Partial Parallel imaging (PPI) in MR for faster imaging

Download Report

Transcript Partial Parallel imaging (PPI) in MR for faster imaging

Partial Parallel imaging
(PPI) in MR for faster
imaging
IMA
Compressed Sensing
June, 2007
Acknowledgement: NIH Grants 5RO1CA092004 and 5P41RR008079,
Pierre-Francois Van de Moortele, Gregor Adriany, Kamil Ugurbil
Our coils

Open face coil

16 Channel
“closed” coil
Intrinsically, surface coils offer a representation of signal as
The sensitivities are complex valued
Acquired k-space,
How to INTERPOLATE most stably to the
Non-acquired data.
we will see why it makes completely sense to
think about interpolation
Field of View
Courtesy: Douglas Noll, University of Michigan
Undersampled images
Undersampled individual images
I
unaliased
1
H
1 " reducedFOV "


 S  S  S  y
H
1
The linear system
The solution of the linear system gives rice to a spatially varying noise
amplification. This is solely dependent on the sensitivities and is referred to
as the geometry factor
The geometry factor
α is the index of an aliased pixel,
βn is the index of an unaliased
pixel.
Overall loss when using PPI is SNRred=SNRfull/(g sqrt(R))
The reduced FOV is the RSOS of all the channels with a reduced FOV, only for illu.
Back to the equation
Image space
S indv. Channels. E encoding.
p un-aliased image
K-space
S indv. Channels. E encoding.
p un-aliased image (all in kspace)
Convolution operator
E is known, but we can make the formal separation of S, as follows:
E”acq” includes all of
k-space for the
sensitivities
Two matrix equations, two unknowns

SENSE/SMASH formalism (get one image)
GRAPPA idea, get multiple images. The
interpolation is essentially similar to Kriging
Courtesy: Yeh, et al, MRM Volume 53, Issue 6 , Pages 1383 - 1392
GRAPPA
Reconstructing
the data for
EACH coil
Courtesy: Griswold et al. MRM, 47(6):1202-1210 (2002)
Several reconstruction is found for EACH k-space pointdue to the blocks. A weigthed average is used to compute
just one

ACS (Auto-Calibration Signal) lines (no x)
,l

GRAPPA formula to reconstruct signal
in one channel
where A represents the acceleration factor. Nb is the number of blocks used
in the reconstruction, where a block is defined as a single acquired line and
A-1 missing lines. 4-8 blocks are needed
Temporal sampling
Interleaved/segmented (2)
PE
Interleaved (2)
½ kspace
½ kspace
time
Works well for imaging of static objects.
For dynamic imaging, each image is
not only undersampled, but also
captures a different part of the
“motion”/”change”. The acquisition is
assumed faster than the motion
PSF considerations (generally)


Let us start with imaging
psf  F ()
Standard PPI used to
“unalias” the effects of
the psf
PE or t
UNFOLD (does not require
multichannels)

Specifically, alternate the sampling
by a factor 2, such that
…
½ kspace
Remove aliasing by
Courtesy: Madore. MRM 48:493 (2002).
fMRI (UNFOLD)
Remove aliased
frequency by
selective filtering
FIG. 16. Results obtained for a single-trial fMRI experiment (4 spiral interleaves, 16 kz phase-encode values, axial images, matrix size
128 3 128, TR 5 250 msec, TE 5 40 msec, 5 mm resolution along z, 24 cm FOV). Bilateral finger tapping was performed while a 2 sec audio
cue was on, and then stopped for 12 sec. The acquisition time for a time frame (16 sec) is longer than a paradigm cycle (14 sec). UNFOLD is
used to reduce the acquisition time by a factor 8, providing 7 frames per paradigm cycle. a: The acquired frames are corrupted by an 8-fold
aliasing in the through slice direction. b: Temporal frequency spectrum for the highlighted image point in a. UNFOLD interleaves 8 spectra
into
the same temporal bandwidth. Marks are placed on the axis at the locations of the DC, fundamental and harmonic frequencies for the
non-aliased material. Selecting only these frequencies, the aliasing seen in a is removed in c.
Courtesy: Madore. MRM 48:493 (2002).
Extend the
concept
of
aliasing
Line in image
Unalias the
support in x-f
space, just like we
unalias in x space
with SENSE
Tsao et al, MRM 50: 1031-1042 (2003)
DATA challenge
1.
Where is the support in x-f space?
Interleaved training
set
Used to
define
support in x-f
space
“Equivalent”
to a
reference
scan
Tsao et al, MRM 50: 1031-1042 (2003)
Similar concepts hold
for radial, where the
center is the “prior”.
This is used in speech
imaging
How do the methods compare?
k-t SENSE, vs. Sliding Window

Consensus (in cardiac imaging) of:
Xu et al. MRM, 57:918930 (2007)
What does the artifact mean
Xu et al. MRM, 57:918930 (2007)
Looking at the temporal variation
(in speech
[radial])
Sliding window
K-t SENSE
y
t
Comments/Conclusions
Michael S. Hansen. Workshop on Non-catesian MRI.
2007
Formally
The mising information can be determined from the acquired data,
if the coeeficients a(i,j,k) are known
With localized sensitivities (smooth in image space)
SMASH

Find weights nk(m)(x) [no x –readout dependence] such that we
get a new synthetic sensitivity profile Cmcomp
WE do parallel Imaging by finding ONE combined image (just like SENSE)
m is selected depending on how FAR the data must be interpolated. Only one
line is used to advance the data
Generalised SMASH

Find weights ak(m)(x) [with x –
readout dependence] such that
(x)
Express
s j ( x, k y )   C j  e
q

m  p

ik y y

q
a
m  p
m
j
C j ( x, y) 
( x)   e
q

m  p
a mj ( x)eimky
 i ( k y  mk ) y

a mj ( x)s ( x, k y  mk )
We use several phase-encoding lines to generate missing
information. For each readout point a new set of weights
are comp.
Two severe issues
The final image is that of a complex
sum image of the individual images.
Not optimal for SNR
 Total cancellation can occur with such
complex sums.
 Coils where phase-aligned PRIOR to
reconstruction

AUTO-SMASH

ACS (Auto-Calibration Signal) lines
(no x), not fitting to a harmonic, but
a “missing” PE-line