Transcript Document
pH / ORP Conductivity Dissolved Oxygen Chlorine Standards CONTROLS Concerns in pH # 6: High Purity pH www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards PROBLEM High purity water has a very high impedance. The high impedance of the water combined with the high impedance of the glass measuring electrode makes the pH signal susceptible to electrostatic interference and sluggish response. www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards Solution Model 615-25 high purity pH panel or model 615-26 ultra-high purity (less than 1 µScm-1) panel www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards pH Tip Configuration Reference Flat Glass Stem Reference Junction SS Flow cell www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards pH Tip Configuration •Flat surface reduces path length between glass and reference for reduced impedance •Flat surface allows small volume at electrode tip •Flat surface close to inlet allows fast sample response •Stainless Steel flow cell can be used for differential measurement technique and eliminate electrostatic pickup. •Use low conductivity buffers for quicker response after calibration www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards Short path length between the glass and the reference minimizes impedance and will speed response time. 1 mm separation www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards Small sample volume at the tip allows fast sample turnover. Also, connection between the glass and reference is accomplished by KCl that leaches into the sample from the reference. The smaller the sample volume, the better the connection between the glass and reference . 1 ml volume at sensor tip www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards Ultra-high purity applications are susceptible to electrostatic interference. The model 615-26 has a differential pre-amplifier which uses the stainless steel flow cell as a third electrode. The differential pre-amp measures the mV potential between the glass and flow cell and then the reference and flow cell and subtracts the two values. Thus, the electrostatic interference is eliminated from the measurement. mV measurement between glass and flow cell then reference and flow cell. www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards Pressure and Flow Fluctuating pressure and flow will have an effect on the stability of the pH measurement. As the KCl from the reference is the conductor, fluctuation of the pressure or flow will allow varying amounts of KCl to flow out of the reference. The varying flow will be read by the pH electrode as varying conductivity and the pH meter will show “jumpy” readings. Pressure and flow regulation is standard for both the 615-25 and 615-26 to create better pH stability. www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards Calibration When calibrating, it is important to use the proper standards for best accuracy and quick response. As the conductivity of the sample is low, if normal pH buffers are used they will be absorbed into the reference junction and take a long time to to purged. By using low conductivity buffers, excellent accuracy and speed of response will be accomplished after calibration. www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards Calibration and Response Time 11 10 pH 9 8 Standard Buffer 7 Low Conductivity Buffer 6 5 4 0 www.iccontrols.com 30 60 90 Time (Minutes) 180 IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards IC Controls High Purity pH Low sample volume Short path length between the glass and reference Differential measurement in ultra-pure applications Pressure and flow regulation Low conductivity buffers available Proven Technology Superior Design Better Results www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards 615 Sensor and Flow cell For more information: Call: 519-941-8161 800-265-9161 Email: [email protected] or www.iccontrols.com We will be happy to help you directly or put you in contact with your local Representative of IC Controls Qualitiy Water Products 615-26 www.iccontrols.com IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS