Aging Measurements on Triple-GEM detectors
operated with CF4-based gas mixtures
M. Alfonsi1, G. Bencivenni1, W. Bonivento2, A.Cardini2,
P. de Simone1, F. Murtas1, D. Pinci3,
M. Poli-Lener1, D. Raspino2 and B. Saitta2
1.
Laboratori Nazionali di Frascati - INFN, Frascati , Italy
2.
Sezione INFN di Cagliari – Cagliari, Italy
3.
Sezione INFN di Roma 1, Roma, Italy
Siena, 24th May 2004
OUTLINE
 Introduction
 Local Aging Tests (X-ray and hadrons at PSI)
 Global Aging at Casaccia (1.25 MeV g)
 Understanding of the aging process
 Conclusions
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
2
Triple-GEM Detector (1)
70 µm
140 µm
50 µm
A Gas Electron Multiplier (F.Sauli, NIM
A386 531 1997) is made of 50 m thick
kapton foil, copper clad on each side and
perforated by an high surface-density of
bi-conical channels.
conversion
and drift
By applying a potential difference (300 - 500 V)
between the two copper sides, an electric field
as high as 100 kV/cm is produced in the holes
acting as multiplication channels for gaseous
detectors.
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
amplification
transfer
3
Triple-GEM Detector (2)
A Triple-GEM detector is built
by inserting three GEM foils
between two planar electrodes,
which act as the drift cathode
and the anode pad.
Conversion & Drift
The volume between the
cathode and the first GEM
(DRIFT gap) is the sensitive
gap in the detector.
Ionisation electrons drift through the four gaps towards the anode, they
are multiplied passing through the three GEM foils and they induce signal
on the readout PCB as they leave the third GEM (i.e. in the INDUCTION
gap)
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
4
GEM detectors in LHCb
In the innermost region of the first muon station (M1R1) of LHCb, the
highest particle rate region of the Muon Subsystem, the experiment
requires:
Rate Capability
Station Efficiency
Cluster Size
Radiation Hardness
Chamber active area
up to 0.5 MHz/cm2
>96% in a 20 ns time window (*)
<1.2 for a 10x25 mm2 pad size
1.8 C/cm2 in 10 years (**)
20x24 cm2
(*) A station is made of two detectors “in OR”. This improves time resolution and provides
some redundancy
(**) Estimated with 50 e-/particle at 184 kHz/cm2 with a gain of ~ 6000.
The value quoted takes into account for both the charge integrated on the third GEM
foil (1/2xQ) and the anode PADs (1/2xQ)
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
5
Why CF4-based gas mixtures?
The intrinsic time spread of a GEM detector is s(t) = 1/nvdrift , where n is the
number of primary clusters per unit length and vdrift is the electron drift
velocity in the ionization gap.
9.7 ns
4.5 ns
5.3 ns
4.5 ns
Considerable improvement with respect
to the Ar/CO2=70/30 gas mixture,
which exhibits a poor time resolution of
about 10 ns RMS, is obtained with the
new CF4 and iso-C4H10 based gas
mixtures, which allow time resolutions
better than 5 ns RMS
Our Choice:
Ar/CO2/CF4 45/15/40
Fast & Non-flammable
Single Chamber Time Spectra
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
6
Aging measurements: summary
 Local Aging:
performed with a high intensity 5.9 keV X-ray tube, irradiated area of about 1
mm2 (about 50 GEM holes). Integrated charge 4 C/cm2  25 LHCb years.
 Large Area Aging:
performed by means of the PSI M1 positive hadron beam, with an intensity up
to 300 MHz and an irradiated area of about 15 cm2. Integrated charge 0.5
C/cm2  3 LHCb years.
 Global Aging:
performed at Casaccia with a 25 kCi 60Co source. Detectors were irradiated at
0.5  16 Gray/h. Integrated charge up to 2 C/cm2  12.5 LHCb years.
