Aqua Glo Inspections

Download Report

Transcript Aqua Glo Inspections

Aqua Glo
Inspections
Presented by
Howard Gammon
Gammon Technical Products Inc.
Aqua Glo Inspections
There are two
standards required for
every Aqua Glo
instrument.
Fluorescing Standard
Calibrating Standard
Aqua Glo Inspections
Remove the Fluorescing Standard held by 4
screws under the hinged flap.
Aqua Glo Inspections
Read the Set Point
on the back of the
Fluorescing
Standard
Aqua Glo Inspections
Set the adjusting
lever to be
perpendicular to
the side of the
comparator
chamber.
Aqua Glo Inspections
Read the position of the
lever on the scale.
If the lever reading is the
same as the Fluorescing
Standard set point, this
confirms that the correct
Standard is being used
and that nobody has
switched with another
instrument.
Aqua Glo Inspections
Compare the Code Number
on the Fluorescing Standard
with the code on the
Calibrating Standard. They
must be the same. However,
the Set Points are usually
different. This is acceptable.
Aqua Glo Inspections
Newer Aqua Glo’s
have the code
number on a decal
on the hinged cover.
If yours does not
have a decal, we
recommend that you
make one or scribe
the code into the
metal.
Aqua Glo Inspections
If the lever reading is
different than the
Fluorescing Standard Set
Point, remove the Calibrated
Scale and install a new one,
positioned correctly.
Order P/N GTP-834-153 and
you will receive a packet of
3, in case you do not install
correctly.
Aqua Glo Inspections
If the code on the Calibrating Standard is
different than the code on the Fluorescing
Standard, order a new set of standards,
GTP-835.
A new calibrated scale will be included.
Aqua Glo Inspections
Somebody is sure to ask why the
Standards are not all the same.
Answer: Each one is manually painted
in ultraviolet light and we have never
mastered a technique to make them the
same. There are too many variables.
Return-Repair Procedure
Call 1-800-524-0287 or 732-223-4600.
Say that you need an RFR number so
that you can return an Aqua Glo for
repair.
State your name, telephone number,
unit number, and location unless it is
classified.
We can ship to an APO address.
Return-Repair Procedure
Package the instrument carefully,
preferably in it’s original case, and mail
it to us with the RFR number on the
OUTSIDE.
Gammon Technical Products
2300 Hwy 34 (P.O. Box 400)
Manasquan, NJ 08736
Return-repair Procedure
Be sure to return the Calibrating and
Fluorescing Standards and the
instrument pack that mounts on the
side of the chamber.
We will telephone you within 1 or 2
days, reporting the damage and the
cost of repairs.
Return-Repair Procedure
If you elect to have us make the repair,
give us an IMPAC card number. If you
do not want the repair made, we will
return it to you at our expense.
Repairs are normally completed in a
day or two.
Other Services
We have an ongoing program
to service the:
U.S. Army
Aviation Fuel Contamination
Test Kit: Petroleum
NSN: 6630-01-008-5524
Other Services
You can ship the entire box to us. We
will quote the cost to repair or replace
all missing items.
Alternatively, rather than shipping the
box to us, we will quote the cost of any
component in response to your inquiry.