Transcript Document
Auto Mode Switch Auto Mode Switch • Changes to a non-tracking mode during high atrial rates • Programmable Atrial Tachycardia Detection Rate • Programmable post-ventricular atrial blanking (PVAB) minimizes inappropriate mode switching due to far-field R-waves Auto Mode Switch • Adapt Filtered Atrial Rate Interval (FARI) based on P-P intervals • The FARI decreases by 38 ms when P-P interval is shorter than FARI • The FARI increases by 25 ms when P-P interval is longer than FARI • FARI < ATDR interval: Entry Mode Switch • FARI > Maximum Tracking Rate (sensor indicated rate interval): Exit Mode Switch Rate (bpm) Interval (ms) 265 Atrium 278 200 ATDR 180 316 FARI < ATDR Interval: mode switch entry 264 277 303 290 289 315 302 300 314 333 354 392 430 468 400 506 150 544 PP < FARI: -38 ms 582 620 PP > FARI: +25 ms 658 696 734 100 600 772 810 848 886 time to mode switch mode switch entry 924 962 1000 50 0 1 1200 2 3 4 5 Time (sec) 6 7 8 9 10 Rate (bpm) Interval (ms) PP < FARI: -38 ms 300 200 300 325 350 180 PP > FARI: +25 ms 375 ATDR 333 400 425 150 400 450 475 MTR 120 FARI > MTR Interval: 500 500 mode switch exit 525 600 100 550 time to mode switch mode switch exit Atrium 1200 50 0 1 2 3 4 5 Time (sec) 6 7 8 9 10 Auto Mode Switch Filtered Atrial Rate Interval (FARI) Atrial cycle lengths DDD 670ms 650ms 630ms 590ms 0ms -38ms -38ms -38ms 670ms 670ms 632ms ATDR: 180bpm 300ms 300ms 300ms 300ms 300ms 350ms 300ms 300ms 300ms 300ms 594ms 333ms -38ms -38ms +25ms -38ms -38ms -38ms -38ms -38ms -38ms 518ms 556ms FARI DDI 442ms 480ms 366ms 404ms 328ms 290ms 315ms 600ms 600ms +25ms +25ms 277ms 302ms 327ms +25ms Auto Mode Switch Base Rate Auto Mode Switch (AMS ) Base Rate Can Be Programmed To: • Base Rate + 0 ppm. to Base Rate + 35 ppm. • (In Steps of 5 ppm.) Auto Mode Switch Base Rate Auto Mode Switch - AMS Base Rate