Protons - Superior Dose Distribution
The Bragg Peak is representative of a delivery of a full, localized, and uniform dose of energy in the target 4 3 2 1
60
0 10 20 Depth in Tissue (cm)
Experience, Expertise and a Commitment to Excellence™
30
System Components
The Synchrotron Accelerator The Beam Transport System The Stationary Beam Gantries (Isocentric Gantries) Patient Alignment and Digital Imaging Systems Facility Control System
Experience, Expertise and a Commitment to Excellence™
Facility Layout
The Loma Linda University Medical Center Proton Treatment Center
The Gantries:
Has two branches: •Eye Tumors •Head and Neck Resembling giant ferris wheels can rotate around the patient and direct the proton beam to a precise point. Each gantry weights about 90 tons and Tumors stands 3 stories tall. It supports the bending and focusing magnets to direct the beam.
The Injector:
Protons are stripped out of the nucleus of hydrogen atoms and sent to the accelerator.
Steel reinforced concrete walls are up to 15 feet thick.
Synchrotron: (Accelerator)
A ring of magnets, 20 ft. in diameter, through which protons circulate in a vacuum tube.
Beam Transport System:
Carries the beam from the accelerator to one of four treatment rooms. This system consists of several bending corners and focus it to the desired location .
Proton Accelerator Control System
Conforma 3000 Integrated Solution
Treatment Control System (TCS) Accelerator Control System (ACS) Treatment Workstation Integrated and Automated “The Brain” Software & Electronics for Treatment Precision Accelerate Deliver Control
Experience, Expertise and a Commitment to Excellence™
Accelerator Control - Advantages
Automatic Beam Centering in Treatment Nozzle Key to Active Beam Scanning Rapid Beam Line Switching <10 sec.
Accurate, Reproducible Beam Energy Within 0.1 MeV Patented Safety Systems Ensure Beam is Delivered at the Right Place & Time
Experience, Expertise and a Commitment to Excellence™
Integrated and automated Rotates >360 o Treatment from multiple “portals” or angles Isocenter is contained within a sphere to 1.0 mm in radius Automated set up Designed for functionality
Gantry Treatment Room
Experience, Expertise and a Commitment to Excellence™
Conforma 3000 Passive Scattering Beam Delivery System Compensator (Bolus) and Block (Aperture) Dose and Flatness Monitor(s) Range Modulation Wheel X-ray Tube Second Scatterer First Scatterer Profile Monitor Beam Movable Snout (cones)
Experience, Expertise and a Commitment to Excellence™
Redundant Dose Detectors Optional Compensator (Bolus) and Block (Aperture) Movable Snout (cones) Conforma 3000 Active Beam System Intensity Modulated Proton Therapy IMPT or Active Beam Scanning (ABS) Supports Compensator-less treatments Variable energy controls penetration depth Capable of continuous (raster scanning) & discrete point based (“Spot” scanning) Scanning Magnets Radiographic Imaging Panels (2)
Experience, Expertise and a Commitment to Excellence™
Beam
Intensity Modulated Proton Therapy
Ability to Treat Complex Volumes & Multiple Target Areas “Spot Scanning” Technique Beam scans from the top down Radiation deposited from side to side
Experience, Expertise and a Commitment to Excellence™
Modular Design (4 gantry, 2 Stationary)
Experience, Expertise and a Commitment to Excellence™
Treatment Planning Software
Experience, Expertise and a Commitment to Excellence™