Transcript PMQ3.ppt
PMQ3 Magnetic Measurements BL vs X fit • BL vs X runs 3, 4 & 5 • Mean Integrated Gradient = 8.1862 +/- 0.00624 T Temperature = 21.7 +/- 0.32 degree C BL vs. X plots Old Data using 3458A • Measured integrated gradient using Metrolab instead of HP 3458A because there were too few points for the 3458A measurement. The deadtime of the 3458A measurement was causing a significant under reporting of the voltage for PMQ3 (~27%). • BL vs X run 2, which uses the 3458A has a Integrated Gradient = 6.7984 +/- 0.016398 T.