MEASUREMENTS OF HEAVY METALS IN FLUIDS
WITH VOLTAMMETRIC METHODS
Øyvind Mikkelsen and Knut H. Schrøder
Norwegian University of Science and technology
Department of Chemistry
N - 7491 TRONDHEIM, NORWAY
Mikkelsen 2000
Voltammetry
• In voltammetry the
information about
the analyte is
achieved by measure
the current as a
function of a applied
potential.
• The current is
proportional to the
concentration.
Mikkelsen 2000
Cross section of the dental amalgam (or somewhat related material)
electrode.
Mikkelsen 2000
Working electrode
• Metal ions are
deposited on the
electrode surface.
Me
Men+
Mikkelsen 2000
Voltammetric cell
+
| |
(a)
(R)
/\/\/\/\/\/\/\/\/\/\/\/\
(d)
•
Counter
electrode
•
Working
electrode
•
Reference
electrode
(c)

(b)
A

Section of elctrode system.

Counter electrode
V
Working electrode
Reference electrode
Mikkelsen 2000
Mikkelsen 2000
Antimony (Sb) in water samples and biological matrices
Bismuth (Bi) in water samples and biological matrices
Cadmium (Cd) in water samples and effluents
Chromium (Cr) in environmental samples and biological matrices
Cobalt (Co) in water samples and effluents
Copper (Cu) in environmental and water samples
Iron (Fe) in water samples
Lead (Pb) in water samples and effluents
Mercury (Hg) in water samples and biological matrices
Molybdenum (Mo) in water samples and biological matrices
Nickel (Ni) in water samples and biological matrices
Platinum (Pt) in environmental samples and biological matrices
Thallium (Tl) in water samples and effluents
Tungsten (W) in surface water samples
Uranium (U) in surface water samples
500 ppt
500 ppt
50 ppt
25 ppt
50 ppt
50 ppt
200 ppt
50 ppt
100 ppt
50 ppt
50 ppt
0.1 ppt
50 ppt
200 ppt
25 ppt
* 1 ppt = 1 part per trillion = 1 ng/kg
KILDE: http://metrohm.ch/products/va.html
Mikkelsen 2000
Signal Source & Measuring Unit (SMU)
- Control UNIT, Microcontroller unit that controls the card in
the SMU (voltammetric analyzer). Communicates with a
external PC via a RS232 interface.
- Signal Source UNIT, this unit consists of two electronic
boards; Signal Source ControlER Board and the
D/A Converting Unit. Digital numeric values of the voltage
signal to the cell are generated by the microcontroller and
placed in this unit. When a scan is started the values is cloked
out from the D/A converter as analog voltage steps.
- Converter/Measure UNIT, consists of one electronic board ref.
as the A/D converting unit. This board receives the current
feedback from the Ampl. Int. UNIT as a voltage signal from a
I/E converter, and converts the signal from the cell to digital
words, which the processor can read and place into a buffer for
transmission to the PC.
Mikkelsen 2000
Ampl. Int. UNIT
Amplifier / Interface UNIT for both the applied and the
received signal, to/from the spesific cell in use. Takes the
asserted potential from the SMU and puts it to the cell.
It also converts the current from the cell to a voltage wich
can be converted to digital values in the voltammetric
analyzer and send it to a PC.
Mikkelsen 2000
Signal Source & Measuring Unit
Control
UNIT
Signal
Source
UNIT
Convert/
Measure
UNIT
Sync
Ampl.
Int.
UNIT
Recording, analysing,
presentation and
documentation
UNIT (PC)
Ext. rec.
connect.
Ext.
trig.
Cntrl.
Cell
UNIT
Mikkelsen 2000
Voltammetric equipment
Mikkelsen 2000
Operational amplifier potentiostat
(OP AMP)
Linear-scan
voltage
generator
Signal source

A
+
Counter electrode
B

+
Referance electrode
Cell I
Potensiostatic control circuit
Working electrode

Eout
C
Recorder
+
Current-to-voltage converter
Mikkelsen 2000
Some typical signalmodulations in
voltammetry
Linear scan
Differential pulse
scan
Cyclic scan
Square wave
Mikkelsen 2000
Typical controlled parameters in voltammetri
C
B
D
A
A = 520 ms, measurement after 480 ms
B = 70 mV
C = 90 ms, measurement after 70 ms
D = 65 mV
Increase in staircase step; 5 mV.
Mikkelsen 2000
Apparatus for voltammetric analyses
Mikkelsen 2000
Software for setting/loading parameters to control unit
LabVIEW
Mikkelsen 2000
Background of dental amalgam electrode in acetatbuffer pH=4,6.
Scan from -1.2 to -0.2 V
Mikkelsen 2000
Zinc, cadmium and lead (25, 50 and 75 ppb) detected by DPASV on
dental amalgam electrode. All scan in NH4Ac (non-buffered) pH=6,4
also containing nickel and copper and gallium (75 ppb). Deposition
time 60 s at -1.2V. Scan from -1.2 to -0.4V, pulse height
50 mV.
Mikkelsen 2000
Plot of a typical scan at silver electrode before and after lead addition.
Scan performed by DPASV using a sine oscillating working silver
electrode. Deposition time 60 s, scan rate 15mV s-1, pulse height 40
mV.
Mikkelsen 2000
Soilwatch Project