Air Sensitive Techniques 1: Using Schlenk Lines

Download Report

Transcript Air Sensitive Techniques 1: Using Schlenk Lines

Air Sensitive Techniques 1:
Using Schlenk Lines
Advanced Research Techniques Workshop
April 20th 2011
James Watson
Outline
Why use air-sensitive techniques?
• Common air-sensitive compounds
Equipment
• Specialist glassware
• The Schlenk line
• Purge and refill
• Cycling tubing
Handling Reagents
• Dry solvents and distillation
• Measuring and weighing
• Spills/Fires
Clean-up
• Reagents
• Glassware
• Syringe/cannula
Workshop 20/04/11
Why Use Air Sensitive Techniques?
•
Many commonly used chemical reagents are extremely reactive towards water
and/or oxygen and will oxidise, decompose, ignite or even explode
•
Air sensitive techniques have been developed to avoid contact with air and
moisture from the atmosphere at every step
!! these techniques do NOT protect from explosive or shock sensitive materials,
and provide limited protection from toxic compounds !!
Common Air Sensitive Compounds
•
•
Examples of pyrophoric compounds:
-
Metal alkyls and aryls e.g. RMgX, RLi, RNa, R3Al, R2Zn
-
Metal hydrides e.g.. NaH, KH, LiAlH4
-
Hydrides e.g.. B2H6, PH3, AsH3
-
Boranes, phosphines, arsines
Examples of chemicals which react violently with water
-
Metal hydrides
-
Metal amides
-
Alkali metals e.g.. Na, K, Cs
-
Many main group halides e.g.. BCl3, BF3, AlCl3, PCl3, SiCl4
Specialist Equipment
•
Glove Box – for solid reagents that must be handled under completely dry conditions
•
Glove Bags are also available cheaply for less sensitive compounds (not suitable for
pyrophoric materials)
Air Sensitive Techniques Apparatus
Cap
Tipper funnel
N2 inlets
Schlenk flask
Ampoule/graduated ampoule
Dropping funnel
Air Sensitive Techniques Apparatus
Glass syringe and plunger
Cannula
Frit
Filter stick
Condenser
The Schlenk Line (or Double Manifold)
C
B
A
E
D
G
H
F
The Schlenk Line (or Double Manifold)
A. Metre-long nitrogen bubbler
E.
B. Nitrogen line
F. Vacuum tubing
C. Tubing to nitrogen supply
G. Trap surrounded by liquid N2 dewar
D. Vacuum line
H. Tubing to vacuum pump
•
Two-way double oblique manifold taps
After assembly, the Schlenk line must be purged with nitrogen as follows (only
suitable for Schlenk lines with 1m long bubbler and sufficient Hg in the reservoir):
1. Connect 2 taps with one piece of tubing, ensuring the nitrogen supply is off
2. Turn one of the taps to vacuum, then slowly open the other tap to N2 to pull the
mercury up the exhaust line, whilst the air is sucked from the manifold
3. Close the tap turned to the N2 manifold, and turn on the nitrogen supply to purge
4. Repeat steps 2 – 4, twice
The Schlenk Line (or Double Manifold)
•
Before switching on the vacuum pump, do a visual check of the line:
 check all taps on the manifold are closed
 check all joints are streak-free
 ensure the trap is empty and connected to the line with no streaks at the joint
 check the nitrogen supply by turning on the tap until there is a good flow through
the mercury bubbler
Liquid Nitrogen
•
Liquid nitrogen is a cryogenic fluid which is extremely cold and boils at -196 oC
•
Small volumes of liquid vaporise into large volumes of gas, ~700 times expansion
!! wear safety goggles and protective clothing at all times !!
!! never draw air through traps containing liquid N2, as explosive liquid O2 can condense !!
Hazards

