Transcript ppt

GLAST LAT Project
Gamma-ray Large
Area Space
Telescope
Instrument Analysis meeting, July 28, 2006
Calibration of high energy diode.
Alexandre Chekhtman
NRL/GMU
A.Chekhtman
1
GLAST LAT Project
Instrument Analysis meeting, July 28, 2006
Problem with energy reconstruction above
1 GeV/crystal
• No reconstructed energy
depositions above 1 GeV
for data collected with
flight gain
A.Chekhtman
2
GLAST LAT Project
Instrument Analysis meeting, July 28, 2006
Calibration process for high energy diode
•
•
When high energy diode is calibrated with muons, preamplifier gain is
set to the value 10 times bigger, than normal (“muon” gain)
Calibration includes two components:
– adc2dac (integral nonlinearity) – relationship between measured signal (in
ADC units) and the charge injected to the preamplifier input
• Should take care of different preamplifier gain
– MeVPerDAC – relationship between deposited energy in a crystal and the
charge (in DAC units) injected to the preamplifier input
•
•
• Should NOT depend on preamplifier gain
In this simplified model to take into account the different preamplifier
gain setting we just have to use the appropriate adc2dac calibration
(for flight gain or for muon gain)
Real situation is more confusing: we also can change the charge
corresponding to one DAC unit by switching charge injection capacitor
(controlled by CALIBGAIN bit in the configuration word)
–
We use CALIBGAIN=OFF setting when doing charge injection calibration with muon
gain, otherwise the step of charge injection calibration would be too big, compare to
muon signal
•
•
•
A.Chekhtman
This setting decrease the value of one DAC unit by factor ~9.3 (CALIBGAIN factor).
to take this into account, we have to multiply MeVPerDAC values by this CALIBGAIN factor, when
reconstructing data collected with flight gain
CALIBGAIN factor could be defined individually for each channel, by comparing charge injection
calibrations done with CALIBGAIN=ON and CALIBGAIN=OFF
3
GLAST LAT Project
Instrument Analysis meeting, July 28, 2006
Modifications to calibGenCAL
•
•
•
•
•
•
ciFit will be able to process data for low energy diodes and high energy
diodes separately
ciFit for high energy diodes will be run 3 times with different runs
from calibGen suite, producing 3 intnonlin files:
1.
2.
3.
Muon gain, CALIBGAIN=OFF (calibGen run 204)
Flight gain, CALIBGAIN=ON (calibGen run 102)
Flight gain, CALIBGAIN=OFF (calibGen run 104)
–
–
compare intnonlin files from 2 and 3 and will generate the file with
CALIBGAIN factors for all channels.
Generate MeVperDAC and Asymmetry files for flight gain
–
–
For muon gain: intnonlin file 1 and MeVperDAC/Asymmetry for muon gain
For flight gain: intnonlin file 2 and MeVperDAC/Asymmetry for flight gain
Intnonlin file 1 will be used in muon calibration procedure when
producing MeVperDAC and Asymmetry files for muon gain
New python script will
We have to store in the calibration database two sets of calibration
files for each time period (with different “flavor”):
When running reconstruction we should select appropriate calibration
flavor, depending on configuration used for data collection.
A.Chekhtman
4
GLAST LAT Project
Instrument Analysis meeting, July 28, 2006
Status
• Zach Fewtrell started to implement the proposed
modifications beginning of this week
• He hopes to get them ready today or on Monday
• He’ll produce the new set of calibration files (for
flight gain) for CU beginning of next week.
A.Chekhtman
5