Die Spitzen des SONICflex bone

Download Report

Transcript Die Spitzen des SONICflex bone

SONICflex bone – Indications and settings
KaVo SONICflex bone tips
in use with INTRAsurg 1000 Air
Indications and settings
Seite 1
16.07.2015 - © KaVo Dental GmbH
SONICflex bone – Indications and settings
Comparison of cutting power
(standardized oscillating cuts made under 50g
working pressure)
From left to right:
Round rotary drill
SONICflex bone sagittal saw
Competitor 1/oscillation
Competitor 2/oscillation
Seite 2
16.07.2015 - © KaVo Dental GmbH
SONICflex bone – Indications and settings
Important recommendations for use
•
Make sure of incoming air pressure level between 4 - 7 bar (technical
adjustement), otherwise required power will not be reached.
•
Don‘t press SONICflex bone tips against bone – minimization of
power. Correct working pressure can be compared with writing by
sharpened pencil.
•
Choose the right power setting on regulation ring of SONICflex 2003 L
according to user instructions of each single tip (see instructional
manual).
Seite 3
16.07.2015 - © KaVo Dental GmbH
SONICflex bone – Indications and settings
Tip no. 80, square
• Sonicflex 2003L setting:
Level 3
• Indications:
- Dividing Sharpey`s fibres
- Extension of alveolar bone for:
- Removal of root sections
- Gentle extraction prior to
implantation
- Separation of multi-rooted
teeth
- Bonesplitting
Dr. F. Heinemann, Morsbach
Seite 4
16.07.2015 - © KaVo Dental GmbH
SONICflex bone – Indications and settings
Tip no. 81, diamond-coated sphere
• SONICflex 2003L setting:
Level 3
• Indications:
- Bone window preparation for
sinuslift
-
In case of massive bone structure cortical
bone could be cut by rotary method and
only remaining bone structure is being
removed by oscillating method.
Dr. Dr. S.-H.Gnoth, Köln
Seite 5
16.07.2015 - © KaVo Dental GmbH
SONICflex bone – Indications and settings
Tip no. 82, non-diamond-coated sphere
• SONICflex 2003L setting:
Level 2 or 3
• Indication:
- Separation of Schneider‘s
membrane during sinuslift
Dr. Dr. S.-H.Gnoth, Köln
Seite 6
16.07.2015 - © KaVo Dental GmbH
SONICflex bone – Indications and settings
Oscillating saws
Tips no. 83 (axial)/84 (sagittal)
SONICflex 2003L settings:
Level 3
•Indications:
•Removal of bone blocks, Osteotomy
(Distraction, Bonesplitting)
84
sagittal
In case of massive bone structure cortical bone could be cut by rotary
method and only remaining bone structure is being removed by oscillating
method
83
axial(quer)
Dr. E.Guez, Paris
Dr. E. Guez, Paris
Seite 7
16.07.2015 - © KaVo Dental GmbH