Cold burns can occur very rapidly and over a wide area – always wear
thermoprotective gloves (available from Gas Safety UK, product code 99-002, £16)
!! do not wear cloth gloves of any kind !!
 Asphyxiation can occur as liquid nitrogen rapidly expands to large volumes of
nitrogen gas.
!! never accompany a dewar containing liquid N2 in a lift !!
 Over pressure will occur if the rapidly boiling liquid N2 is used
in a closed system. Beware of an “ice-dam” forming in apparatus.
!! never screw a lid on a thermos flask containing liquid N2 !!
Pump Maintenance
•
•
Oil Change: every 6 months or when necessary
1.
open outlet valve and decant oil into waste bottle/drum via funnel
2.
pour fresh oil into the inlet valve using a clean funnel to flush out old oil
3.
close outlet valve and fill pump
Measuring Pressure using a McLeod gauge
1.
attach to the Schlenk line using rubber tubing
2.
turn the two-way double oblique manifold tap to vacuum slowly
3.
carefully rotate the gauge to the vertical position to read the vacuum pressure
4.
turn back to horizontal and repeat step 3 until the reading is constant

ideal pressure should be 10-2 torr or lower
Pump Maintenance
•
Poor Vacuum: cannot achieve a good vacuum
1.
check vacuum tubing for degradation and replace if necessary
2.
check oil level and for discolouration – change if necessary
3.
check joints and taps in line and ensure there are no streaks in grease
4.
if adequate vacuum cannot be achieved, take the pump to the workshop
to be tested and cleaned
!! ensure liquid N2 is always topped up around traps
to prevent volatiles contaminating pump oil !!
Purge-and-Refill: How to Cycle Tubing
•
Cycling tubing and glassware using the purge-and-refill method ensures all air is
removed and replaced with inert gas
1.
turn on the vacuum pump and have liquid N2 around the solvent traps
2.
turn on the nitrogen supply and ensure the bubbler is at a gentle rate
3.
connect glassware to a tap on the line using thick-walled rubber tubing
4.
cycle the tubing only (the tap on the glassware remains closed):
a.
turn the two-way double oblique manifold taps to vacuum to purge
b.
after a few seconds, rotate the tap to refill with nitrogen (the
mercury/oil in the bubbler will be sucked up). If there are already
other taps in use under nitrogen, close these before turning new
taps to nitrogen, and only re-open when the pressure is equalised
c.
wait for the pressure to equalise and the bubbler to begin bubbling
d.
repeat steps a – c twice more to purge-and-refill
Purge-and-Refill: How to Cycle Tubing
5.
with tubing under nitrogen, the stopcock on the glassware can now be opened
6.
when the bubbler begins to bubble again, the pressure is equalised and the
flask and tubing are ready for further manipulation
n.b. if connecting more than one piece of glassware and tubing to the line, the
above steps must be carried out simultaneously for each piece
Schlenk line tap
two-way double oblique manifold taps
stopcock
Positive Nitrogen Flow
•
at all times during a reaction, the system should be
under a slight positive pressure of nitrogen (positive
flow) which can be visually indicated by the mercury or
oil bubbler
•
when removing stoppers or caps from apparatus, there
must be a positive flow of N2. This provides a protective
atmosphere “blanketing” the glassware from entry of air
and moisture
Glass-Aware
Plan ahead!
1. Plan the whole reaction ahead of time (a diagram for each step can be helpful at first!)
 how many Schlenk flasks will I need?
 will I need a dropping funnel/condenser/frit?
Laboratory glassware contains a thin film of adsorbed moisture which must be removed
2. Pre-heat all glassware in an oven (>100 oC, > 30mins), then assemble whilst hot using
silicone grease at the joints (see next slide), and allow to cool under vacuum.
Open joints must be closed with stoppers or glass caps.
 Schlenks may also be flamed under vacuum with great care using a Bunsen
burner or heat gun
!! never heat a sealed system !!
3. When cool, purge the glassware with nitrogen and cycle. Schlenk flasks can then be
weighed with their stoppers and stirrer bars.
 Mark Schlenks and stoppers with corresponding symbols to avoid mix-up
Greasing Joints
•
when assembling glassware using grease there should be no streaks in the grease at
the joints. This ensures that the joints will not seize and that no air can enter.
1.
apply a small amount of grease to the male joint and spread along the length
2.
assemble the joint and rotate carefully to distribute the grease. If streaking
occurs, add an additional small amount of grease to cover this area
3.
the joint is correctly greased if no streaks appear
streaked joint
streak-free joint
Dry Solvents
•
Dry solvents must always be used, as standard bench top solvents can contain up to
30% water. These must be dried and handled carefully.
Distillation stills