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
7
Local Aging Measurements
performed with an high intensity X-ray beam on 1 mm2 (about 50 GEM holes) of
small size (active area 10x10 cm2) prototypes of the detector
Ar/CO2/CF4 (60/20/20)
G = 2.5·104
I ~ 270 nA
Ar/CO2/CF4 (45/15/40)
G = 6·103
I ~ 160nA
Ar/CF4/C4H10 (68/25/7)
G = 104
I ~ 160nA
Φgas=166.6 cc/min
Φgas=200 cc/min
10 LHCb
Years
6.6 C/cm2
10 LHCb
Years
1.8 C/cm2
10 LHCb
Years
3 C/cm2
G/G < 5%
G/G < 5%
G/G ~ 10%
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
Φgas=154 cc/min
8
Large Area Aging with hadron beam
At PSI (M1) we exposed small size prototypes of detectors to a high intensity
hadron beam (up to ~300 MHz on a spot size of about 15 cm2)
The beam was composed by low momentum (350 MeV/c) positive pions + ~7%
proton contamination
Gas mixture used: Ar/CO2/CF4 (60/20/20), Ar/CO2/CF4 (45/15/40),
Ar/CF4/iso-C4H10 (65/28/7), with a Φgas=200 cc/min
During the test we measured an average current on the anode of ~10 µA;
With the chosen gas mixture, Ar/CO2/CF4 (45/15/40), detectors integrated a
charge of about 0.5 C/cm2, equilavent to about 3 years of operation at LHCb
New measurements have been performed on these detectors at T11/PS at CERN:
NO LOSSES in performances (time resolution, working region width..) have been
observed
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
9
Global Aging Test at ENEA-Casaccia: Set-up
 A full size (20x24 cm2 active area) prototype (C)
in low irradiation position ~ 0.5 Gy/hour,
TWO full size prototypes (A,B) and a smaller
(10x10 cm2 active area) prototype (D) in high
irradiation position ~ 16 Gy/hour,
Gas flow
C -> B -> A -> D
 Gas mixture: Ar/CO2/CF4 = 45/15/40,
at reference Gain ~ 6x103
 Gas flows: C -> B -> A -> T/H2O Probe -> D -> Out
initially Φgas=200 cc/min, then Φgas=350 cc/min
Chamber
A,B,D
4.5m
 Monitored H2O (± 1ppm), T (± 0.1°K) in the gas
line, and external P (± 0.1mbar)
Chamber C
60Co
25 kCi
 Stainless steel Gas pipes used, except for the
exhaust and a short insertion to chamber inlet/outlet where polipropilene is used
(proved to be radiation-hard and not hygroscopic).
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
10
Gas mixture pollution
Normalized Pad Current (a.u.)
Gas flow
B
OFF
OFF
A
D
i-UP
During the whole test,
high-irradiated chambers
suffered a problem of gas
mixture pollution due to the
impossibility to increase the
gas flow over 350cc/min.
This was evident in the
downstream chamber D,
switching the upstream
chambers B and A off.
This effect causes a
SYSTEMATIC DAMAGE on
high-irradiated chambers!
Time (days)
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
11
Two clearly different trends! (1)
Normalized Currents (%)
11 LHCb years
PSI
Casaccia / Big Detector A / 16 Gy/h
Casaccia / Big Detector B / 16 Gy/h
Casaccia / Big Detector C / 0.5 Gy/h
Casaccia / Small Detector D / 16 Gy/h
X-Ray / Small Detector / Local Aging
High-irradiated chambers
exhibit a drastic current
drop during the test:
A  -89% after 11.5 y
B  -80% after 8.5 y
H2O injection
CO2 problem
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
12
Two clearly different trends! (2)
Normalized Currents (%)
The first 1.25 LHCb years
Casaccia / Big Detector A / 16 Gy/h
Casaccia / Big Detector B / 16 Gy/h
Casaccia / Big Detector C / 0.5 Gy/h
Casaccia / Small Detector D / 16 Gy/h
X-Ray / Small Detector / Local Aging
Chamber C as well as
X-ray and PSI test results
show no current drops after
an integrated charge of
0.15 C/cm2
G/G ~ -10%
after the same integrated
charge !
Integration time: 3  35 days
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
13
Preliminary Conclusions
 A current drop was observed at Casaccia … but this was not
observed in previous local X-ray (>10y) and PSI tests (~3y).
 This systematic effect could be due to the fact that for the heavily
irradiated detectors the gas flow was not increased proportionally with
the irradiation rate
 Nevertheless all tested chambers were still working after the
irradiation
 Several tests and checks to understand the aging process
have been performed:
 uniformity measurement
 gain, charging-up, rate capability measurements with X-rays
 performance of aged chambers at beam test
 electron microscope analysis (SEM) of aged chambers
 reproducing the low gas flow effect observed at Casaccia
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
14
Measurements on aged chambers:
Pad current uniformity
A
B
Uniformity 12%
Uniformity 22%
A remind ...
the chamber A was downstream in the gas line ...
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
15
Gain and Rate Capability
Gain
Gain @ 1280 V
A
B
6.103
6.103
-89% (11.5 Y)
-80% (8.5 Y)
-55%
-32%
Reference Xray LNF
Casaccia drop @ 15-20
MHz/cm2
X-ray
1.6 MHz/cm2
Rate Capability
Pad current (nA)
Chamber A current drop (~89%)
≈ gain drop (~55%)
+ rate cap. loss at 15MHz/cm2 (~30%)
BUT no losses at the LHCb
rate!
M. Alfonsi, LNF-INFN
Casaccia B
Siena, 24th May 2004
Casaccia A
~50%
~30%
XRay
LHCb
Rate (Mhz/cm2)
16
Test beam results on aged chamber
Before irradiation, performances of these detectors were measured at BTFLNF Facility (500 MeV electron beam) [see P.Valente presentation]
Chamber A
OR station
20ns efficiency
.
