Transcript Slide 1

Inhaled anesthetic Delivery Systems

Sahmeddini MD Department of Anesthesia Shiraz medical university

Inhaled anesthetic Delivery Systems

Safety Standards

ANSI - (American National Standards

Institute) 1979 -- Standards set for all machines sold in

the U.S.

ASTM -- (American Society for Testing and

Materials 1988 1994: ASTM F1161-94 2000: ASTM F1850-00

 

To comply with the 2000 ASTM F1850-00 standard Continuous breathing system pressure

Exhaled tidal volume Ventilatory carbon dioxide concentration Anesthetic vapor concentration Inspired oxygen concentration Oxygen supply pressure

    

Arterial oxygen saturation of hemoglobin Arterial blood pressure Continuous electrocardiogram.

  

Functions of anesthesia machine Convert supply gases from high pressure to

low pressure Convert liquid agent to gas Deliver in a controlled manner Provide positive pressure for ventilation Alert the provider to malfunction

   

Prevent delivery of a hypoxic mixture

Testing Specific Components of the Anesthesia Delivery System 1)Calibration of the oxygen analyzer 2) The low-pressure circuit leak test 3) The circle system tests.

High Pressure System

Receives gasses from the high pressure E cylinders attached to the back of the anesthesia machine. ( 2200 psig for O2, 745psig for N2O

)

High Pressure System Receives gasses from the high

pressure E cylinders attached to the back of the anesthesia machine 2200 psig for O2 745 psig for N2O Usually not used, unless

  

pipeline gas supply is off

High Pressure System

Hanger Yoke

Hanger Yoke: orients

and supports the cylinder Providing a gas-tight

seal Ensuring a

unidirectional gas flow into the machine

Pin Index Safety System(PISS )

Prevents tank swaps

Pin positions

Air 1-5 Oxygen

2-5

Nitrous oxide

3-5

Pin Index Safety System(PISS)

Two sources of gas: Pipeline 50 psig Tanks

»Oxygen: 2200 psig »Nitrous oxide: 745 psig »Both reduced to 45 psig upon entering the

machine

• • • • •

Tank

H Tank E Tank

E Size Compressed Gas Cylinders

Cylinder Characteristics

Colour

Oxygen

White

Nitrous Oxide

Blue

Air

Black State Contents (L) Empty Weight (kg) Gas 625 5.90

Liquid and gas 1590 5.90

Gas 625 5.90

Full Weight (kg) Pressure Full (psig) 6.76

2000 8.80

750 6.50

1800

Approximate remaining time# Oxygen cylinder pressure(psig)

200 .oxygen flow rate(L/min)

Intermediate Pressure System

Intermediate Pressure System

Receives gasses

from the regulator or the hospital pipeline at pressures of 40-55 psig

Pipeline Inlet Connections

Mandatory N2O and O2,usually have air and suction too Inlets are non interchangeable due to

• •

specific threading as per the Diameter Index

Safety System (DISS)

Diameter Index Safety System (DISS)

Oxygen Pressure Failure Devices

Machine standard requires that an anesthesia machine be designed so that whenever the oxygen supply pressure is reduced below normal, the oxygen concentration at the common gas outlet does not fall below 19%

Oxygen Pressure Failure Devices

A Fail-Safe valve is present in the gas line supplying each of the flow meters except O2.

This valve is controlled by the O2 supply pressure and shuts off or proportionately decreases the supply pressure of all other gasses as the O2 supply pressure decreases

Oxygen Pressure Failure Devices

Historically there are 2 kinds of fail-safe valves Pressure sensor shut-off valve (Ohmeda)

Oxygen failure protection device (Drager) 

Pressure Sensor Shut-Off Valve

Oxygen supply pressure opens the valve as long as

it is above a pre-set minimum value (e.g.20 psig).

If the oxygen supply pressure falls below

the threshold value the valve closes and the gas in that limb (e.g..N2O), does not advance to its flow control

Pressure sensor shut-off valve

Oxygen Failure Protection Device

(OFPD)

Based on a proportioning principle rather than a shut-off principle.

The pressure of all gases controlled by the OFPD will decrease proportionately with the

Oxygen failure protection device

Oxygen Supply Failure Alarm

The machine standard specifies that whenever

the oxygen supply pressure falls below a manufacturer specified threshold (usually 30 psig)

alarm shall blow within 5 seconds.

Limitations of Fail-Safe Devices/Alarms

Fail-safe valves do not prevent administration

of a hypoxic mixture because they depend on pressure and not flow.

Do not prevent hypoxia from accidents such as

pipeline crossovers or a cylinder containing the wrong gas.

