Bangladesh BEST Programme Calibration of Volumetric Glassware By Gravimetric Method Nihal Gunasekara Sri Lanka

Bangladesh BEST Programme

Laboratory used Glassware

Burette

Single-volume pipette

Graduated pipettes

One-mark volumetric flasks

Graduated measuring cylinders

Bangladesh BEST Programme

DEFINITIONS

Unit of volume Cubic centimeter : ml

Ref. temperature 20 °C

To deliver Apparatus is calibrated to accurately transfer the stated volume to another container

.

To contain Apparatus is calibrated to an accurate measure of the volume contained

Bangladesh BEST Programme

Apparatus Needed

Balance

Thermometer

Hygrometer

Barometer

Water Distilled water /De-ironized

Pipette Sucker

Bangladesh BEST Programme

Recommended Apparatus - Balance

Selected Volume (V) 10 ml < V < 1000 ml 1000 ml < V < 2000 ml V > 2000 ml Resolution of Balance (mg) Std. Deviation (mg) 1 1 10 10 100 100

Bangladesh BEST Programme

Recommended Apparatus/Liquids

Apparatus/Liquid Thermometer Hygrometer Barometer Calibration Liquid Maximum Error 0.1 °C 5% 1 k Pa Distilled /de-ionized water

Bangladesh BEST Programme

Test Room Conditions

Temperature Humidity

M

Stabilization Maintained at 20 ± 1 °C and during calibration ± 0. 5 °C stability Maintained 35% to 85% Prefer to maintain at 60 ± 10% Volumetric instruments and test water 1 to 2 hours

Bangladesh BEST Programme Preparation

   

Cleaning of volumetric glassware

Internal surface of volumetric vessels shall be cleaned sufficiently before the calibration Oil or grease is removed by suitable solvent The vessel should be nearly filled with an aqueous solution of a soap less detergent and shaken vigorously. It should then be repeatedly rinsed with distilled water Be sure that vessel is cleaned before commence calibration

Bangladesh BEST Programme Calibration Procedure

   

Prior to any calibration test make sure that all instruments reached the ambient temperature in the laboratory During the calibration maximum temperature variation in the laboratory should not exceed ±0.5

o C.

Set or read a meniscus against a reference line or scale Be sure that the vessel or weighing bottle and the distilled water are at room temperature.

Bangladesh BEST Programme Calibration Procedure

  

Fill distilled water up to the required volume Weigh the testing container with distilled water.

Measure the temperature of distilled water using the thermometer.

Repeat the above step three times and record in the data recording form.

Repeat the above steps for cylinders at least at four different volumes

Bangladesh BEST Programme Calibration Procedure –Flask & Cylinder (contain) 1. Weigh the empty testing container and record the value 2. Fill distilled water up to the required volume 3. Weigh the testing container with distilled water .

4. Measure the temperature of distilled water using the thermometer.

5. Repeat the above step 2 through 4 three times and record in the data recording form.

6. Repeat the above step 2 through 5 for cylinders at least at three different volumes

Bangladesh BEST Programme Calibration Procedure – Pipette (deliver) 1. Weigh a empty container and record the value 2. Fill pipette with distilled water up to the index mark 3. Any water on the outside of the tip is removed with filter paper after the filling is completed 4. Transfer the distilled water to empty vessel, weighted in step (1) 5. Weigh the vessel with distilled water .

6. Measure temperature of distilled water using the thermometer 7. Repeat the above steps three times and record in the data recording form.

Bangladesh BEST Programme Calibration Procedure – Burette (deliver) 1. Weigh empty container and record the value 2. Fill burettes with distilled water up to the index mark 3. Any water on the outside of the tip is removed with filter paper after the filling is completed 4. Transfer the distilled water to empty vessel, weighed in step (1) 5. Weigh the vessel with distilled water 6. Measure temperature of distilled water using the thermometer.

7. Repeat the above steps three times and record in the data recording form.

8. Repeat the above steps 1 through 7 for Burettes at least three different levels

Bangladesh BEST Programme Principle of calibration - volumetric glass ware

V

20   (

R L

R E

 *    

w

1  

a

   *    1   

b a

   *  1   

t

 20   Bouncy effect Temperature effect

Z values are given in Tables B6, B7 and B8 in ISO 4787 : 2010 for different types of glass at common air pressure Vs temperature

14

Bangladesh BEST Programme Notations

R L-

is the balance reading of vessel with water reference liquid in grams

R E -

is the balance reading of empty vessel in grams Ρ

a -

is the density of air, in grams per milliliter ρ

b

- when weighing in air as though the density of the weight was 8.0 g/ml.

ρ

w

- is the density of water at t o C, in grams per milliliter.

ϒ - is the coefficient of cubical thermal expansion of the material of which the item of glassware, tested is made in reciprocal degrees Celsius, t - is the temperature of the water used in testing in degrees Celsius.

Bangladesh BEST Programme

Density of Air-

Table B3 ISO 4787 Temperature Air Pressure hPa 930 940 950 960 970 980 990 1000 1010 19 1.104 1.116 1.1128 1.140 1.152 1.164 1.176 1.188 1.200

20 1.100 1.112 1.124 1.136 1.148 1.160 1.172 1.183 1.195 21 1.096 1.108 1.120 1.132 1.144 1.155 1.167 1.179 1.191

Bangladesh BEST Programme Density of water

Table B4 ISO 4787 Temperature Density (g/ml) °C ρ

w

19 0.99840

20 0.99820

21 0. 99799

Bangladesh BEST Programme Coefficient of Cubical Thermal Expansion Material Coefficient of cubical thermal

expansion ϒ -1 °C 10 -6

Borosilicate glass 3.3 9.9

Borosilicate glass 5.0 15 Soda-lime glass 27

Bangladesh BEST Programme Data Recording

Use spread sheet for data recording

Record Temperature, Humidity, Barometric pressure

Record final data in a data recording sheet

Review from a third party

Prepare the final calibration certificate

Bangladesh BEST Programme