Transcript Slide 1

Nurse Call Systems & Emergency Call
Systems
Overview of ANSI/UL 1069 and
ANSI/UL 2560
The Association of Electrical and Medical Imaging Equipment Manufacturers
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Overview
• National Consensus Standards
 American National Standards Institute (ANSI)
• Nurse Call Systems
• Emergency Call Systems
 Overview of new UL2560 Standard
• Differences between UL1069 and UL2560
• Applicability of Standards
3
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Nurse Call Systems & UL1069
4
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Consensus Standards
• UL1069 complies with all relevant articles of
NFPA70 National Electrical Code and NFPA99
Health Care Facilities Code
• NFPA101 (Life Safety Code) 18.1.1.1.2/3
defines Healthcare facilities to be “…where the
occupants sleep there, but are not reasonably
able to evacuate themselves in an emergency”
 As such, these facilities have 24-hour staff on duty
5
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Nurse Call Systems
• UL1069 Hospital Signaling and Nurse Call
Equipment
 Created in 1977 by UL and Nurse Call industry
 7th edition (October 2007) clarifications/additions
focused on three areas:
• Wireless nurse call devices
• Definition and verification of core system elements
• Definition of oxygen-enriched testing for pendant controls
6
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Nurse Call Systems
Wireless Nurse Call Devices

Wireless initiating devices, as part of the fundamental
system, have clear limitations:
1) The intent of wireless devices is to extend the coverage of a
hard-wired system, however, only initiating devices can be
wireless
2) The fundamental notification devices MUST be wired
3) Wireless devices utilize a “shared” radio frequency (RF)
space
–
–
Not guaranteed to be available or work in all real world
environments
UL1069 adds requirements to assure best possible reliability
7
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Nurse Call Systems
Wireless Nurse Call Devices

Wireless initiating devices, as part of the fundamental
system, have clear limitations:
4) Devices are supervised. Loss of contact with receiver for more
than 90 seconds (including interference, loss of battery power,
damage, etc.) = a supervisory alarm
–
–

This requirement is based on testing the LARGEST possible
system
Note: On systems listed prior to Ed 7, the supervision time may
be 24 hours
Extensive testing required to assure devices will work
in predictable interference scenarios
•
Systems must have designs that “hop” or move frequencies to
work around potential interferences
8
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Nurse Call Systems
Fundamental System Requirements
 UL1069 defines the fundamental nurse call system
devices:
•
•
•
•
•
•
•
Patient Station
Bath Station
Emergency Station
Corridor Light
Console
Power and Control
Duty Station
A system without corridor lights cannot be UL1069 listed
9
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Nurse Call Systems
Oxygen-Enriched Environments
 UL1069-listed systems fully comply with NFPA 99 for
safety of pendant devices (call cords and pillow
speakers) in oxygen-enriched environments
10
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems & UL 2560
11
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL2560 Emergency Call Systems for Assisted
Living and Independent Living Facilities
 1st edition published September, 2011
 Addresses minimum performance of emergency call
(e-call) systems in senior living communities
12
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL2560 Requirements
 Placement
• Permanently fixed e-call station must be located in every
resident bathroom
• Additional optional stations at resident beds, elsewhere in
living areas and common areas are allowed
 Pendants
• Pendants are optional, but if included, must comply with the
standard
• Assigning pendants to residents does not remove
requirements for at least one fixed e-call station in each
resident bathroom
13
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL2560 Requirements
 Call Indicator
• E-call stations and pendants must include call assurance
indication
 Central Notification Station
• There must be at least one notification station to receive all
calls
• Must be at a fixed location
• Additional optional portable notification stations are allowed
14
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL2560 Requirements
 Origination Reporting
• The origin of a resident call must be reported at a
notification station
• The origin must be specific
– e.g.“Apt 117” or “3rd Floor Men’s Room”
 Cancellation
• Call (from fixed station or pendant) must be canceled at the
source of the call
• Exception: If the system can separate non-emergency
resident calls from emergency resident calls and there is
voice communication between the e-call station and
notification station
– Non-emergency calls can be canceled from notification station
after voice contact with resident has been made
15
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
Other UL2560 Requirements





Connection Testing
System Testing
Back-up Power
Low Battery Alarm
Interference Testing
To obtain a copy of UL2560:
http://ulstandards.ul.com/standards-catalog/
16
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Comparison of the Standards
17
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Standards Comparison
UL1069
UL2560
Required Call Notification
Permanently
installed primary
console and
corridor lights
One or more call
notification stations , at
least one of which must
be in a fixed location
Supervision of Wireless
Initiating Devices
Every 90 seconds
Every 24 hours
Assumes patient is
not mobile
Assumes resident is
mobile
Required
Not required
Mobility
Nursing Staff 24/7
18
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Applicability of the Standards
19
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Applicability of Standards
When UL1069 Applies
 Systems covered by the standard are intended to be
installed in either General or Critical Care areas in
accordance with:
• National Electrical Code (NFPA 70)
• Health Care Facilities Code (NFPA 99)
• Life Safety Code (NFPA 101 - Chapter 18)
20
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Applicability of Standards
When UL2560 Applies
 Systems covered by standard are intended to be
installed in:
• Assisted and independent living facilities
– 24/7 staffing is not required
– Residents can reasonably be assumed to evacuate
themselves in an emergency
21
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Importance of Listing
22
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Importance of Listing
• When nurse call systems are listed to UL1069
and emergency call systems to UL2560, you
can count on:
 Product safety for all users
 System reliability
 Corporate integrity to design to the highest
standards now and as new requirements are
implemented
 Peace of mind for staff, patients, residents, and
relatives
23
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Additional Resources
24
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Additional Resources
NEMA Presentation
 A copy of this presentation is available for download at
http://www.nema.org/Products/Pages/Health-CareCommunications-and-Emergency-Call-Systems.aspx
UL Guide Cards for Standards
 UL Guide Cards for the UL 1069 and UL 2560 standards are
available at http://www.ul.com
• Click Nurse Call Systems for a link to the Nurse Call System Guide Card
• Click E-Call Systems for a link to the Emergency Call System Guide Card
Note: for the links to work, go to Slide Show mode
25
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Questions?
Thank You!
26
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Back Up Slides
27
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL 2560
 Connection Testing/Electrical Supervision
• Wireless and telephony based e-call systems
must be self-testing so to report device faults
• Maximum time for fault to reported cannot exceed
24 hours
28
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL 2560
 System Testing/Program Access and Control
• Many modern e-call systems incorporate one or
several microcontrollers
• If one of these microcontrollers fails, a fault
condition must be reported within 90 seconds
29
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL 2560
 Backup Power
• UL 2560 requires a secondary power source that
provide power when main power fails
– Typically accomplished with batteries
– Assisted living/independent living facility can have an
on-site power generator but, if none exists, a secondary
power source must be provided to self-power ECS
• Manufacturer must specify length of time the
system will operate on backup power
– e.g. “With 10% of the emergency call stations in alarm,
backup batteries will continue to power the system for
12 hours”
30
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL 2560
 Low Battery Alarm
• Wireless e-call system transmitters are typically
powered by batteries (replaceable or permanent)
• Low battery must be reported at least 7 days prior
to transmitter failure
• Manufacturer must state expected life of batteries
and conditions that comprise expectation
– e.g. “If five emergency calls are placed each day, the
battery will last at least 38 months”
31
The Association of Electrical and
Medical Imaging Equipment Manufacturers
Emergency Call Systems
UL 2560
 Interference Testing
• Wireless e-call systems have to pass tests to
minimize possibility that operation will be
impaired by interfering signals
32