ORB-SCAN-1390
Download
Report
Transcript ORB-SCAN-1390
Corneal topography
orbscan
S.A.A MORTAZAVI MD
Associate professor of ophthalmology
Feiz hospital 1390
2
ORBSCAN SYSTEM
Use
the principle of projection
Forty
scanning slit beams (20 from the
left and 20 from the right with up to
240 data points per slit ) to scan the
cornea and measure independently the
X,Y & Z locations
3
Orbscan imaging
Forty slit images are acquired in two 0.7
second periods
Each of the 40 slit images triangulates one
slice of ocular surface
Distance between data slices average 250
microns
5
ORBSCAN
Orbscan
I only slit scan topography
Orbscan
II the placidodisc added in
orbscan I
7
ORBSCAN
The images used to construct the anterior
corneal surface,posterior corneal
surface,anterior iris and anterior lens
surfaces
Data regarding the corneal pachymetry and
anterior chamber depth
8
Elevation
Orbscan measure elevation
Elevation is important the only complete
scaler measure of surface shape
Both slope & curvature can be
mathematically derived from a single
elevation map
BEST FIT SPHERE (BFS)
The computer calculates a hypothetical sphere
that matches as close as possible to the actual
corneal shape being measured
Compares the real surface to the hypothetical
sphere showing areas above the surface of the
sphere in warm colours and areas below the
surface in cool colours
10
Topography quad map
The
upper left : anterior float
The
upper right : posterior float
The
lower left : keratometric pattern
The
lower right : pachymetry map
12
13
14
NORMAL BAND SCALE
Highlights the abnormal areas in the
cornea in orange to red colors
The normal areas are all shown in green
Helpful in generalized screening in
preoperative examination
16
AXIAL MAP
Provides detailed keratometric
information across the diameter of the
cornea
K readings are between certain values
the cornea must be neither too steep nor
too flat
17
AXIAL MAP
To create a good quality corneal flap in
LASIK if either extremes (too steep or too
flat) is the case, this can lead to surgical flap
complications
K readings of more than 48 D are an
indication of potential keratoconus
18
19
20
Middle box
Keratometric readings
White to white distance in mm
Angle kappa readings
The thinnest point of cornea
irregularity within the central 3 mm & 5
mm
22
PACHYMETRY MAP
The orbscan measures thickness from the
tear film layer to descemet’s membrane and
is thicker than that obtained with
ultrasound
Adjustment factor (acoustic factor) ,the
default setting is 92%
Provides a reading showing the thinnest
point of the cornea that may not
necessarily be the central reading
24
PACHYMETRY MAP
Thinnest
In
point <470 micron
pathological corneas, thinnest point is
often displaced inferotemporal
Difference
of >100 microns from the
thinnest point to the values at 7mm
optical zone
25
ELEVATION MAP
The
green colour is referred as
refrence sphere (at sea level )
The
warmer colours are above this
level and the cooler colours are below
26
27
28
ANTERIOR ELEVATION MAP
Looking at a proper scale in the cornea ,
can see height differences
Compare the height of the actual cornea
to a best fit sphere
Posterior map
The
highest elevation value as a
keratoconus indicator or at least as a
screen for patients may be at risk of
developing keratectasia
55
D elevation as an absolute cut off
30
ELEVATION DATA
The
difference between the highest
and lowest points is a potential
keratoconus indicator if over 100
microns (Rousch criteria)
31
DIAGNOSTIC CRITERIA
Power map changes
Posterior elevation maps
Pachymetry
Composite/integrated topography
information
32
POWER MAP
Mean
corneal power >45D
In addition to steep corneal
curvatures-the bowtie or broken
bowtie appearance indicative of early
keratoconus
Central corneal asymetry a change
within central 3mm optical zone of
the cornea of more than 3D
33
Irregularity in central cornea
Greater than 1.5 D in 3 mm zone and
greater than 2.0 D in 5 mm zone is
considered abnormal and cause for
concern
POSTERIOR ELEVATION MAP
Many
surgeons think the first sign
of keratoconus appears on the
posterior surface of the cornea
3.13% of population screened for
laser surgery had posterior ectasia
criteria by orbscan , despite having
axial topography classified as
normal
35
36
POSTERIOR ELEVATION MAP
The
most common reference
surface for viewing elevation maps is
the best fit sphere
A
best fit sphere (BFS) >55D on the
posterior profile , indicative of
posterior ectasia
37
Posterior float difference
Greater than 50 micron generally
accepted as abnormal
In corneas thinner than normal over 40
as abnormal
Posterior elevation map
The location of the steepest part of the
posterior float should be relatively central ,
but is a more concern it be located away from
the center and in an area of corneal thinning
Posterior float difference;40 to 50 microns
seems to be the maximum difference
40
Correlation of signs of the
highest point
Highest
point on the posterior
elevation coincides with the highest
point of anterior elevation , the
thinnest on pachymetry and the point
of steepest curvature on the power
map
41
42
Although
high posterior elevation
and ratio between two elevation
maps rarely used as exclusion criteria
alone , but by considering these
together , more conclusive
information can be obtained
43
Risks of ectasia indices
Number of abnormal maps
Posterior float difference >0.050
3mm & 5mm irregularity
Peripheral thickness changes
Astigmatism variance between eyes
Steep k’s –mean power map
45
Three step rule
One abnormal map ; perform with
caution
Two abnormal map ; with concern
Three abnormal map ;contraindicated
47
48
49
Composite/integrated information
Similarly between anterior & posterior
profiles a forward bending of areas shown
above the BFS and association with the
thinnest point on the cornea
Inferotemporal displacement of the highest
point
50
51
52
Abnormal tear film
Can significantly distort the readings
The significant change in surface quality
and validity of the dry eye
54
55
56
57
58
59
60
61
62
63
64
65