Transcript Document

Optical Microscopy

• Introduction • Lens formula, Image formation and

Magnification • Resolution and lens defects

Basic components and their functions • Common modes of analysis • Specialized Microscopy Techniques • Typical examples of applications

Diffraction of Light

Diffraction of light occurs when a light wave passes by a

corner

(or a opening

barrier

) (or a or through an slit) that is physically the approximate size of, or even smaller than that light's wavelength.

Sin  =  /d  /d

1 st 2 nd 3 rd film

Light waves interfere constructively and destructively.

Resolution of Microscope – Rayleigh Criteria

Rayleigh Criteria: Angular separation 

of the two points is such that the

central maximum of one image falls

on the first diffraction minimum of the other

=

m

1.22

/d

Resolution of Microscope – in terms of Linear separation

   To express the resolution in terms of a linear separation r, have to consider the Abbe’s

theory

Path difference between the two beams passing the two slits is

d

sin

i

d

sin    Assuming that the two beams are just collected by the objective, then i =  and

d min =

/2sin

I II I II

  

Resolution of Microscope – Numerical Aperture

If the space between the specimen and the objective is filled with a medium of refractive index n, then wavelength in medium  n =  /n The d

min

=  /2n sin  =  /2(N.A.) For circular aperture d

min

= 1.22

 /2(N.A.)=0.61

 /(N.A.) where N.A. = n sin  is called

numerical aperture Immersion oil n=1.515

Numerical Aperture (NA) NA=1 theoretical maximum numerical aperture of a lens operating with air as the imaging medium

Angular aperture

(  72 degrees)

One half of A-A NA

of an objective is a measure of its ability to gather light and resolve fine specimen detail at a fixed object distance.

NA = n(sin

)

n: refractive index of the imaging medium between the front lens of objective and specimen cover glass

Resolution of a Microscope (lateral)

The smallest distance between two specimen points that can still be distinguished as two separate entities

d

min

= 0.61

/NA

NA=nsin () 

illumination wavelength (light) NA – numerical aperture  one half of the objective angular aperture n-imaging medium refractive index d

min

~ 0.3

 m for a midspectrum  of 0.55

 m

  

Factors Affecting Resolution

Resolution = d min = 0.61

 /(N.A.) Resolution improves (smaller d min ) if   Assuming that sin  = 0.95 (  = 71.8°) or

n

 or   Wavelength Red 650 nm Air (n= 1) Oil (n = 1.515) Yellow Green 600 nm 550 nm Blue Violet 475 nm 400 nm  (The eye is more sensitive to blue than violet)

Optical Aberrations

Reduce the resolution of microscope Two primary causes of non-ideal lens action:

• Spherical (geometrical) aberration – related to the spherical nature of the lens • Chromatic aberration – arise from variations in the refractive indices of the wide range of frequencies in visible light Astigmatism, field curvature and comatic aberrations are easily corrected with proper lens fabrication.

Defects in Lens

   Spherical Aberration – Peripheral rays and axial rays have different focal points (caused by spherical shape of the lens surfaces.

causes the image to appear hazy or blurred slightly out of focus .

and very important in terms of the resolution of the lens because it affects the coincident imaging of points along the optical axis and degrade the performance of the lens.

Defects in Lens

Chromatic Aberration

Axial - Blue light is refracted to the greatest extent followed by green and red light, a phenomenon commonly referred to as dispersion  Lateral - chromatic difference of magnification: the blue image of a detail was slightly larger than the green image or the red image in white light, thus causing

color ringing

of specimen details at the outer regions of the field of view

A converging lens can be combined with a weaker diverging lens, so that the chromatic aberrations cancel for certain wavelengths: The combination – achromatic doublet

Axial resolution – Depth of Field Depth of focus (F mm) Depth of Field (F

m)

m) NA F 0.1 0.13

0.4 3.8

.95 80.0

F 15.5

5.8

0.19

The distance above and below geometric image plane within which the image is in focus

M NA M NA F F F F

The axial range through which an object sharpness can be focused without any appreciable change in image F is determined by NA.

Basic components and their functions

camera Beam splitter Reflected light Olympus BX51 Research Microscope

Cutaway Diagram

Transmitted light

Functions of the Major Parts of a Optical Microscope

   Lamp and Condenser: project a parallel beam of light onto the sample for illumination Sample stage with X-Y movement: sample is placed on the stage and different part of the sample can be viewed due to the X-Y movement capability Focusing knobs: since the distance between objective and eyepiece is fixed, focusing is achieved by moving the sample relative to the objective lens

Light Sources

Condenser

Light from the microscope light source

Condenser gathers light and concentrates cone of light that illuminates uniform intensity over the entire viewfield it into a the specimen with

Specimen Stage

Functions of the Major Parts of a Optical Microscope

   Objective: does the main part of magnification and resolves the fine details on the samples (m o ~ 10 – 100) Eyepiece: forms a further magnified virtual image which can be observed directly with eyes (m e ~ 10) Beam splitter and camera: allow a permanent record of the real image from the objective be made on film

Microscope Objectives

d

min

= 0.61

 /NA

Objective specifications Anatomy of an objective rical ture Objectives are the most important components of a light microscope: image formation , magnification , the quality of images and the resolution of the microscope

Eyepiece

(Diaphragm) M=(L/f o )(25/f e ) Eyepieces ( Oculars ) work in combination with microscope objectives to further magnify the intermediate image

Common Modes of Analysis

• •

Depending on the nature of samples, different illumination methods must be used Transmitted OM

thin section of rocks, minerals and single crystals

Reflected OM

-

transparent specimens opaque specimens most metals, ceramics, semiconductors

Specialized Microscopy Techniques

Polarized OM

-

specimens with anisotropic optical character

Characteristics of materials can be determined

morphology (shape and size), phase distribution (amorphous or crystalline), transparency or opacity, color, refractive indices, dispersion of refractive indices, crystal system, birefringence, degree of crystallinity, polymorphism and etc.

camera Beam splitter Reflected light Olympus BX51 Research Microscope

Cutaway Diagram

Transmitted light

Polarization of Light When the electric field vectors of light are restricted to a single plane by filtration, then the the light is said to be polarized with respect to the direction of propagation and all waves vibrate in the same plane.

