Transcript Standards & Guidelines for Creating Presentations using
3M™ Attest™ Sterile U Web Meeting March 19, 2009
How Long Did They Go? Healthcare Facilities Share Flash Reduction Best Practices
© 3M 2009. All Rights Reserved.
Welcome!
Facilitator:
Tammy Torbert, 3M Sterilization Assurance
Speaker:
Dorothy Larson, 3M Technical Service
Panelist
: Francis Zieman, 3M Technical Service
Housekeeping:
Questions; -Mute feature (*7=unmute; *6=mute) -“Chat” feature Technical difficulties CE Credits Post session follow-up
For more information:
www.3M.com/AttestSterileUOnline
2
© 3M 2009. All Rights Reserved.
Discussion Topics
1.
2.
3.
4.
5.
Define flash sterilization Discuss the Association of periOperative Registered Nurses (AORN) and the Association for the Advancement of Medical Instrumentation’s (AAMI’s) recommended practice on flash sterilization Discuss issues associated with flash sterilization Describe an effective quality assurance program for flash sterilization Discuss Flash Reduction Best Practices
3
© 3M 2009. All Rights Reserved.
Quality Control Recommended Practices
ANSI/AAMI Comprehensive guide to steam sterilization and sterility assurance in health care facilities, ST79:2006/A1:2008
AORN
– Recommended Practices for Sterilization in
Perioperative Practice Setting, 2009
– Recommended Practices for Selection and Use of
Packaging Systems for Sterilization, 2009
– Recommended Practices for Surgical Instruments
and Powered Equipment, 2009
4
© 3M 2009. All Rights Reserved.
AAMI and AORN
AAMI
Guidelines for the use, care and/or processing of medical device or system to ensure a device is used safely and effectively and its performance will be maintained
AORN
Standards and recommended practices serve as the basis for policies and procedures
Promote safety and optimal outcomes for patients
5
© 3M 2009. All Rights Reserved.
Flash Sterilization
Process designed for the steam sterilization of patient care items for immediate use.
6
AAMI ST79 © 3M 2009. All Rights Reserved.
Flash Sterilization - History
Reprocessing of dropped instruments mid-procedure Unwrapped, 270 deg. F. Gravity cycle only -No dry time High Temperature Sterilization
7
AAMI ST79 © 3M 2009. All Rights Reserved.
Flash Sterilization - Today
High temperature (270-275
° F/ 132-135 ° C)
No dry time/ No storage Sterilization Process: - Gravity - Dynamic-air-removal Packaging expanded beyond mesh-bottom surgical tray Photo courtesy of Rose Seavey, The Children's Hospital - Denver
8
AAMI ST79 © 3M 2009. All Rights Reserved.
Flash Sterilization - Packaging
Perforated, mesh bottom, open surgical tray Rigid sterilization container system Protective organizing case Single-wrapped surgical tray
9
AAMI ST79 © 3M 2009. All Rights Reserved.
Flash Sterilization - Packaging
AORN recommends the use of rigid sterilization containers
Reduce risk of contamination during transport to point of use
Ease of presentation to sterile field
10
AORN © 3M 2009. All Rights Reserved.
Types of Steam Sterilization Processes
Gravity Displacement Dynamic-Air-Removal by
Prevaccuum
Steam-flush pressure-pulse (SFPP)
11
AAMI ST79 © 3M 2009. All Rights Reserved.
Flash Sterilization
Minimum Flash Cycle Times for Gravity Steam Sterilization Cycles
Load Contents
Unwrapped nonporous items Unwrapped nonporous & porous items in mixed load
Temp
270 ° F/132 ° C 275 ° F/135 ° C 270 ° F/132 ° C 275 ° F/135 ° C
Time(Min)
3 3 10 10
12
AAMI ST79 © 3M 2009. All Rights Reserved.
