www.mars-energo.com
Mars-Energo
V.O. 13 Line, 6-8, office 41H
Saint-Petersburg, Russia, 199034
Tel./Fax: +7 812 331-87-36
E-mail: [email protected]
MTS ME 3.3
Portable Multipurpose Test System
The System is meant for on-site testing
and calibration of service-class and
reference class measuring instruments
including electric energy meters,
instrument converters etc.
Testing of electric energy meters
MTS ME 3.3 portable test system is specifically designed to provide on-site meter test conditions
identical to in-lab tests.
A portable computer as part of the system
makes it possible to automatically conduct and facilitate test procedures. Energoform program oriented
at the Test System users makes it possible, among
other functions, to create and manage user-defined
test signals, profiles and test procedures, e.g. the
user can specify and store on the hard disc any load
conditions, parameters of test points etc. for the
three-phase system.
Requirements for meters under test are as follows:
Design: Single- or three-phase electromechanical and electronic meters.
Measured power (energy): active or reactive.
Input current: up to 7.7 A.
Actual standards: IEC standards defined as
series 62053 (11-23).
Accuracy: for service-class meters: 0.2S accuracy
class (or less accurate); for reference-class meters: power measurement error shall be 0.05% or worse (PF=1).
On-site testing of 0.2S meters
Competitive advantages of the Test System:
1. In addition to on-site testing and calibration of high-precision meters, the system can also
be used for various purposes owing to the extended
functionality of reference standard Energomonitor
3.1K-02, namely:
ƒƒTo measure power quality parameters and,
with Network Energomonitoring (EmWorkNet)
program, perform power quality analysis); e.g., before
testing meters, mains frequency, THD and deviation
of voltage can be measured;
ƒƒTo test current and voltage instrument
transformers of 0.2 accuracy class (complete with
the appropriate set of accessories);
ƒƒTo test DC meters provided that such design
option is selected.
2. Waveform generator Energoforma 3.3 provides test signals applicable to test instruments
measuring power quality as per Class A 61000-4-30
power quality measurement standard.
Test signals generated with a high degree of
accuracy and stability are distinguished by the following characteristics:
ƒƒ1st harmonic: 42.5…57.5 Hz.
ƒƒHarmonic composition: harmonics in the
range of (1st…50th) ; interharmonics and subharmonics
in the range of (0.5th; 1.5th; 2.5th;…50.5th).
ƒƒDesired network conditions (distorted or
not) can be imitated by specifying signal parameters
either directly via the menu or from a PC.
Energomonitor 3.1KM-02 (portable)
DESCRIPTION
Energomonitor 3.1KM-02 is a reference standard of
0.02* accuracy class introduced for testing/calibration/
adjustment of reference-class and service-class instruments that measure electric quantities and power quality
parameters. The instrument is supplied in durable case
handy for transportation and resistant to shock and vibration. The instrument supports data exchange with a PC
via serial interfaces RS-232 and/or USB.
Energomonitor 3.1KM is equipped with the following:
ƒƒ three voltage and three current measurement
channels;
ƒƒ inputs to accept signals from pulse outputs of
electric energy meters (Fin);
ƒƒ pulse outputs supposed to generate signal of the
frequency proportional to measured power (Fout);
ƒƒ signal input to accept signals from scaling
converters (current clamps) or CT test device (CTCS).
TESTING OF ENERGOMONITOR 3.1KM
A period of 1 year is considered maximum interval
between tests.
__________________
* Other options available: 0.05 and 0.1 accuracy class.
DELIVERY PACKAGE
Item description
Energomonitor-3.1KM reference standard
220V mains cable
User’s Manual
Calibration certificate
Manufacturer’s package
PC connection cable, RS232 interface
PC connection cable USB, interface (1.8 m)
Energomonitoring software on optical media
Optional accessories
Current clamps (10A, 100A or 1000A)
Flexible current clamps (300/3000A)
Current Transformer Calibration Switch with 220V mains cable and two cables for connection to the
instrument
Cable “Fout”
Cable “Fin”
Test leads 4 mm (1000V, 4 colors, cross section – 1 mm2)
Test leads 4 mm for current up to 12A (2 colors, cross section – 2.5 mm2)
Cable 100A (cross-section – 6 mm2 with banana plug and terminal lug)
Photoelectric scanning head SH-E
Photoelectric scanning head SH-I
Pulse Former PF
Quantity
1
1
1 copy
1 copy
1
1
1
1 CD
3
3
1
1
1
4
6
6
1
1
1
TECHNICAL CHARACTERISTICS
Description
Power supply
Maximum power consumption from mains (220V, 50Hz), VA