OXYGEN FLUSH VALVE

By passes vaporizer

Delivers large volumes

of oxygen to breathing circuit Is under high(er)

pressure caution!!!

OXYGEN FLUSH VALVE

Receives O2 from pipeline inlet or cylinder reducing device and directs high, unmetered flow directly to the common gas outlet (downstream of the vaporizer)  Machine standard requires that the flow be between

35 and 75 L/min

The ability to provide jet ventilation Hazards: May cause barotraumas Dilution of inhaled anesthetic     

Second-Stage Reducing Device Located just upstream of the flow control valves Receives gas from the pipeline inlet or the cylinder reducing device and reduces it further to

26 psig for N2O and 14 psig for O2

Purpose is to eliminate fluctuations in      pressure supplied to the flow indicators caused by fluctuations in pipeline pressure

Low Pressure System

Extends from the flow control valves to  the common gas outlet Consists of:

Flow meters Vaporizer mounting device Check valve

    

Common gas outlet

Flow Meter Assembly

Flow Meter

When the flow control valve is opened the gas enters at the bottom and flows up the tube elevating the indicator • The indicator floats freely at a point where the downward force on it (gravity) equals the upward force caused by gas molecules hitting •

Flow Meter Standards

Oxygen flow control knob Physically different

Larger and projects further Different shape All knobs are colour coded Knobs are protected

   

Electronic flow sensors Some newer anaesthesia workstations have now replaced the conventional glass flow tubes with electronic flow sensors that measure the flow of the individual gases. • These flow rate data are then presented to the anaesthesia care provider in either numerical format, graphic format, or a combination of the two.

Cracked tubes

In the presence of a  flow meter leak (either at the “O” ring or the glass of the flow tube) a hypoxic mixture is less likely to occur if the O2 flow meter is downstream of all other flow meters

Proportioning Systems

Mechanical integration of the • N2O and O2 flow control valves • Maintain a minimum 25% concentration of oxygen with a maximum N2O:O2 ratio of 3:1 •

Proportioning Systems

Proportioning Systems

Proportioning Systems

Vaporizers

A vaporizer is an instrument designed to change a liquid anesthetic agent into its vapor and add a controlled amount of this vapor to the fresh gas flow •

Vaporizers

Classification of Vaporizers

Methods of regulating output concentration Concentration calibrated Method of vaporization Flow-over Bubble through Injection Temperature compensation: Thermocompensation Supplied heat         

Applied Physics

Vapor pressure

Based on characteristics of agent Varies with temperature Boiling point: Vapor pressure equals atmospheric pressure Latent heat of vaporization: Heat required to change liquid into a vapor Comes from liquid environment • • • • • • •

Ohmeda and Drager Characteristics

Variable bypass Flow over Temperature compensated Agent specific Out of circuit

• • • • •

Basic Design

Gas enters vaporizer Flow is split Majority is by passed Some enters vaporizing chamber Saturated gas leaves chamber Diluted by bypass gas Delivered to patient • • • • • • •

Generic Bypass Vaporizer

Flow from the flow meters enters the inlet of the vaporizer The function of the concentration control valve is to regulate the amount of flow through the bypass and vaporizing chambers • Splitting Ratio = flow though vaporizing chamber/flow through bypass chamber • •

Factors that Effect Output

Flow rate

Accurate at most flows Lower than dial setting at both extremes of

 

flow Temperature Vapor pressure varies with temp

 

Accurate at 20 - 35 C

Factors that Effect Output

Intermittent back pressure

Retrograde flow Higher than dial setting especially at low flows and high ventilator pressures

Carrier gas composition

N2O causes transient drop • • • • • • •

Carbon Dioxide Absorbents

Two formulations of carbon dioxide absorbents are commonly available today: soda lime calcium hydroxide lime Baralyme, or barium hydroxide lime

• • • •

CO2 Absorption (con’t)

Soda lime –94% calcium hydroxide –5% sodium hydroxide –1% potassium hydroxide –silica to harden granules –ethyl violet as an indicator

• • • • • •

CO2 Absorption (con’t)

Ethyl violet is the pH indicator added to both soda lime Ethyl violet changes from colorless to violet when the pH of the absorbent decreases as a result of carbon

dioxide absorption. • •

Unfortunately, in some circumstances ethyl violet may not always be a reliable indicator of the functional status of absorbent

Anesthesia Ventilators

The ventilator on the modern anesthesia workstation serves as a mechanized substitute for the hand of the anesthesia care provider in manipulating the reservoir bag of the circle system, or another breathing system.