apparatus and drying agents
Storage

molecular sieves
Transferring to reaction vessel

cannulae / syringes
Distillation Apparatus
1.
Round bottom flasks (still pot)
2.
Heating mantle
3.
Condenser
4.
Inert gas via bubbler
5.
Solvent collector
6.
Solvent outlet
7.
Syringe port
8.
Overflow
4
3
7
5
8
1
6
2
Drying Agents
Solvent
Diethyl ether* / petroleum ether*
Drying agent
Distil over sodium pieces (0.5% w/v) and
benzophenone (1 spatula per litre)
Dichloromethane
Calcium hydride 5% w/v
Toluene*
Distil over sodium pieces (1% w/v)
DMF*
Distil over calcium hydride (1% w/v)
THF*
Distil over potassium pieces (1% w/v)
Acetonitrile*
Distil over calcium hydride (1% w/v)
* these solvents can be pre-dried over sodium.
!! never add sodium to chlorinated solvents – an explosion may occur !!
Collecting and Storing Dry Solvent
•
Once collected, dry solvent must be stored in a suitably dry manner. This can be
achieved by using 4Å molecular sieves which have a cage-like crystalline structure
which captures water.
•
To activate molecular sieves:
1.
Pre-heat molecular sieves in hot oven overnight. Leave ampoule in oven for at
least 1h
2.
!! Young's taps will be damaged if placed in the oven !!
Pour sieves into ampoule and insert Young’s tap (but do not close)and connect
to Schlenk line. Turn the Schlenk line tap to vacuum.
3.
Place ampoule in a heating mantle, using a clamp stand. Turn heat dial to 7.
4.
Heat for at least two hours. Turn off heat, and allow ampoule to cool under
vacuum
5.
When cool, cycle ampoule, and leave under nitrogen until
ready to use
Collecting and Storing Dry Solvent
•
Collecting solvent from distillation apparatus
1.
turn on the heating mantle, water supply and check
N2 flow. If the collector has a tap close it.
2.
when collector is full, turn off heating mantle and
allow solvent to cool
3.
connect ampoule with sieves to an inert gas line
using PVC tubing; turn the gas on first
4.
insert a needle through the tubing to purge air from
the Young's joint
5.
remove Young's tap and place ampoule under
solvent outlet. Turn off N2 supply to prevent blowback.
6.
collect
solventapparatus
into the ampoule
!!
distillation
can be dangerous,
do not use without proper training !!
Transferring Dry Solvent from Ampoule to Schlenk
1.
connect your solvent ampoule and dried Schlenk flask (see Glass-Aware) to the
Schlenk line and cycle tubing
2.
unscrew (anticlockwise) Young's tap and remove. Maintain a positive pressure of N2,
creating an inert blanket over the solvent.
3.
quickly and carefully insert the correct sized subaseal (usually no. 21) containing a
bleed needle (to purge the cavity with N2) into the neck of the ampoule
4.
open the stopcock of the Schlenk flask.
Maintain a positive pressure of N2 and
remove the stopper. Insert a subaseal
containing a bleed needle. Be sure to rotate
this bleed needle to purge the large cavity,
then remove the needle.
Transferring Dry Solvent
5.
Insert one end of a cannula/double-tipped needle into the ampoule and check the
nitrogen is flowing through against your hand. Insert the other end into the Schlenk
flask.
6.
Put a bleed needle through the subaseal of the Schlenk flask, and check nitrogen
flow against your hand. Close the Schlenk flask tap and lower the cannula/doubletipped needle into the solvent to transfer.
6. The pressure of nitrogen will now force the solvent into
the empty Schlenk.
7. After transferring the desired amount, lift the cannula
out of the solvent in the ampoule.
8. Now open the tap to restore N2 flow, remove the bleed
needle and cannula. Clean cannula as described later
(Syringe/ Cannula/ Needle Clean Up slide)
Cannula
Measuring and Weighing Dry Reagents
•
Dry reagents must be carefully measured or weighed to ensure that their potential
contact with air and moisture is minimised.
Dry Solid Reagents: from the bottle

powder funnels
Dry Solid Reagents: from a Schlenk flask

transfer of solids from Schlenk flask to Schlenk flask
Dry Liquid Reagents: from the bottle

cannulae/syringes
Weighing Out: Dry Solid from a Bottle
1. prepare a dry, pre-weighed Schlenk flask (see Glass-Aware) and a dried powder
funnel (long-necked funnels are better to avoid compound sticking to greasy joint!)
2. clamp the Schlenk flask and connect to the Schlenk line. Cycle the tubing.
3. while under a gentle flow of nitrogen, remove the stopper. Test the flow against
your hand (too strong a flow will blow your solid back out of the flask)
4. do not block the joint completely with funnel (solid will blow out), hold it just
above or use a ring clamp
5. add solid to flask. Remove any solid sticking to the grease with a tissue and regrease stopper. Ensure all solid is removed, it can cause the joint to crack.
6. replace the stopper and weigh the flask.
Weighing Out: Dry Solid from Schlenk flask to Schlenk flask

