Transcript Document

X Rays
Medical Physics Notes
Ideal X-Ray Examination
• a film that showed sufficient
contrast between the features that
the doctor wanted to examine
• while putting the patient at minimal
risk from the ionizing effect of the
radiation.
The Beam
• An X-ray tube does not produce a
monochromatic beam
• It produces a spectrum of X-ray
energies limited at the high energy
end by the accelerating voltage
applied.
• With characteristic peaks relating to
the material of the target
Attenuation (reduction of the
beam strength)
• Attenuation occurs as the Xrays pass through matter.
• This attenuation is
exponential.
– Let Io = Intensity of the
incident beam
– I = Intensity of emerging
beam
– x = the thickness of the
material the beam travels
through
 m = linear attenuation
coefficient
Syllabus Extract
• Exponential attenuation Linear
coefficient m, mass attenuation
coefficient mm and half-value thickness x
• Remember r is density!!
• You did something similar with gamma
ray absorption
Half Thickness
I = Ioe-mx
I/Io = e-mx
natural log of (I/Io) = -mx
If the intensity halves
I = 0.5 Io
ln 0.5 = -mx
ln 0.5 = -ln2
x = ln2/m
Half Thickness
• This is similar to half life – the
thickness of material that the rays
must pass through in order for the
intensity of the beam to be cut in
half.
• This varies with density of material
– the denser the material the smaller
the half thickness.
Half Thickness
• Ensure you can calculate half thicknesses as well
as find them off graphs.
• In a similar way you can find out the thickness
needed to reduce penetration to a tenth etc. ( just
put I = 0.1 Io into the equation).
• Values for the mass attenuation coefficient mm can
be changed into m by using the equation
mm = m / r
Where r is the density of the material
Attenuation mechanisms
• The lower energy rays are more likely to
be attenuated by the body than the high
energy ones.
• Attenuation occurs as the radiation passes
through the body of the patient by two
principal mechanisms:
– photoelectric absorption and
– Compton scattering.
Syllabus Extract
• Differential tissue absorption of X-
rays (excluding details of the
absorption processes)
• Note that you do NOT need details
of the attenuation processes!
• But you do need to know the names!
Attenuation
• The lower energy rays are more
likely to be attenuated by the body
than the high energy ones.
Attenuation occurs as the radiation
passes through the body of the
patient by two principal mechanisms:
photoelectric absorption and
Compton scattering.
Attenuation Processes _ Detail
NOT required
• Photoelectric absorption occurs when
a photon of energy is absorbed by an
orbital electron and this electron is
then promoted to a higher energy
level (more outer orbit) or leaves the
influence of the nucleus completely
(ionization).
Attenuation Processes – Detail
NOT required
• A.H. Compton discovered that if he
bombarded graphite with
monochromatic X-rays, the scattered
X-rays had lower energies (longer
wavelengths) than the undeflected
ones: the greater the deflection the
bigger the energy loss.
Attenuation Processes – Detail
NOT required
• The bombarding X-ray photon has a
lot of energy - the force binding the
electron to the atom is insignificant
compared to the force exerted by
the photon on impact.
• When the photon 'bounces off' the
electron, the electron recoils and
thereby picks up some of the
photon's energy. This is called
Compton Scattering.
Attenuation Processes – need
to know!
• Photoelectric absorption is the
dominant mechanism for low energy
X-ray photons (used in soft tissue)
whereas Compton Scattering becomes
more significant for higher energy
photons (bone).
Contrast in Breast Tissue
• Low energy photon energies produce a
better contrast between media of
similar density (as in mammography)
but overall absorption is greater.
• This means that a higher anode
current (resulting in a more intense
beam) has to be used the lower the
accelerating potential employed
across the tube.
Contrast in Chest XRay
• In a chest X-ray the densities of
tissue to be investigated is diverse
(bone/lung/heart) and ‘harder’ X-rays
can be employed.
• These still give the contrast required
in the image but absorption is reduced
by using high energy rays and filtering
out the lower energy ones (soft X-rays)
produced by the tube.
Contrast in Chest XRay
• This can be done using an aluminium
filter. Suitable energy for a chest Xray would be 60-100 keV depending
upon the exact nature of the detail
required to make the diagnosis.
Collective Dose
• When calculating the collective dose to the
population the average dose received per person is
multiplied by the number of persons.
Average Dose
• To calculate the average dose received because of
X-ray examinations the number of each type of
investigation would be found and then the typical
dose given for each procedure would become the
multiplying factor.
Typical Doses
X-ray examinations of
• limbs, joints and teeth involve a typical
effective dose of about 0.01 mSv whereas
a
• chest CT scan involves 8.0 mSv. and a
• barium enema 7.2 mSv.
Hence one CT scan is equivalent in dose to
about 800 knee X-rays!!
Typical Doses
• This is why although many more low
dose X-rays are carried out, they do
not contribute very much to the
population dose.
• The much lower number of major
scans make a significant contribution
to population dose because they
individually are equivalent to a vast
number of low dose investigations.
Effective Dose
• The effective dose of each
procedure varies because dose
depends on:
– X-ray intensity,
– energy and
– application time.
Real Time Investigations
• A real time investigation such as
Barium meal involves the patient
being bathed in X-rays as the doctor
watches an image on a TV monitor.
• The dose is minimized by pulse
application and image freezing but
necessarily involves a much bigger
dose than a simple ‘snapshot’ method
as used in a chest X-ray.
Real Time Investigations
• The dose varies in its effect on
tissue too as this is dependent
upon the quantity of cell division
taking place and summed
absorption of layers of tissue.