Ventilators Classified by:

Power source pneumatic – electric – both – Drive mechanism double circuit – driven by oxygen – • •

Ventilator Problems (con’t)

Leak in bellows assembly • Mechanical problems • Electrical problems •

Setting the Ventilator

Based on the principle that PaCO 2 is directly proportional to alveolar ventilation

AV X CO

2

= AV X CO

2 (what you have) (what you want) AV = alveolar ventilation CO 2 = carbon dioxide If you know 3, you can solve for the 4th

The Circuit: Circle System

Arrangement is variable, but to prevent re-breathing of CO2, the following rules must be followed: Unidirectional valves between the patient and the reservoir bag

Fresh-gas-flow cannot enter the circuit between the expiratory valve and the patient

Adjustable pressure-limiting valve (APL) cannot be located between the patient and the inspiratory valve

– •

Circle System

Relative stability Advantages: of inspired concentration – Conservation of respiratory moisture and heat – Prevention of operating room pollution – PaCO2 depends only on ventilation, not fresh gas flow – Low fresh gas flows can be used – High Complex resistance Disadvantages: design = potential for malfunction – (multiple one-way valves) = higher work of breathing – • •

The Adjustable Pressure Limiting (APL) Valve

User adjustable valve that releases gases to the scavenging system and is intended to provide control of the pressure in the breathing system Bag-mask Ventilation : Valve is usually left partially open. During inspiration the bag is squeezed pushing gas into the inspiratory limb until the pressure relief is reached, opening the APL valve. Mechanical Ventilation : The APL

• • •

The Adjustable Pressure Limiting (APL) Valve

Scavenging Systems

Scavenging is the collection and removal of vented anesthetic gases from the OR Since the amount of anesthetic gas supplied usually far exceeds the amount necessary for the patient, OR pollution is decreased by scavenging • •

Scavenging Systems

Workers should not be exposed to an eight hour time-weighted average of > 2 ppm halogenated agents (not > 0.5 ppm if nitrous

oxide is in use) or > 25 ppm nitrous oxide.

Evidence of harm to anesthesia personnel from waste gases is suggestive but unproved (strongest relationship is N2O and reproductive difficulties).

Type of Scavenging Systems

Scavenging may be active (suction applied) • passive (waste gases proceed passively down corrugated tubing through the room ventilation exhaust grill of the OR). •

Hazards of scavenging

Obstruction distal to interface

Occupational exposure

Barotrauma or inability to ventilate

System Checkout

High-Pressure System Checkout Check Oxygen Cylinder Supply.

Open the O 2 cylinder and verify that it is at

least half full (≈1000 psi) Close the cylinder.

Check Central Pipeline Supplies.

 

Check that hoses are connected and pipeline

gauges read about 50 psi.

Low-Pressure System

Check Initial Status of Low-Pressure System .

a. Close the flow control valves and turn the

vaporizers off. b. Check the fill level and tighten the

vaporizers’ filler caps.

Perform Leak Check of Machine Low Pressure System.

Verify that the machine master switch and flow control valves are OFF Attach a “suction bulb” to the common (fresh) gas outlet Squeeze the bulb repeatedly until it is fully collapsed.

   

Verify that the bulb stays fully collapsed for at least 10 seconds. Open one vaporizer at a time and repeat steps c and d as above. Remove the suction bulb and reconnect the fresh gas hose.

 

Turn on Machine Master Switch and All Other

a.

Necessary Electrical Equipment.

Test Flow Meters.

Adjust the flow of all gases through their

 

full range while checking for smooth operation of the floats and undamaged flow tubes. b.

Attempt to create a hypoxic O 2 /N 2 O mixture and verify correct changes in flow

and/or alarm.

Calibrate O

2

Monitor

c.

d.

a.

Ensure that the monitor reads 21% in

b.

room air. Verify that the low-O 2 alarm is enabled and functioning.

Reinstall the sensor in the circuit and flush

the breathing system with O 2 . Verify that the monitor now reads greater

than 90%.

Check Initial Status of Breathing System .

a.

Set the selector switch to “Bag” mode. b.

Check that the breathing circuit is

complete, undamaged, and unobstructed. c.

d.

Verify that CO 2 absorbent is adequate. Install the breathing circuit accessory

 

equipment (e.g., humidifier, positive end expiratory pressure [PEEP] valve) to be used during the procedure.

Perform Leak Check of Breathing System .

Set all gas flows to zero (or minimum).

Close the APL (pop-off) valve and occlude the

Y-piece. Pressurize the breathing system to about

30 cm H 2 O with an O 2 flush. Ensure that the pressure remains fixed for at

least 10 seconds. Open the APL (pop-off) valve and ensure that

the pressure decreases