Flash Sterilization
Minimum Flash Cycle Times for Dynamic-Air-Removal Steam Sterilization Cycles
Load Contents
Unwrapped nonporous items Unwrapped nonporous & porous items in mixed load
Temp
270 ° F/132 ° C 275 ° F/135 ° C 270 ° F/132 ° C 275 ° F/135 ° C
Time(Min)
3 3 4 3
13
AAMI ST79 © 3M 2009. All Rights Reserved.
QUESTION
: Review information in following table –
is this a flash cycle?
Sterilization Temperature Sterilization Time Sterilization Method Packaging Type Dry Time 270 deg. F.
4-minute exposure Prevacuum Rigid Container None 14
© 3M 2009. All Rights Reserved.
QUESTION:
Review information in following table – is this a flash cycle?
Sterilization Temperature 270 deg. F.
Sterilization Time Sterilization Method Packaging Type Dry Time 4-minute exposure Prevacuum Rigid Container None ANSWER: YES!
Remember, flash sterilization is the steam sterilization of patient care items for immediate use !
15
© 3M 2009. All Rights Reserved.
ANSI/AAMI ST79 Section 10 Quality Control
Four levels of testing Routine load release Testing of each non-implant and implant load Routine sterilizer efficacy monitoring Establishing a regular pattern of testing the efficacy of the sterilization process Sterilizer qualification testing Testing of the sterilizer after events occur which could affect the ability of the sterilizer to perform Periodic product testing Testing of routinely processed items to ensure the effectiveness of the sterilization process and to avoid wet packs
16
© 3M 2009. All Rights Reserved.
Routine Load Release Nonimplants
Physical monitors External process indicator (Class 1) on every package Internal CI (Class 3, 4 or 5) inside every package
Optional monitoring of load
with a PCD containing a
– BI – BI and a Class 5 CI – Class 5 integrating indicator
Evaluation of all data by an experienced, knowledgeable person
Do not distribute load if any data suggests a sterilization process failure
17
AAMI ST79 © 3M 2009. All Rights Reserved.
Routine Load Release Implants
Physical monitors External process indicator (Class 1) on every package Internal CI (Class 3, 4 or 5) inside every package
A PCD containing a BI and a Class 5 integrating indicator
Evaluation of all data by an experienced, knowledgeable person Do not distribute load if any data suggests a sterilization process failure
18
AAMI ST79 © 3M 2009. All Rights Reserved.
Sterilization Process Monitoring Tools
Physical Monitors Chemical Indicators Biological Indicators
19
© 3M 2009. All Rights Reserved.
Physical Monitors
Verify that parameters of sterilization cycle are met
Recording charts
Gauges Tape
Printouts Digital displays
20
AORN & AAMI © 3M 2009. All Rights Reserved.
ANSI/AAMI/ISO 11140-1, 2005 Sterilization of health care products-Chemical indicators - Part 1: General requirements
Class 1
Process Indicators
Class 2
Indicators for use in Specific Tests
Class 3
Single Variable Indicators
Class 4
Multi-variable Indicators
Class 5
Integrating Indicators
Class 6
Emulating Indicators
21
© 3M 2009. All Rights Reserved.
Chemical Indicators
Class 1: Process Indicators
Use with individual units to indicate that the unit has been directly exposed to the sterilization process
– Exposure Control (e.g., indicator tapes) – Distinguish between processed and unprocessed units
Designed to react to one or more the critical process variables of
22
AAMI ST79 © 3M 2009. All Rights Reserved.
Chemical Indicators
Class 2: Indicators for Specific Tests Equipment Control Testing sterilizer performance
Bowie-Dick Test monitors efficacy of air removal and steam penetration in 270-275 sterilizers) ° F dynamic-air removal sterilizers (i.e., vacuum assisted
23
AAMI ST79 © 3M 2009. All Rights Reserved.
Chemical Indicators
Internal CIs – Class 3, 4, 5
Tests conditions inside individual packs
Internal chemical indicator inside each package Use inside each pack, wrapped tray, flash tray or container in area least accessible to sterilant – identifies sterilant penetrated
24
AORN & AAMI © 3M 2009. All Rights Reserved.
AORN
Placement of internal chemical indicators
25
AORN © 3M 2009. All Rights Reserved.