Protection provided by the enclosure
Dimensions (width × depth × height), mm, not exceeding
Weight, kg, not exceeding
Display of data
Lifetime estimate, years, at least
Mean time to first failure (MTFF), hours, at least
Operating conditions
Ambient temperature, °С
Relative humidity, %,
Atmospheric pressure, kPa (mm Hg)
Value
230V+10%; 50/60 Hz+5%
100
IP 20
480 × 380 × 200
10
Graphical LCD display; 240 × 128
10
35 000
20 ± 10
up to 80 at 20 °С
84–106.7 (630–800)
www.mars-energo.com
METROLOGICAL CHARACTERISTICS FOR EM-3.1KM-02 (accuracy class 0.02)
Parameter
Measured
Range
Type and Limit
of permissible error
Note
RMS of AC voltage
0.01 to
960V
Relative to the measured value
E< ± 0.015% (30…576V)
E< ± 0.055% (6…30V)
UN = 1, 2, 5, 10, 30, 60, 120,
240, 480, 800*V
RMS of AC current (direct
measurements)
0.005A
to 120A
Relative to the measured value
E< ± 0.015% (0.05…120A)
E< ± 0.1% (0.005…0.05A)
IN = 0.05, 0.1, 0.25, 0.5, 1.0,
2.5, 5.0, 10, 25, 50, 100 А
RMS of AC current (via current
clamps or CT Test device)
0.1…
4500A
relative
E< ±0.2% ... ±2%
Depends on current clamps
or CT Test device
Phase angle between the 1st
harmonics of phase voltages and
Phase angle between 1st voltage and
1st current harmonics in the same
phase
0° to 360°
absolute
±0.01°
Phase angle between nth voltage and
nth current harmonics, n = 2 to 40
0° to 360°
absolute
Active power (Р), I > 0.05A, U > 30V;
in relation to the apparent power;
for lower range error values see
Technical description manuals
0.03W
to 345 kW
E< ±0.015%
E< ±0.025%
cosφ = 1
cosφ =0.5
Reactive power (Q), I > 0.05A,
U > 30V; in relation to the apparent
power; for lower range error values
see Technical description manuals
0.03Var
to 345
kVar
E< ±0.03%
E< ±0.05%
sinφ = 1
sinφ = 0.5
Apparent power (S), I > 0.05A,
U > 30V; in relation to the measured
value; for lower range error values
see Technical description manuals
0.03VA
to 345kVA
E< ±0.03%
Voltage and current errors
are added up
Power factor (PF)
PF = P / S
0.1 to 1
absolute
±0.001
0.01…120A
0.2…960V
AC frequency
40Hz to
70Hz
absolute
±0.001Hz
0/01…120A
0.2…960V
Voltage dip duration
0.02 sec.
or longer
absolute
±0.02 sec.
49 Hz < F < 51 H
Voltage dip depth (Udip)
10% to
100%
relative
10.0 %
49 Hz < F < 51 H
Voltage swell height (over-voltage
factor)
110%
to 799%
relative
2.0 %
49 Hz < F < 51 H
Voltage swell duration
0.02 sec.
or longer
absolute
±0.02 sec.
49 Hz < F < 51 H
Flicker short-term perceptibility, Pst
0.25 to 10
relative
5.0 %
Reading of real-time clock
__________________
* subject to special order.
±0.3°
±1.0°
N/A
absolute
±2 sec. per 24 hours
0.01…120A
0.2…960V
12…528V; 0.01…120A
nth harmonic ratio: 2% to
15%
2 < n < 10
11 < n < 40
49 Hz < F < 51 Hz
∆U/U ≤ 20%, provided that
voltage envelope is meanderlike
Ambient temperature
10 to 35 °C
Energoforma 3.3
PURPOSE AND FIELDS OF APPLICATION
Hand-held Waveform Generator Energoforma 3.3
generates single- and three-phase system of current/voltage signals that can be used for testing and adjustment of
instruments that measure:
ƒƒ Active, reactive or apparent electric power and energy;
ƒƒ Power quality parameters as per IEC61000-4-30;
ƒƒ RMS of currents and voltages within commercial
frequency range.
The sphere of application includes mobile calibration (testing) laboratories and meter/transformer test systems where the Generator is used together with reference
meters, e.g. Energomonitor-3.1K or Energomonitor 3.3T1.
DESCRIPTION
Energoforma 3.3 supplied as a self-contained handheld module is provided with three channels for generating
voltages (phase voltages) and three independent channels
for generating currents. Waveforms and signal parameters
are viewed either on the built-in graphic LCD or PC via
“Energoform” software.
Delivery package
Energoforma 3.3
Power supply cable
User manuals
PC-connection cable (RS-232)
Energoform software
Item description
Optional accessories (included when specified in the Supply Agreement)
Set of measurement cables (7 pcs.)
Quantity
1
1
1 copy
1
1 CD
1
Technical characteristics
Parameter
Range of output voltage setting per each phase, V
Increments of voltage setting
Range of output current setting per each phase, A
Increments of current setting
1st harmonic frequency, Hz
Increments of frequency setting
Phase angle between 1st and nth voltage harmonics
of the same phase, angular degrees
Phase angle between 1st and nth current harmonics
of the same phase, angular degrees
Increments of phase angle setting
Harmonic composition, n
Power instability (during 1 minute or less ), %
Power output of current source, VA
Power output of voltage source, VA
Value
20...264
0.01
0.005...7.7
0.0001
42.5...57.5
0.01
0...360
0...360
0.01
1...50
±0.05
5
10