Solids likely to decompose over time may be better stored in a dried Schlenk under N2
1. Prepare a dry pre-weighed Schlenk flask (see Glass-Aware), and a tipper funnel capped
on one end, with an N2 inlet tap at the other end.
2. Clamp all glassware and connect your pre-weighed Schlenk, your reagent Schlenk, and
the tipper funnel to the Schlenk line. Cycle tubing simultaneously.
3. Under a positive flow of nitrogen, remove the stopper
from your pre-weighed flask and the cap from the tipper
funnel, and connect the two. Rotate the flask to ensure a
streak-free joint.
4. Close the N2 inlet tap on the tipper funnel and remove
this, along with the stopper from the reagent Schlenk,
and connect the funnel to the Schlenk flask. Rotate
joint to remove streaks.
Weighing Out: Dry Solid from Schlenk flask to Schlenk flask
5. Secure the joints in place with elastic banding so
they stay together while tipping.
6. Close the tap of your reagent Schlenk flask, and
carefully turn your pre-weighed flask to vacuum
(if there is a stirrer bar in your pre-weighed flask,
use a magnet on the outside of the flask to
secure it)
7. Carefully tip an estimated amount of solid into
the pre-weighed Schlenk.
8. Clamp the reagent Schlenk flask. Carefully turn
the tap of the pre-weighed Schlenk back to N2.
Open the tap of the reagent Schlenk flask to
restore the flow of N2.
Weighing Out: Dry Solid from Schlenk flask to Schlenk flask
9.
Under positive nitrogen pressure, remove the pre-weighed flask and replace the
stopper. Also replace the cap on the tipper funnel and turn this flask back to
vacuum.
10.
Now re-weigh your prepared Schlenk and calculate how much solid was transferred.
11.
If there was too much or too little, connect to the Schlenk line again and cycle the
tubing. Turn the reagent flask and tipper back to nitrogen, and repeat from step 4.
12.
When you are happy with the mass of solid transferred, remove the tipper under
positive nitrogen pressure and place stoppers in both flasks.
Measuring Dry Liquid Reagents: Syringes
Syringe transfer tips:
•
Small quantities (up to 20 mL) of air-sensitive reagents and dry solvents may be
transferred with a syringe equipped with a thin, 12-18inch long needle. This avoids
having to tip reagent bottles and storage flasks. Tipping often causes the liquid to
come in contact with the septum/subaseal causing swelling, deterioration and spills
Using syringes:
1. Pressurise the vessel with N2 (for bottles see next slide). Subaseals can be inserted
into Schlenks/ampoules (as described in transferring solvents).
2. Purge the syringe with N2 by inserting into the septum/subaseal and filling with N2
only; the N2 pressure should push the plunger back, so that the syringe fills, and
then remove from the bottle and push N2 out into fume hood.
3. Repeat step 2 twice more
Measuring Dry Liquid Reagents
Sure/Seal bottles
•
Sure/Seal bottles have a crown cap and liner crimped in place
which gives some protection against air and moisture from the
atmosphere. Needles and cannulae can be inserted through
this seal.
1. Pressurise the bottle with N2 by inserting a needle connected to
Crown cap with
hole
a Schlenk line via a Luer adapter (check the N2 flow first)
2. The reagent can then be dispensed using a syringe or doubletipped needle inserted through the hole in the metal cap.
1. Liquids likely to decompose over time may be better transferred
to a dried ampoule under nitrogen for storage.
Luer adapter
Measuring Large Volumes of Dry Liquid Reagents
•
When measuring large volumes of liquid, it is not advisable to use a syringe. Instead
use a graduated ampoule and a cannula.
1.
Connect a dry graduated ampoule and your reagent flask or bottle to the Schlenk line.
Cycle tubing simultaneously.
2.
Have ampoules and Schlenks under N2, and replace stoppers/ Young's taps with
subaseals. Reagent bottles can be pressurised with N2 as described on the previous
slide.
3.
Transfer liquid from a reagent bottle/ampoule to the graduate ampoule using an oven
dried cannula/double-tipped needle, as described in Transferring Dry Solvents. When
you have the desired amount, return nitrogen flow and remove the bleed needle
4.
Then transfer the liquid reagent from the graduated ampoule to reaction vessel, using
the cannula/double-tipped needle, and opening/closing the stopcock to alter the
nitrogen flow and adjust rate of addition.
NMR tube holder
•
Some reactions can be monitored by 31P and 11B
NMR spectroscopy, by preparing an NMR
sample in an NMR tube holder, as shown.
•
The NMR tube is placed in the holder and dried
in the same fashion as a Schlenk flask.
•
Transfer a sample of the reaction solution in
the same fashion as transferring solvent.
Attaching Glass Apparatus
Connecting condensers and dropping funnels
1.
Dry glassware and assemble as described in Glass-Aware.