Chemical Indicator Placement
Rigid container
Place two CIs inside rigid containers
Place one in each of two opposite corners
26
AORN © 3M 2009. All Rights Reserved.
Chemical Indicator Placement
Multi-layer container
Place two CIs in each level of multi-level rigid container
Place one in each of two opposite corners on each level
27
AORN © 3M 2009. All Rights Reserved.
Chemical Indicator Placement
Multi-level container
Supplied by the manufacturer, holes in tray, has to be wrapped Place a CI in center on each level
28
AORN © 3M 2009. All Rights Reserved.
Chemical Indicator Placement
Wrapped instrument trays
Place CI in geometric center, not on the top
29
AORN © 3M 2009. All Rights Reserved.
Routine Load Release Nonimplants – Flash Sterilization
Class 5 Integrating Indicator PCD for releasing flash loads
PCD must be representative of load
– – – – Open surgical tray Rigid sterilization container Protective organization case Single-wrapped surgical tray
Class 5 Integrating Indicator also serves as internal CI
30
AAMI ST79 © 3M 2009. All Rights Reserved.
Biological Indicators
Definition Test system containing viable microorganisms providing a defined resistance to a specified sterilization process
31
AAMI ST79 © 3M 2009. All Rights Reserved.
3M™ Attest™ Rapid Readout Biological Indicators - Flash Cycles 1291, Blue Cap BI
270 ° F/132 ° C, gravity
1292, Brown Cap BI
270 ° F/132 ° C, dynamic-air-removal
32
© 3M 2009. All Rights Reserved.
Biological Indicators AAMI ST79
AAMI and AORN - Weekly, preferably daily and implants
33
AAMI ST79 & AORN © 3M 2009. All Rights Reserved.
Routine Load Release Implant Loads
“Emergency situations should be defined”
34
AAMI ST79 © 3M 2009. All Rights Reserved.
Implant Loads – Flash Cycles
“In an emergency, when flash sterilization of an implant is unavoidable, a enzyme-only indicator) rapid-action BI and a Class 5 chemical integrating indicator (or should be run with the load. The implant should be quarantined on the back table and should not be released until the rapid-action BI provides a negative result.”
35
AORN © 3M 2009. All Rights Reserved.
Biological Indicators
“Rationale: The use of BIs provides evidence of efficacy by challenging the sterilizer with a large number of highly resistant bacterial spores. cycle.
Biological monitoring provides the only direct measure of the lethality of a sterilization Sterilizer manufacturers validate their sterilization cycles using BIs; therefore, routine sterilizer efficacy monitoring in health care facilities should also be conducted using BIs.” (section 10.5.3.2)
36
AAMI ST79 © 3M 2009. All Rights Reserved.
Record Keeping
“Documentation establishes accountability.”
AORN Sterilization Recommended Practice
37
AORN © 3M 2009. All Rights Reserved.
Implant Loads - Flash Cycles
“Flash sterilization should not be used for implantable devices except in cases of emergency when no other option is available.”
Photo courtesy of Rose Seavey, The Children's Hospital Denver
38
AORN © 3M 2009. All Rights Reserved.
Flash Sterilization - AORN Documentation
Log or data base to trace sterilized items used on patients should include: Information on each load
Item(s) processed
Patient receiving the item(s) Cycle parameters used (e.g., temperature, duration of cycle)
Day and time cycle is run Operator information Reason for flash sterilization
39
AORN © 3M 2009. All Rights Reserved.
Routine Sterilizer Efficacy Monitoring Sterilizer Qualification Testing
Routine Efficacy Testing
Establishing a regular pattern of testing the efficacy of the sterilization process
Qualification Testing
Testing of sterilizer after events occur that affect the ability of the sterilizer to perform
Sterilizer installation
Relocation Malfunction Major repairs of sterilizer or utilities Sterilization process failures
40
AAMI ST79 © 3M 2009. All Rights Reserved.
