New Modes of Mechanical Ventilation Mazen Kherallah, MD, FCCP Consultant Intensivist King Faisal Specialist Hospital.

Download Report

Transcript New Modes of Mechanical Ventilation Mazen Kherallah, MD, FCCP Consultant Intensivist King Faisal Specialist Hospital.

New Modes of Mechanical Ventilation
Mazen Kherallah, MD, FCCP
Consultant Intensivist
King Faisal Specialist Hospital
Key Ideas for Understanding
Mechanical Ventilation
1. Mathematical Models
–
–
–
Equation of motion
Time constant
Mean airway pressure
2. Control Variables
–
Pressure, volume, dual
3. Phase Variables
–
Trigger, Limit and Cycle
4. Breath Types/Patterns
5. Optimum Mode Selection
Lung
Mechanics
resistance = Dpressure / Dflow
flow
transairway
pressure
transrespiratory
pressure
volume
transthoracic
pressure
elastance = Dpressure / Dvolume
Equation of Motion
ventilation
pressure =
(to deliver
tidal volume)
resistive
pressure
(to make air flow
through the
airways)
+
elastic
pressure
(to inflate lungs
and chest wall)
P = Presistive + Pelastance
P= RxV + ExV
Phase Variables
• Trigger (start)- begins inspiratory flow
• Cycling (end)- ends inspiratory flow
• Limiting (continue)- places a maximum value on a “control
variable”
– pressure
– volume
– flow
– time
Trigger VariableStart of a Breath
•
•
•
•
•
Time - control ventilation
Pressure - patient assisted
Flow - patient assisted
Volume - patient assisted
Manual - operator control
Inspiratory - delivery limits
• Maximum value that can be reached but
will not end the breath– Volume
– Flow
– Pressure
End of Insp…cycle mechanisms
• The phase variable used to terminate
inspiration– Volume
– Pressure
– Flow
– Time
Breath Type… Only Two (for now)!
• Mandatory
– Ventilator does the work
– Ventilator controls start and stop
• Spontaneous
– Patient takes on work
– Patient controls start and stop
The Control VariableInspiratory Breath Delivery
• Flow (volume) controlled
– pressure may vary
• Pressure controlled
– flow and volume may vary
• Time controlled (HFOV)
– pressure, flow, volume may vary
Volume/Flow Control
Inspiration
Pressure Control
Expiration
20
Inspiration
Expiration
20
Paw
Pressure
Paw
0
1
0
2
20
0
1
2
0
1
2
20
Volume
0
0
1
0
2
3
Flow
0
-3
3
Time (s)
0
-3
Time (s)
Volume Control Breath Types
60
Paw
SEC
cmH20
-20
120
1
2
3
4
5
6
INSP
SEC
Flow
L/min
1
2
3
4
5
120
6
EXH
If compliance decreases the pressure increases to
maintain the same Vt
New Modes of Ventilation
Dual-Controlled Modes
Type
Manufacturer; ventilator
Name
Dual control within a breath
VIASYS Healthcare; Bird
8400Sti and Tbird
VIASYS Healthcare; Bear 1000
Volume-assured pressure
support
Pressure augmentation
Dual control breath to breath:
Pressure-limited flow-cycled
ventilation
Siemens; servo 300
Cardiopulmonary corporation;
Venturi
Volume support
Variable pressure support
Dual control breath to breath:
Pressure-limited time-cycled
ventilation
Siemens; servo 300
Pressure-regulated volume
control
Adaptive pressure ventilation
Autoflow
Variable pressure control
Dual control breath to breath:
SIMV
Hamilton; Galileo
Hamilton; Galileo
Drager; Evita 4
Cardiopulmonary corporation;
Venturi
Adaptive support ventilation
Dual Control within a Breath
volume-assured pressure support
• This mode allows a feedback loop based on the volume
• Switches even within a single breath from pressure
control to volume control if minimum tidal volume has
not been achieved
Bear 1000
Bird 8400Sti
Tbird
Dual Control within a Breath
volume-assured pressure support
• The Respiratory Therapist sets :
–
–
–
–
–
–
–
pressure limit = plateau seen during VC
respiratory rate
peak flow rate (the flow if TV < target)
PEEP
FiO2
trigger sensitivity
minimum tidal volume
Pressure limit
overridden
40
Set pressure limit
Paw
cmH20
-20
0.6
Volume
L
Set tidal volume cycle threshold
Tidal volume
Tidal volume
not met
met
0
60
Inspiratory flow
greater than set flow
Flow cycle
Set flow limit
Inspiratory flow
equals set flow
Flow
L/min
60
Switch from Pressure control to
Volume/flow control
trigger
Pressure at
Pressure support
no
flow=
25% peak
yes
yes
delivered VT
≥ set VT
no
Cycle off
inspiration
Insp flow
> Set flow
yes
no
Switch to flow control
at peak flow setting
no
yes
delivered VT
= set VT
no
PAW <PSV
setting
yes
Control logic for volume-assured pressure-support mode
Dual Control within a Breath
volume-assured pressure support
• If pressure too high, all breaths are pressure-limited.