Cap the bottom joint of the glassware with a greased cap
and insert a greased N2 inlet into the top joint, ensuring
there are no streaks.
2.
Connect the N2 inlet to the Schlenk line using PVC tubing,
and turn the two-way manifold tap vacuum. Allow to cool
under vacuum.
3.
Elastic band the cap and inlet, and cycle the glassware.
4.
Attach reaction vessel to Schlenk line and cycle tubing.
5.
Under positive N2 flow, remove stopper from flask and
the cap from the apparatus, and connect the two
Attaching Glass Apparatus
Connecting a frit and filtering solutions
1.
Dry glassware and assemble as described in Glass-Aware. Cap both joints of the
glassware with greased caps and insert a greased tap into the stopcock, ensuring
there are no streaks.
2.
Connect the N2 inlet to the Schlenk line using PVC tubing, and turn the two-way
manifold tap vacuum. Allow to cool under vacuum.
3.
Elastic band the caps and cycle the glassware.
4.
Attach reaction vessel to Schlenk line and cycle tubing.
5.
Under positive N2 flow, remove stopper from flask and bottom cap from the frit, and
connect the two.
6.
Transfer reaction solution by removing the top cap and inserting a subaseal
containing a bleed needle. Transfer using cannula techniques as described in
Transferring Dry Solvents
Removing Solvent
Solvent must be removed on the high-vac line using an external trap
1.
Connect external trap to the Schlenk line by the top outlet.
Connect the external trap to the Schlenk flask with a piece of
tubing from the side-arm. The Schlenk flask must be closed.
2.
Place a Dewar flask around trap and fill with liquid N2.
3.
Turn the Schlenk line to vacuum, and agitate the solution with a
stirrer bar or swirling. Slowly open the Schlenk stopcock. Solvent
will begin to bubble/bump
4.
When the bumping subsides the tap can be fully opened.
5.
The flask will become cold and solvent may freeze, warm very
gently heating with a heat gun.
7.
When all solvent is removed close the Schlenk stopcock, and close the Schlenk manifold
tap. Remove the tubing and take the trap out of the Dewar. Connect the flask to the line,
and cycle the tubing. Open the flask to N2.
Elastic Banding
Elastic bands are required to secure joints during flushing, as
the nitrogen pressure may open the seals of unsecured joints.
•
Rubber bands are more secure than clips which are quite
flexible. Stoppers could easily come out of their joints if
there is a build up of vapour pressure inside the Schlenk,
exposing your material to the atmosphere.
•
Schenk flask taps can also be secured with an elastic band.
Open:
during a reaction to prevent accidental removal
Closed:
Schlenk can be left in a basket with tap secured
and cannot be knocked out of position
Spills / Fires
Spills
•
Phosphines/ mercaptans must be cleaned using bleach
•
Lithium aluminium hydride can be wiped up using barrel roll, and can be quenched
by immediately and quickly submerging tissue in a sink full of water
•
Sodium hydride can be destroyed with alcohol
Fires
•
Always alert someone else first!
•
Small fires can often be allowed to burn out
•
In case of larger fires, use a dry powder extinguisher or fire blanket
Reagent Clean Up
•
Empty Sure/Seal bottles – remove the crown cap and leave in the fume hood to allow
the last traces of reagent to hydrolyse or oxidise
•
Destroy phosphines/mercaptans with bromine solution (IMS:bromine 100:1 approx.)/
bleach, followed with IMS, water and acetone (in that order).
•
Destroy organometallic compounds with IPA/IMS followed by water, then acetone
•
Alkali metals must be placed in a beaker filled with toluene, and destroyed using a
isopropanol:toluene solution (1:4). When doing this on a large scale (i.e. Still pot) do
this under an inert atmosphere.
Glassware Clean-Up
1.
Disassemble glassware
2.
Destroy phosphines, mercaptans and organolithiums as described on the
previous slide.
3.
Dispose of reagents as appropriate and rinse the glassware with water
4.
Wipe all grease from joints, stoppers, taps, caps etc.
5.
Place glassware in a base bath (KOH in IMS) ensuring there are no air bubbles and
the glassware is completely submerged, for a minimum of 4 hrs; 16hrs max
6.
Rinse glassware with water and place in an acid bath (HCl in water) for 1 hr min.
7.
Rinse glassware with water then acetone and put back in oven
!! always use thick nitrile gloves or marigolds !!
!! never put greasy glassware in the oven !!
Syringe/ Cannula/ Needle Clean Up
•
Clean all syringes and needles immediately after use (hydrolysis and oxidation
of reagents can cause needles to plug)
•
Syringes rinsed at the sink , using the plunger to clean the needle. To rinse a
cannula/double tipped needle, use a Buchner funnel as shown with the
Subaseal clamped into place. Apply vacuum to pull solvent through cannula.