Polarized Light Microscope Configuration

Typical examples of applications

Grain Size Examination Thermal Etching 1200C/30min

a

1200C/2h

b

20

 A grain boundary intersecting a polished surface is not in equilibrium (a). At elevated temperatures (b), surface diffusion forms a grain-boundary groove in order to balance the surface tension forces.

m

Grain Size Examination

Objective Lens x100

Grain Growth - Reflected OM

5

m 30

m Polycrystalline CaF 2 illustrating normal grain growth . Better grain size distribution.

Large grains in polycrystalline spinel (MgAl 2 O 4 ) growing by secondary recrystallization from a fine-grained matrix

Liquid Phase Sintering

Reflective OM Amorphous phase 40

m Microstructure of MgO-2% kaolin body resulting from reactive-liquid phase sintering.

Image of Magnetic Domains

Magnetic domains and walls on a (110)-oriented garnet crystal (Transmitted LM with oblique illumination).

The domains structure is illustrated in (b).

Polarized Optical Microscopy (POM) Reflected POM Transmitted POM (a)Surface features of a microprocessor integrated circuit (b)Apollo 14 Moon rock

Phase Identification by Reflected Polarized Optical Microscopy

YBa

2

Cu

3

0

7-x

superconductor material: (a) tetragonal phase and (b) orthorhombic phase with multiple twinning (arrowed) (100 x).

Specialized LM Techniques

Enhancement of Contrast Darkfield Microscopy Phase contrast microscopy Differential interference contrast microscopy Fluorescence microscopy-mainly organic materials • Confocal scanning optical microscopy (

new ) Three-Dimensional Optical Microscopy inspect and measure submicrometer features in semiconductors and other materials

Hot- and cold-stage microscopy

melting, freezing points and eutectics, polymorphs, twin and domain dynamics, phase transformations

In situ microscopy

E-field, stress, etc.

Special environmental stages-

vacuum or gases

Contrast

Contrast is defined as the difference in light intensity between the specimen and the adjacent background relative to the overall background intensity.

Image contrast, C is defined by S specimen -S backgroud

S C = = S specimen S A S specimen

and S

backgroud

are intensities measured from specimen and backgroud, e.g., A and B, in the scanned area.

  

Angle of Illumination

Bright filed illumination – The normal method of illumination, light comes from above (for reflected OM) Oblique illumination – light is not projected along the optical axis of the objective lens; better contrast for detail features Dark field illumination – The light is projected onto specimen surface through a special mirror block and attachment in the objective – the most effective way to improve contrast.

I min I max Light stop C= I max I min I max C-contrast

Transmitted Dark Field Illumination Oblique rays specimen I distance I distance

Contrast Enhancement

OM images of the green alga Micrasterias

Crystals Growth-Interference

contrast microscopy Growth spiral on cadmium iodide crystals growing From water solution (1025x).

w

Confocal Scanning Optical Microscopy

Three-Dimensional Optical Microscopy Critical dimension measurements in semiconductor metrology Cross-sectional image with line scan at PR/Si interface of a sample containing 0.6

m-wide lines and 1.0

m-thick photoresist on silicon.

The bottom width, w, determining the area of the circuit that is protected from further processing, can be measured accurately by using CSOP.

Measurement photoresist is important because it allows the of process the patterned engineer to simultaneously monitor for defects, misalignment, or other artifacts that may affect the manufacturing line.

Hot-stage POM of Phase Transformations in Pb(Mg 1/3 Nb 2/3 )O 3 -PbTiO 3 Crystals

 n T(

o

C) (a) and (b) at 20

o

C, strongly birefringent domains with extinction directions along <100> (c) at 240

o

cubic , indicating a tetragonal symmetry; C, phase transition from the tetragonal into cubic phase with increasing isotropic areas at the expense of vanishing strip domains.

E-field Induced Phase Transition in Pb(Zn 1/3 Nb 2/3 )O 3 -PbTiO 3 Crystals

Schematic diagram for in situ domain observa tions.

a b Single domain c Domain structures of PZN-PT crystals as a function of E-field; (a)E=20kV/cm, (b) e=23.5kV/cm (c) E=27kV/cm

Rhombohedral at E=0 and Tetragonal was induced at E>20kV/cm

Review Optical Microscopy

Use visible light as illumination source • Has a resolution of ~o.2

 m • Range of samples characterized - almost unlimited for solids and liquid crystals • Usually nondestructive; sample preparation may involve material removal •Main use preliminary – direct observation visual for observation; final charac terization with applications in geology, medicine, materials research and engineering, industries, and etc.

• Cost - $15,000-$390,000 or more

Characteristics of Materials Can be determined By OM:

morphology (shape and size), phase distribution (amorphous or crystalline), transparency or opacity, color, refractive indices, dispersion of refractive indices, crystal system, birefringence, degree of crystallinity, polymorphism and etc.

Limits of Optical Microscopy

• Small depth of field <15.5

m

Rough surface

• Low resolution ~0.2

m • Shape of specimen

Thin section or polished surface Cover glass specimen Glass slide resin 20

m

• Lack of compositional and crystallographic information

Optical Microscopy vs Scanning Electron Microscopy

25 m radiolarian

OM SEM Small depth of field Low resolution Large depth of field High resolution