• If the peak flow setting is too high , all breaths will
be volume-controlled
• If the pressure is set too high or the minimum tidal
volume is set too low; the volume guarantee is
negated
• If peak flow set too low, the switch from pressure to
volume is late in the breath, inspiratory time is too
long.
Dual Control within a Breath
volume-assured pressure support
• Amato et al Chest 1992;102: 1225-1234
• Compared VAPS to simple AC volume
– Lower WOB
– Lower Raw
– Less PEEPi
Dual Control Breath-to-Breath
pressure-limited flow-cycled ventilation
Volume Support
• Tidal volume is used as feedback control to
adjust the pressure support level
• All breaths are patient triggered, pressure
limited, and flow-cycled.
• Automatic weaning of pressure support as long
as tidal volume matches minimum required VT
(VT set in a feedback loop to adjust pressure).
Dual Control Breath-to-Breath
pressure-limited flow-cycled ventilation
Volume Support
Servo 300
Maquet Servo-i
VS vs VAPS
• How does volume support differ from
VAPS ?
– In volume support, we are trying to adjust
pressure so that, within a few breaths, desired
VT is reached.
– In VAPS, we are aiming for desired VT tacked
on to the end of a breath if a pressure-limited
breath is going to fail to achieve VT
VS (Volume Support)
• Entirely a spontaneous mode
• Delivers a patient triggered (pressure or flow), pressure
targeted, flow cycled breath
– Can also be timed cycled (if TI is extended for some reason) or
pressure cycled (if pressure rises too high).
• Similar to pressure support except VS also targets set VT.
It adjusts pressure (up or down) to achieve the set volume
(the maximum pressure change is < 3 cm H2O and ranges
from 0 cm H2O to 5 cm H2O below the high pressure
alarm setting
• Used for patients ready to be “weaned” from the
ventilator and for patients who cannot do all the WOB
but who are breathing spontaneously
VS (Volume Support)
(1), VS test breath (5 cm H2O); (2), pressure is increased slowly until target volume is achieved; (3),
maximum available pressure is 5 cm H2O below upper pressure limit; (4), VT higher than set VT
delivered results in lower pressure; (5), patient can trigger breath; (6) if apnea alarm is detected,
ventilator switches to PRVC
yes
Calculate new
Pressure limit
trigger
no
Pressure limit
Based on VT/C
Volume from
Ventilator=
Set tidal volume
calculate
compliance
Flow= 5% of
Peak flow
yes
cycle off
no
Control logic for volume support mode of the servo 300
Dual Control Breath-to-Breath
pressure-limited flow-cycled ventilation
Volume Support
• Little data to show it actually works.
• If pressure support level increases to maintain
TV in pt with increased airways resistance,
PEEPi may increase.
• If minimum TV set too high, weaning may be
delayed.
