Transcript Slide 1
Accommodation measurement in eyes implanted with Tek-Clear Accommodative Intraocular Lens
Studený P. , Dědková B., Farkaš A., Vokrojová M., Siveková D., Břešťák M.
Oční klinika FNKV a 3 LF UK Praha Oční oddělení , Karlovarská krajská nemocnice a.s.
Centrum Fetální medicíny Praha Tek-Clear Accommodative IOL is manufactured by: TEKIA, INC.
www.tekia.com
Ways to restore accommodative ability after cataract surgery
Predictability: intermediate vision Disphotopic phenomena Contrast sensitivity far vision near vision
Multifocal IOL Refractive/diffractive
+ + + + Stability of refraction +
Accommodating IOL
+ + + 0 + -
Tek-Clear Accommodative IOL
Disc-shaped; excellent stability and centration in capsular bag 5.5 mm optic, 10.4 or 10.2 mm haptic Biconvex shape Hydrophilic acrylic A-constant 118.0
Thin junctions between optic and haptic part ; lens movement optimization during accommodation (pseudophakic accommodation) Square-edge design to minimize PCO
Patients, materials and methods
Number of patients: n=18 (32 eyes) Average age of patients: 60.6 (28-81) Follow up time: 3 months at least Preoperative UCVA: 0.28 (0.05- 0.7) Preoperative BCVA: 0.65 (0.05-1.0) Preoperative refraction (SE): 2.125 (0-6) Cataract Examination objectives: UCVA, BCVA, accommodative amplitude
Patients, materials and methods
1 phaco surgeon Standard cataract surgery, CCI 2.2mm, incision u.12, Infinity vision system, bimanual I/A, PCIOL implantation without OVD
IOL movement in eye
Ultrasound system Vivid7Pro, Probe L12M
Ultrasound examination
Eye accommodation examination
Push-up test Defocus Plusoptix
Push-up test – results
Subjective method Optimal correction for long distance Text is moved towards the patient – he/she notifies when the text becomes blurred
Push-up test – results
After 7 days After 1 month After 3 months 3.92 D (1-5) 3.52 D (1-5) 4.17 D (1-5)
Push-up test – results
16 14 12 10 2 0 8 6 4 1D 1-2D 2-3D above 3D After 7 days After 1 month After 3 months
Defocus method
Subjective method Optimal correction for long distance Patient watches optotypes placed 5m in front of him Minus lens correction is progressively increased until the image become blurred
Defocus method – results
After 7 days – 0.79D (- 0.5D -1.25D) After 1 month – 0.85D ( -0.5D – 1.25D) After 3 months – 0.87D ( -0.5D – 1.5D)
Plusoptix
Objective method Binocular auto-refractometer is placed 1 meter in front of the patient Fixation target is moved towards the patient Changes in eye refraction are recorded – refraction difference between short and long distance
Plusoptix – results
After 7 days – 1.92D ( 0.88 – 3.04) After 1 month – 2.16D ( 0.89 – 4.18) After 3 months – 2.2D (1.0 – 3.6)
Plusoptix – results
16 14 12 10 8 6 4 2 0 below 1D 1-2D 2-3D above 3D After 7 days After 1 month After 3 months
Summary
Excellent accommodation ability of Tek-clear IOL was demonstrated by means of subjective and objective examination of accommodation amplitude at 3 months follow up.
Anterior optic vault during accommodation was captured using high-frequency ultrasound – this effect explains increase in optical power of lens
Summary
Longer follow up time is needed More patients
Thank you for your attention