Order of rinse:
Bromine solution (phosphines); IMS; Water; Acetone

Dilute acid can be used to unblock a needle

Dry plunger and syringe separately in the oven
•
Organolithiums must be destroyed using IMS
•
Phosphines/mercaptans must be rinsed with IMS into bromine solution
(IMS:bromine roughly 100:1)
Summary
•
Using air- and moisture- sensitive techniques is a necessary requirement
for handling many types of reagents
•
Dry glassware is essential
•
When removing stoppers/caps from Schlenk flasks and other apparatus
ensure that the inlet taps are open to nitrogen/inert gas flow
•
Sure/seals found on reagent bottle gradually degrade. Store liquid
reagents in ampoules for prolonged use
•
Solid reagents can be stored in Schlenk flasks for prolonged use
•
Remember to label all flasks and co-ordinating stoppers
Workshop 20th April 2011
1.
Weigh zirconocene dichloride (0.5g) into a sample vial on the balance and
transfer to a separate Schlenk flask using a powder funnel.
2.
Add dry THF (15mL) to the zirconocene using a syringe, then cool to -78oC.
3.
Weigh out dry potassium tbutoxide from a Schlenk flask via tipper funnel into a
pre-dried, pre-weighed Schlenk flask. Weigh out approx. 0.5g
4.
Add dry toluene from an ampoule using a syringe. Use 15mL.
5.
Add dry nbutyl lithium from an ampoule using a syringe. Use 2.8mL of 1.6M
solution.
6.
Now transfer the benzyl potassium into the zirconcene solution using a
cannula.
7.
Concentrate the solution to around 1/3 of the volume and filter through a frit.
References
•
Sigma-Aldrich Technical Bulletin AL-134
•
Advanced Practical Organic Techniques. J. Leonard, B. Lygo, G. Procter
•
http://www.safety.ncl.ac.uk/uploads/Handling%20of%20Liquid%20Nitroge
n.pdf
•
Advanced Practical Inorganic and Metalorganic Chemistry. R.J. Errington
Acknowledgements
Dr K. Izod
Mr J. Dyson
Miss L. Cotterill
The Postgraduate Committee
Other Workshops in the Series
Workshop
Date
Using Mestre-Nova and Delta to analyse NMR data
23rd March 2011
Air Sensitive Techniques 1: Using Schlenk Lines
20th April 2011
Effective Quenching and Working-up of Reactions
18th May 2011
Recrystallisation and Growing Samples for X-Ray
15th June 2011
Industry Techniques 1: Flash Chromatography
13th July 2011
Air Sensitive Techniques 2: Using a Glove box
10th August 2011
Industry Techniques 2: Microwave Reactions
7th September 2011
Dates, locations and titles to be confirmed
•
Sign-up sheet for each workshop will be available on the Postgraduate
Notice board two weeks in advance
•
PhD students: record attendance at each workshop in your e-portfolio
1 school PGRDP credit is available per workshop