PowerPoint-Präsentation - PULSION Medical Systems SE

Download Report

Transcript PowerPoint-Präsentation - PULSION Medical Systems SE

PULSION Medical Systems SE
PulsioFlex Software V4.0 – New Features
Data Visualization – SpiderVision
SpiderVision
• SpiderVision screen needs to be activated in the layout settings
• Shows dynamic changes of continuous parameters graphically
• Screen display and settings similar to PiCCO2
2
Data Visualization – SpiderVision
•



SpiderVision configuration
To be opened by touching the Spider diagram
Basic: predefined basic settings
Individual: individual configuration (number of arms and parameters)
3
Data Visualization – Profile
Profile
• Profile screen needs to be activated in the layout settings
• Static display of the values during the last thermodilution measurement
• Screen display and configuration similar to PiCCO2
4
Data Visualization – Profile
• Profile configuration
 To be opened by toucing the Profile diagram
 Left hand side buttons enables switching between the predefined
parameter configurations
 Individual configuration is not possible
5
PiCCO Thermodilution Measurement
Values with insufficient pressure curve
• In case a calibration cannot be performed due to an insufficient pressure
signal, a related message appears on the screen
• After confirmation the discontinuous thermodilution results are available
6
DO2 Calculation
•
•
•
Calculation of oxygen delivery (DO2) is now also available without CeVOX
module
Input screen is opened by touching the DO2 button
in the main menu
After entering the SaO2 and Hb values DO2 is calculated and displayed
continuously
7
Data Acquisition – Start
Trend Data Recording
• Activation in the Main Menu
• Pushing the Start button starts data recording
• All parameter values are stored every 12 seconds for later download
• File name is created automatically and indicated on the screen
8
Data Acquisition – Event Input
•
•
•
Events can always be put in by touching the input field next to ‘Add Event’
(max. 25 characters per event)
All entered events will be stored in the event list
Event inputs while data recording is running will also be stored in the
export data file
9
Data Acquisition – Stop & Data Export
•
•
•
•
•
Stop button stops the recording and the save file is closed
Transferable files are listed
Plug in USB storage device to one of the USB ports and push Export Data
Observe screen messages on USB device detection and finishing of data
export
File format is ‘.csv’ (comma separated values) which can be opened by
spread sheet programs (e.g. MS Excel)
10
Thank you very much
For your attention
11