Dissolved Oxygen
Tel: +44 (0) 1726 879800 www.partech.co.uk
Applications
• • • •
Wastewater Treatment Plants
– Used in activated sludge tanks for the control of nitrification strength
Surface Water Monitoring
– Oxygen measurement in rivers, lakes or seas as an indicator of the water quality.
Water Treatment
– Intake Protection – Reservoir Water Quality
Fish Farms
– Oxygen measurement and regulation for optimum living and growth conditions
Product Application – Wastewater Treatment
Activated Sludge - Dissolved Oxygen River - Dissolved Oxygen Tertiary Inlet - Dissolved Oxygen Storm Water Inlet works
Simplified Diagram of Activated Sludge System
Aeration tank From Primary Settlement To Outfall or Tertiary Treatment Clarifier Aeration System Sludge Recirculation Excess sludge extraction • • • • Activated sludge is a biochemical process for treating sewage and industrial wastewater that uses air (or oxygen) and microorganisms to biologically oxidize organic pollutants, producing a waste sludge (or floc) containing the oxidized material. The oxygen reacts with ammonia to form nitrate (nitrification process). Atmospheric air or pure oxygen is bubbled through primary treated sewage (or industrial wastewater) and combined with organisms to develop a biological floc which reduces the organic content of the sewage The combination of raw sewage and biological mass is commonly known as Mixed Liquor.
Why Measure Dissolved Oxygen?
Air is forced into the aeration basins, it increases the activity of the micro organisms and helps keep the organic waste thoroughly mixed Oxygen is added to the aeration basin to enhance the nitrification process by providing oxygen to combine with ammonia to form nitrates Micro organisms must have at least 0.1 to 0.3mg/l DO. Plants maintain the DO to 2 mg/l. If the DO is less than 2mg/l the bugs in the centre of the floc may die as the bugs on the outside of the floc use up the DO first. If this happens the floc breaks up and settles poorly When the DO comes too high it wastes energy. The aeration process takes between 30-60% of the total electrical power in a WwTW. Automated aeration process using DO according to the ESEPA may reduce energy costs by 50%.
Also adds benefit of mixing, bring the bugs, oxygen and nutrient together. This keeps the floc suspended and prevents it from settling
Ensure Success – Pick Your Location
Platform over aeration process • • • Find Representative sample point of the process Well mixed sample Safe to reach sensors Over aerated process with inaccessible probes Sample point is critical • • • • Avoid Dead Zone Extreme Turbulence Area next to Aerator Hazardous access to sensor • Reaching over handrails
Outlet
Oxidation Ditch
Aerator Inlet RAS Aerator Aerator
Where should the DO sensor go?
Outlet Aerator
Oxidation Ditch
Inlet RAS Aerator Not representative Not fully mixed Good Balance False High Aerator Here
Technologies
Sensors Technologies - Comparisons
The Sensing Environment for an Optical Cell consist of an optical arrangement where the membrane is excited by an LED to cause a luminescent response. The oxygen reacting with the surface of the membrane causes the luminescent properties to change and hence DO can be monitored.
Variations LED Wavelength: Green or Blue Membrane Material Membrane Life Active Membrane with Luminescent Molécules on an Inerte Material
Receiver Emitter
The Sensing Environment for a Galvanic Cell consists of a cathode and anode surrounded by an electrolyte. The oxygen passing through the membrane reacts with the electrolyte generating a potential difference proportional to the concentration of dissolved oxygen. Variations: Clarke Cell – electrolyte is consumed and sensing heads needs replacing.
Mackereth Cell – electrolyte is not consumed and no sensor replacement is required.
Membrane Materials Electrolyte Composition Anode/Cathode Materials Inert Membrane allows DO molecules to pass into the electrolyte
Anode Cathode
Dissolved Oxygen Sensors Technologies – Galvanic Membrane
The cell is cylindrical in construction, the anode is so constructed as to give the maximum surface area to improve response.(the type 3 electrode –used for sewage applications has an iron anode) The membrane may be of polyethylene, silicone rubber, PTFE, or any membrane showing good permeability to oxygen. In this cell the oxygen reacts with the metal anode and ceramic cathode generating a current, which is directly proportional to the partial pressure of the oxygen passing through the membrane. It is in effect an electrochemical cell. The electrode has a high temperature coefficient but a thermistor is built into the cell to give temperature compensation over a range from +2 to +40˚C. The signal is amplified and displayed either as concentration or percentage saturation.
Does the Sensing Technology Matter?
• • • Compare a good optical sensor and a good galvanic sensor and there is little to choose between them.
Optical has a shorter response time (5 times faster than galvanic) and can monitor in static water.
Optical Membrane caps can have a short life.
– 12 – 18 month enforced replacement for some.
Good Galvanic sensors will last 3+ years between refurbishments.
Good Optical and Galvanic sensors work • Choose a good sensor, the principal of operation is only part of the decision.
• Correct installation is key to success.
Maintenance is required and must be included in operating budgets • New monitors need to be quickly integrated into operator and technicians work programmes, not left until they fail.
• Or use the supplier/ICA contractor to look after the equipment.
• Consider who will do the calibration, cleaning and servicing of the instrument.
Product Information
Sensors – Dissolved Oxygen
OxyTechw 2 ™ OPT • Dissolved Oxygen & Temperature Sensor • Optical Luminescent Technology • Applications • Activated Sludge Control OxyTechw 2 ™ GAL • Dissolved Oxygen • Galvanic principle.
• Applications.
• Activated Sludge Control • Exceptional resistance to damage. Long term stability of 3 years without need for sensor replacement or membrane replacement
Fast response – Minimal Maintenance
• Provides self cleaning action on face of sensor • Prone to ragging and fats/grease
Sensors Installation & Cleaning
Floating Ball Rigid staff Flexible Shaft • Places sensor below the surface.
• Prone to ragging • Ragging falls off by moving sensor • Places the sensor below the surface • Provides self cleaning action on face of sensor
7300w
2
Monitor/Controller
Multiple Sensors • 1 or 2 into the base unit • Upto 8 using expansion boxes • Any w² sensor is compatible with the monitor Graphic Display with trending • Easy to configure • Multiple language ability Flexible control and monitoring options • Profibus output – anticipated late 2011 • Triple Validation option • Dosing control and datalogging available on request
Example of a Dissolved Oxygen Application
Example of connection layout for 7300w
2 (1 or 2 sensors)
Example of connection layout for 7300w
2 (1 to 8 sensors)
Portable Option
Galvanic Sensor • Self Polarising and Self Temperature Compensated No warm up time, short response time • 90% of end value in less than 20 seconds Automatic Calibration and self check 1400 hours from one 9V alkaline battery • Approximately 2 years with 1 hour use per day Large easy to read graphical LCD display
Angus Fosten Sales and Marketing Director Mobile: 0776 846 5581 Email: [email protected]
Andrew Wallace Field Sales Engineer Mobile: 0777 815 0118 Email: [email protected]
Clive Teobald Sales Engineer Mobile: 0796 601 3446 Email: [email protected]
Derek Bourne Export Sales and Marketing Director Mobile: 0780 857 2573 Email: [email protected]
Jan Johnson Sales Administrator Email: [email protected]
Tel: +44 (0) 1726 879800 www.partech.co.uk