VS (Volume Support)
• Indications
• Spontaneous breathing patient who require
minimum E
• Patients who have inspiratory effort who
need adaptive support
• Patients who are asynchronous with the
ventilator
• Used for patient who are ready to wean
VS (Volume Support)
• Advantages
• Guaranteed VT and E
• Pressure supported breaths using the lowest
required pressure
• Decreases the patient’s spontaneous respiratory rate
• Decreases patient WOB
• Allows patient control of I:E time
• Breath by breath analysis
• Variable I to meet the patient’s demand
VS (Volume Support)
• Disadvantages
• Spontaneous ventilation required
• VT selected may be too large or small for
patient
• Varying mean airway pressure
• Auto-PEEP may affect proper functioning
• A sudden increase in respiratory rate and
demand may result in a decrease in ventilator
support
Dual Control Breath-to-Breath
pressure-limited time-cycled ventilation
Pressure Regulated Volume Control
Servo 300
Maquet Servo-i
Dual Control Breath-to-Breath
pressure-limited time-cycled ventilation
Pressure Regulated Volume Control
• Delivers patient or timed triggered, pressure-targeted
(controlled) and time-cycled breaths
• Ventilator measures VT delivered with VT set on the
controls. If delivered VT is less or more, ventilator
increases or decreases pressure delivered until set VT
and delivered VT are equal
Dual Control Breath-to-Breath
pressure-limited time-cycled ventilation
Pressure Regulated Volume Control
• The ventilator will not allow delivered
pressure to rise higher than 5 cm H2O below
set upper pressure limit
• Example: If upper pressure limit is set to 35 cm H2O and
the ventilator requires more than 30 cm H2O to deliver a
targeted VT of 500 mL, an alarm will sound alerting the
clinician that too much pressure is being required to deliver
set volume (may be due to bronchospasm, secretions, changes
in CL, etc.)
PRVC (Pressure Regulated
Volume Control)
PRVC. (1), Test breath (5 cm H2O); (2) pressure is increased to deliver set volume; (3), maximum
available pressure; (4), breath delivered at preset E, at preset f, and during preset TI; (5), when VT
corresponds to set value, pressure remains constant; (6), if preset volume increases, pressure
decreases; the ventilator continually monitors and adapts to the patient’s needs
yes
Calculate new
Pressure limit
trigger
no
Pressure limit
Based on VT/C
Volume from
Ventilator=
Set tidal volume
calculate
compliance
time= set
Inspiratory time
yes
cycle off
no
Control logic for pressure-regulated volume control and autoflow
PRVC (Pressure Regulated Volume
Control)
• Disadvantages and Risks
• Varying mean airway pressure
• May cause or worsen auto-PEEP
• When patient demand is increased, pressure level
may diminish when support is needed
• May be tolerated poorly in awake non-sedated
patients
• A sudden increase in respiratory rate and demand
may result in a decrease in ventilator support
PRVC (Pressure Regulated Volume
Control)
• Indications
• Patient who require the lowest possible
pressure and a guaranteed consistent VT
• ALI/ARDS
• Patients requiring high and/or variable I
• Patient with the possibility of CL or Raw
changes
PRVC (Pressure Regulated Volume
Control)
• Advantages
•
•
•
•
•
•
Maintains a minimum PIP
Guaranteed VT and E
Patient has very little WOB requirement
Allows patient control of respiratory rate and
Variable E to meet patient demand
Decelerating flow waveform for improved gas
distribution
• Breath by breath analysis
E
A New Twist…
Volume Targeted
60
Paw
SEC
cmH20
-20
120
1
2
3
4
5
6
INSP
SEC
Flow
L/min
1
2
3
4
5
6
120
EXH
Many Dual Modes start out looking like PCV
Volume Targeted
(Pressure Controlled)
60
Paw
SEC
cmH20
-20
120
1
2
3
4
5
6
INSP
SEC
Flow
L/min
120
1
2
3
4
5
6
EXH
As compliance changes - flow and volumes change
New Volume Targeted Breath
Pressure Variability is Controlled
60
Paw
SEC
cmH20
-20
120
1
2
3
4
5
6
INSP
SEC
Flow
L/min
120
1
2
3
4
5
6
EXH
Pressure then raises to assure that the set tidal
volume is delivered
Dual Control Breath-to-Breath
adaptive support ventilation
ASV (Adaptive Support Ventilation)
• A dual control mode that uses pressure
ventilation (both PC and PSV) to maintain a set
minimum E (volume target) using the least
required settings for minimal WOB depending on
the patient’s condition and effort
– It automatically adapts to patient demand by
increasing or decreasing support, depending on the
patient’s elastic and resistive loads
ASV (Adaptive Support Ventilation)
•
•
•
•
•
•
The clinician enters the patient’s IBW, which allows the
ventilator’s algorithm to choose a required E. The ventilator
then delivers 100 mL/min/kg.
