IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 15.1: Optimization of protection in radiography Practical.
Download ReportTranscript IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 15.1: Optimization of protection in radiography Practical.
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 15.1: Optimization of protection in radiography Practical exercise IAEA International Atomic Energy Agency Overview / Objectives • Subject matter : quality control of general radiography system • Step by step procedure to be followed to implement QC test: • kVp Accuracy and Reproducibility • Interpretation of results IAEA 15.1: Optimization of protection in radiography 2 IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology Part 15.1 : Optimization of protection in radiography kVp measurement (Accuracy and Reproducibility) IAEA International Atomic Energy Agency kVp measurement Purpose : • Image quality and patient dose are dependent on any variation in the generator kilovoltage (kV) of the x-ray set. Therefore an accurate kV calibration is required. • A non-invasive tube voltage check over the range of kV used clinically should be performed Equipment : • An electronic device (like multi-function meter) to measure : kVp, time, dose. IAEA 15.1: Optimization of protection in radiography 4 Multi-Function Meters IAEA 15.1: Optimization of protection in radiography 5 kVp Accuracy • Before making measurements, one must understand what “kVp” means • There are 3 definitions of kVp, only one is commonly used • CAUTION : some mobile x-ray units use a capacitor, and the kVp decreases during the exposure IAEA 15.1: Optimization of protection in radiography 6 kVp Definitions IAEA 15.1: Optimization of protection in radiography 7 kVp Accuracy Method : • set meter at about 100 cm from focal spot or as prescribed by the manufacturer • set ~ 20 mAs • measure kVp from about 60 - 120 kVp (e.g., 60, 70, 80, 100, and 120) • calculate error between selected and measured value IAEA 15.1: Optimization of protection in radiography 8 kVp Accuracy Analysis : • The difference between selected and measure kVp should be within 5% Frequency : • acceptance • tube change, generator maintenance • annual IAEA 15.1: Optimization of protection in radiography 9 kVp Reproducibility • This test checks that the kVp will stay the same from exposure to exposure Method : • same setup, but at fixed 70 kVp • make 5 exposures about 30 seconds apart (be careful of tube load!) IAEA 15.1: Optimization of protection in radiography 10 kVp Reproducibility Analysis : • calculate coefficient of variation (CV) • CV = (standard deviation/mean) of the 5 readings • CV must be less than 2% Frequency : • as for accuracy IAEA 15.1: Optimization of protection in radiography 11 kVp Reproducibility Further test : • older x-ray units may have a mechanical selector for kVp • test this by moving the kVp dial up, down, then back to 70 before each exposure • a variation of 3 kVp is permitted IAEA 15.1: Optimization of protection in radiography 12