldbus systems". PTB 1994, ISBN 3-89429-512-0
Rick Ales 8/20/2003
© Swagelok Company, 2003
NeSSI Architecture
Information - Level 2 LAN (Ethernet)
OPC
Gateway
Field - Level 1 LAN (EtherNet)
Control Room
Enclosure
Class 1 Zone 2
24 VDC
CAN Master
Circuit 0
Rick Ales 8/20/2003
IS Barrier
Sensor
Actuator
Manager
Field
By-Line
Sample Point
Circuit 1
Circuit 2
Logical
Segment
Class 1 Zone 0/1
Sensor - Level 0 LAN
DeviceNet <250 ft @14VDC
P
V
A
T
Gas
Analyzer
NeSSI Device Sample System
Wireless
Handheld
HMI