A series of test breaths measures the system C, resistance and
auto-PEEP
If no spontaneous effort occurs, the ventilator determines the
appropriate respiratory rate, VT, and pressure limit delivered for
the mandatory breaths
I:E ratio and TI of the mandatory breaths are continually being
“optimized” by the ventilator to prevent auto-PEEP
If the patient begins having spontaneous breaths, the number of
mandatory breaths decrease and the ventilator switches to PS at
the same pressure level
Pressure limits for both mandatory and spontaneous breaths are
always being automatically adjusted to meet the E target
ASV (Adaptive Support Ventilation)
• Indications
•
•
•
•
•
•
•
Full or partial ventilatory support
Patients requiring a lowest possible PIP and a guaranteed VT
ALI/ARDS
Patient requiring high and/or variable
Patients not breathing spontaneously and not triggering the
ventilator
Patient with the possibility of work land changes (CL and
Raw)
Facilitates weaning
ASV (Adaptive Support Ventilation)
• Advantages
–
–
–
–
–
–
–
Guaranteed VT and E
Minimal patient WOB
Ventilator adapts to the patient
Weaning is done automatically and continuously
Variable to meet patient demand
Decelerating flow waveform for improved gas
distribution
Breath by breath analysis
ASV (Adaptive Support Ventilation)
• Disadvantages and Risks
•
•
•
•
•
Inability to recognize and adjust to changes in alveolar VD
Possible respiratory muscle atrophy
Varying mean airway pressure
In patients with COPD, a longer TE may be required
A sudden increase in respiratory rate and demand may result
in a decrease in ventilator support
Automode
• The ventilator switch between mandatory and
spontaneous breathing modes
• Combines volume support (VS) and pressure-regulated
volume control (PRVC)
• If patient is paralyzed; the ventilator will provide
PRVC. All breaths are mandatory that are ventilator
triggered, pressure controlled and time cycled; the
pressure is adjusted to maintain the set tidal volume.
• If the patient breathes spontaneously for two
consecutive breaths, the ventilator switches to VS. All
breaths are patient triggered, pressure limited, and flow
cycled.
• If the patient becomes apneic for 12 seconds; the
ventilator switches back to PRVC
MMV (Mandatory Minute
Ventilation)
•
•
•
AKA: Minimum Minute Ventilation or Augmented
minute ventilation
Operator sets a minimum E which usually is 70% 90% of patient’s current E. The ventilator provides
whatever part of the E that the patient is unable to
accomplish. This accomplished by increasing the
breath rate or the preset pressure.
It is a form of PSV where the PS level is not set, but
rather variable according to the patient’s need
MMV (Mandatory Minute Ventilation)
• Indications
– Any patient who is spontaneously and is
deemed ready to wean
– Patients with unstable ventilatory drive
• Advantages
–
–
–
–
Full to partial ventilatory support
Allows spontaneous ventilation with safety net
Patient’s E remains stable
Prevents hypoventilation
MMV (Mandatory Minute Ventilation)
• Disadvantages
• An adequate E may not equal sufficient A (e.g.,
rapid shallow breathing)
• The high rate alarm must be set low enough to
alert clinician of rapid shallow breathing
• Variable mean airway pressure
• An inadequate set E (>spontaneous E) can lead
to inadequate support and patient fatigue
• An excessive set E (>spontaneous E) with no
spontaneous breathing can lead to total support
PAV (Proportional Assist Ventilation)
• Provides pressure, flow assist, and volume
assist in proportion to the patient’s
spontaneous effort, the greater the patient’s
effort, the higher the flow, volume, and
pressure
• The operator sets the ventilator’s volume and flow
assist at approximately 80% of patient’s elastance
and resistance. The ventilator then generates
proportional flow and volume assist to augment
the patient’s own effort
Drager Evita 4
PAV (Proportional Assist Ventilation)
• Indications
• Patients who have WOB problems
associated with worsening lung
characteristics
• Asynchronous patients who are stable and
have an inspiratory effort
• Ventilator-dependent patients with COPD
Drager Evita 4
PAV (Proportional Assist Ventilation)
• Advantages
– The patient controls the ventilatory variables ( I, PIP,
TI, TE, VT)
– Trends the changes of ventilatory effort over time
– When used with CPAP, inspiratory muscle work is
near that of a normal subject and may decrease or
prevent muscle atrophy
– Lowers airway pressure
Drager Evita 4
PAV (Proportional Assist Ventilation)
• Disadvantages
– Patient must have an adequate spontaneous respiratory drive
– Variable VT and/or PIP
– Correct determination of CL and Raw is essential (difficult).