New measurements has been performed after aging, at T11-PS at CERN(*):
∑VGEM (V)
∑VGEM (V)
.
Chamber A (with the larger gain loss) shows only a shift of the working point of
~ 15V, without affecting working region width.
This shift for chamber B is negligible, so performances of the “OR station” is
practically unaffected.
(*) Different T, p of the two tests has been taken into account
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
17
SEM analysis on aged chambers (1)
Gem 3 down sampling
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
18
SEM analysis on aged chambers (2)
- No damage on gold plated drift cathode and Pad PCB.
They are perfectly clean.
- Fluorine found only on the bottom side of G2 and G3; could be present as
some Cu-F compound, forming a thin insulating layer
(no carbon deposits observed on the surfaces).
G1: No fluorine No etching
Ch A G1 down sample 5
 Inner → 44 µm
 Outer → 76 µm
M. Alfonsi, LNF-INFN
G2:Small fluorine,etching started
Ch A G2 down sample 9
 Inner → 47 µm
 Outer → 79 µm
Siena, 24th May 2004
G3:Large fluorine etching enhanced
Ch A G3 down sample 11
 Inner → 64÷65 µm
 Outer → 77÷78 µm
19
SEM analysis on aged chambers (3)
The fluorine strongly etched the third GEM, not only widening the
copper holes but also etching the kapton inside hole (from the bottom to
the top), changing the hole shape. The effective inner hole diameter,
from the standard 45-50 m becomes 60-65 m.
Cross section of the first GEM foil
of the aged chamber A.
M. Alfonsi, LNF-INFN
Cross section of the third GEM foil
of the aged chamber A.
Siena, 24th May 2004
20
3rd GEM X-Ray Spectroscopy
Fluorine (F- ions) was found only on the bottom side of the GEM foil
and not on the upper side
Cu-F compound layer around the hole,
forming a thin insulating layer
A strong Kapton etching inside the hole has
been observed mainly on the 3rd Gem.
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
21
Fluorine etching explains observed effects
The effects of fluorine etching is twofold:
 Widening of copper holes
gain reduction (*)
 Kapton etching, especially on the bottom part of the hole, and Cu-F
compound forming an insulating layer near the hole
Enhanced charging-up effects
rate capability reduction
(*) S.Bachmann et al., NIM A 438(1999),376-408
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
22
Aging induced by low gas flow
We tried to reproduce the Casaccia test results, irradiating with X-rays
a 10x10 cm2 chamber (total current ~2 A on ~ 1 cm2 irradiation spot)
flushed with a low gas flow (20 cc/min).
A current drop of ~ 40%
for a 0.55 C/cm2 integrated
charge (~3 LHCb years) is
found on the low gas flow
measurement.
High gas flow
Low gas flow
NO current drop is observed
on the high gas flow
measurement.
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
23
Conclusions
 Casaccia results seem to be understood :
the etching observed is correlated with bad gas
flow rate condition;
 No corrosion effect observed on the cathode and anode;
 Etching effects mainly on the third gem with
fluorine deposits near the copper holes edges;
 No aging occur if the gas flow is properly set,
as in LHCb running condition;
 Detectors, even after a severe irradiation in bad conditions,
exhibit good time and efficiency performances.
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
24
Other slides...
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
25
Measurements on aged chambers:
Pad current uniformity [1]
Chamber A (nA/pad)
Chamber B (nA/pad)
Good uniformity on the chambers
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
26
Measurements on aged chambers:
effective gain
X-ray beam measurements show that
aged chambers has a gain lower than
the gain of a new chamber, but not so
low as expected from the current drop
during the test at ENEA-Casaccia
Gain @ 1280 V
Reference Xray LNF
Casaccia drop @ 15-20
MHz/cm2
X-ray
1.6 MHz/cm2
A
B
6.103
6.103
-89%
(11.5 Y)
-80%
(8.5 Y)
-55%
-32%
But during the measurement photon
rate is lower than the one of aging test.
In fact..
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
27
Measurements on aged chambers:
Rate Capability
Charging-up is a typical effect of micro-pattern
detectors using insulating material as support.
Casaccia B
In a new GEM detector, charging-up of kapton
inside GEM holes causes an increase of gain
Casaccia A
that last for few tens of minutes from start-up.
~50% On the contrary, aged detectors exhibit a
decrease of gain (a modification of the internal
kapton profile of the GEM holes?)
Pad current (nA)
Rate Capability
~30%
XRay
Thus, losses in rate capability explain the
current drop observed during aging test
LHCb
Rate (Mhz/cm2)
Chamber A current drop (~89%) 
gain drop (~55%) + rate capability (~30%)
BUT there are no losses at the LHCb rate!
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
28
Charging-up before and after aging
Chamber A before aging..
..and after aging
Time (min)
Time (s)
M. Alfonsi, LNF-INFN
Siena, 24th May 2004
29