Direct detection of C 2 H 2 in air and human breath by cw-CRDS Florian Schmidt, Olavi Vaittinen, Markus Metsälä and Lauri Halonen Laboratory of Physical Chemistry, University of Helsinki, Finland

Cavity ring-down spectroscopy (CRDS)

Laser Scan

 int 

Sn

c

 1 0  

Detector

  0    Absorption enhancement due to cavity: 10 3 to 10 5 CRDS independent of laser power fluctuations No calibration needed June 21, 2010 2

Experimental CRDS setup - schematic

Tuning range

: 1520-1575 nm (6350-6575 cm -1 )

Cavity Finesse

: 390 000

Sensitivity

: 7 ×10 -11 cm -1

Acquisition time

: 1 s - 20 min

Detected at ppt level

: C 2 H 2 , NH 3 , HCN

Other detectable species

: CH 4 , N 2 O, CO, CO 2 , H 2 O ECDL = external cavity diode laser OI = optical isolator AOM = acousto-optic modulator MFC = mass flow controller 3 June 21, 2010

Experimental CRDS setup June 21, 2010 4

Motivation to measure acetylene    One of the most common hydrocarbons in the troposphere Mostly anthropogenic sources (fossil- and biofuels, biomass burning) Correlates with other combustions gases and VOCs   Trace pollution and exposure to combustion gases (exhausts, smoking) Baseline of breath C 2 H 2 in the healthy population? Is C 2 H 2 a biomarker for a medical condition (e.g.

Proteus mirabilis

bacterial infection)?

    Not previously detected in ambient air with high time resolution and without pre-concentration.

Not previously quantified in breath GC-MS slow and expensive Not measurable with PTR-MS and SIFT-MS with standard pre-cursors June 21, 2010 5

Spectra from breath and ambient air Air acetylene concentration 0.93 ppbv; Scan time 20 min; Detection limit 120 ppt June 21, 2010 6

System response to C 2 H 2 absorption June 21, 2010 7

Outdoor air – continuous flow

F. M. Schmidt, O. Vaittinen, M. Metsälä, P. Kraus, L. Halonen, Appl. Phys. B (2010) June 21, 2010 8

Day – Night comparison

F. M. Schmidt, O. Vaittinen, M. Metsälä, P. Kraus, L. Halonen, Appl. Phys. B (2010) June 21, 2010 9

Indoor – outdoor air comparison F. M. Schmidt, O. Vaittinen, M. Metsälä, P. Kraus, L. Halonen, Appl. Phys. B (2010) June 21, 2010 10

Breath collection and sample handling  No adsorption/desorption phenomena in the bags, the gas system tubing and the quartz-coated ring-down cavity  Acetylene can be collected and stored in aluminum breath bags (statistical error for 10 bags: 0.4 ppb)  Acetylene concentration stays within 0.4 ppb for 5 days in the bag  No quantitative difference between alveolar/full breath samples and nose/mouth breathing (non-smokers)  Thus: static measurements were possible  Spectrometer temperature: 22 degrees (room temperature)  Sample pressure in the cavity: 0.1 bar June 21, 2010 11

Study - Background level of C 2 H 2 in breath  40 healthy volunteers 22 male, 18 female, 20 to 63 years old, 31 non smokers, 9 smokers  Subjects requested to have stayed at sampling location for 30 min prior to breath collection  One full breath sample collected to a breath bag  At the same time: one ambient air sample collected to a second bag  Samples were analyzed on the same day or at least within 48 hours.

June 21, 2010 12

Results – Breath acetylene minus air acetylene Breath C 2 H 2 baseline = Ambient air C 2 H 2 level Smokers could be identified with a sensitivity and specifity of 100 % June 21, 2010 13

Timer after smoking June 21, 2010 14

Exposure studies  Subject smokes a cigarette or takes a puff  Continuous breathing into CRDS system for up to 30 min  C 2 H 2 decay recorded standing on peak of absorption profile (3 points per second)  For long-term decay: bag samples or shorter intervals of continuous breathing  Fit exponential functions to decay  Simple 3-compartment model (respiratory tract, blood, tissue) June 21, 2010 15

Breath cycles June 21, 2010 16

Acetylene residence times June 21, 2010 17

Summary   CRDS suitable for fast and direct detection of C 2 H 2 in air and breath Mean C 2 H 2 mixing ratio in Helsinki outdoor air of ~1 ppbv  C 2 H 2 correlates with CO, human activity and temperature  Strong fluctuations (up to 60 ppbv on a minute time scale) in urban outdoor air during daytime. Single bag measurements unreliable.

  Breath C 2 H 2 baseline is equal to ambient air C 2 H 2 level Smokers were identified with 100 % sensitivity and specifity  Breath acetylene returns to baseline level around 3 hours after exposure. Urban air pollution could impair smoking status assessment  High-time resolution decay measurements could improve knowledge about residence times of gases and related physiological processes June 21, 2010 18