Both under and over estimates of CL and Raw during
ventilator setup may significantly impair proper patientventilator interaction, which may cause excessive assist
(“Runaway”) – the pressure output from the ventilator can
exceed the pressure needed to overcome the system
impedance (CL and Raw)
– Air leak could cause excessive assist or automatic cycling
– Trigger effort may increase with auto-PEEP
BiLevels
What is BiLevel Ventilation?
• Is a spontaneous breathing mode in
which two levels of pressure and hi/low
are set
• Enabled utilizing an active exhalation
valve
• Substantial improvements for
spontaneous breathing
– better synchronization, more options for
supporting spontaneous breathing, and
potential for improved monitoring
BiLevel Ventilation
Spontaneous Breaths
Synchronized Transitions
60
Spontaneous Breaths
Paw
cmH20
-20
1
2
3
4
5
6
7
What is BiLevel Ventilation?
• At either pressure level the patient can breathe
spontaneously
– spontaneous breaths may be supported by PS
– if PS is set higher than PEEPH, PS supports
spontaneous breath at upper pressure
BiLevel Ventilation
60
PEEPH
Paw
PEEPL
cmH20
-20
Pressure Support
PEEPHigh + PS
1
2
3
4
5
6
7
Then What Is APRV?
• Is a Bi-level form of ventilation with sudden short
releases in pressure to rapidly reduce FRC and allow for
ventilation
• Can work in spontaneous or apneic patients
• APRV is similar but utilizes a very short expiratory time
for PRESSURE RELEASE
– this short time at low pressure allows for ventilation
• APRV always implies an inverse I:E ratio
• All spontaneous breathing is done at upper pressure
level
APRV (Airway Pressure Release
Ventilation)
•
•
Provides two levels of CPAP and allows spontaneous
breathing at both levels when spontaneous effort is present
Both pressure levels are time triggered and time cycled
APRV (Airway Pressure Release
Ventilation)
•
•
Allows spontaneously breathing patients to breathe at a
high CPAP level, but drops briefly (approximately 1
second) and periodically to allow CPAP level for extra
CO2 elimination (airway pressure release)
Mandatory breaths occur when the pressure limit rises
from the lower CPAP to the higher CPAP level
APRV (Airway Pressure Release
Ventilation)
• Indications
• Partial to full ventilatory support
• Patients with ALI/ARDS
• Patients with refractory hypoxemia due to
collapsed alveoli
• Patients with massive atelectasis
• May use with mild or no lung disease
APRV (Airway Pressure Release
Ventilation)
• Advantages
• Allows inverse ratio ventilation (IRV) with or without
spontaneous breathing (less need for sedation or paralysis)
• Improves patient-ventilator synchrony if spontaneous
breathing is present
• Improves mean airway pressure
• Improves oxygenation by stabilizing collapsed alveoli
• Allows patients to breath spontaneously while continuing
lung recruitment
• Lowers PIP
• May decrease physiologic deadspace
APRV (Airway Pressure Release
Ventilation)
• Disadvantages and Risks
• Variable VT
• Could be harmful to patients with high expiratory
resistance (i.e., COPD or asthma)
• Auto-PEEP is usually present
• Caution should be used with hemodynamically
unstable patients
• Asynchrony can occur is spontaneous breaths are
out of sync with release time
• Requires the presence of an “active exhalation
valve”
Airway Pressure Release Ventilation
Spontaneous Breaths
60
Paw
Releases
cmH20
-20
1
2
3
4
5
6
7
8
Modes of Ventilation
• Questions?