Flash sterilization cycles 41 Routine Efficacy Testing:
Representative BI PCD using one or more BIs and one or more CIs Perforated, mesh bottom, open surgical tray Rigid sterilization container system Protective organizing case Single-wrapped surgical tray Test each type of tray configuration – Must always test the open surgical tray configuration Place on bottom shelf over the drain in otherwise empty chamber Weekly, preferably daily After Bowie-Dick test if dynamic air-removal
Qualification Testing:
Representative BI PCD using one or more BIs and one or more CIs Perforated, mesh bottom, open surgical tray Rigid sterilization container system Protective organizing case Single-wrapped surgical tray Select one representative tray Place on bottom shelf over the drain in
otherwise empty chamber
Three consecutive cycles Before removal Bowie-Dick test if dynamic air-
© 3M 2009. All Rights Reserved.
Process Challenge Device (PCD) Definition
“Item designed to constitute a defined resistance to a sterilization process and used to assess performance of the process.” Representative of load and creates the greatest challenge
Equal to or greater than the most difficult item to sterilize Placed in most difficult area in sterilizer for the sterilant to penetrate
42
AAMI ST79 © 3M 2009. All Rights Reserved.
Process Challenge Device (PCD)
User assembled Challenge test pack or tray (e.g., flash tray) *Note there are no commercially available – FDA Cleared PCDs available for flash sterilization
43
AAMI ST79 © 3M 2009. All Rights Reserved.
Quality Assurance Program
Sterilization Process Monitoring
Monitoring Tools
Physical monitors
Chemical indicators
Biological indicators
Process challenge devices (PCDs) Test or challenge pack Results – Integrate results of all monitoring controls; determine if an effective sterilization process
AAMI ST79
44
© 3M 2009. All Rights Reserved.
QUESTION: If a biological indicator (BI) is positive, do you only recall that load?
45
© 3M 2009. All Rights Reserved.
46 QUESTION: that load? If a BI is positive, do you only recall ANSWER: If determined to be an operator error
e.g.,.. using incorrect sterilization cycle
No recall, don’t use load
If not operator error or don’t know reason
Recall all items processed since last negative BI
Reprocess all retrieved items
© 3M 2009. All Rights Reserved.
Biological Indicator Monitoring Frequency
Why monitor every load?
Universal standard of patient care Reduces variability and chance for errors To be certain all implants, including those in loaners sets, are appropriately monitored To ensure all cycle types used are tested with a BI To ensure all types of packaging used in flash sterilization are tested with a BI To reduce the risk, cost, and impact of a recall To reduce the risk/cost of healthcare-associated infections (HIAs)
47
© 3M 2009. All Rights Reserved.
The Joint Commission Organizational policies/procedures should be based on most stringent:
Laws and regulations
Accepted practice guidelines Current scientific knowledge And, are consistent throughout facility
48
© 3M 2009. All Rights Reserved.
Issues Associated with Flash Sterilization
Time Pressures
– skipped steps in cleaning/decontamination
Flashing large and/or multiple trays
– increases cycle and cooling time
Loaner Instrumentation
49
AAMI ST79 © 3M 2009. All Rights Reserved.
Loaner Instrumentation Issues
Patient Safety Timelines (flashing is not recommended) Communication (OR, SPD, Vendor) Quality
– MDM Reprocessing Guidelines – Adequate time – Implants
Potential for lost items
50
© 3M 2009. All Rights Reserved.
Toxic Anterior Segment Syndrome - TASS
The Issue: serious damage to a patient’s intraocular tissue and vision loss as a result of contaminants in the eye during ophthalmic surgery Causes: contaminated irrigating fluids; antiseptics; antibiotic ointments; powder from surgical gloves
“Most cases of TASS appear to result from inadequate instrument cleaning and sterilization”
Nick Mamalis, MD, Toxic Anterior Segment Syndrome Journal of Cataract and Refractive Surgery 2006; 32:324-333.
51
© 3M 2009. All Rights Reserved.
Toxic Anterior Segment Syndrome - TASS
Inadequate or inappropriate instrument cleaning “Detergents Heat stable endotoxin from overgrowth of gram negative bacilli in water baths or ultrasonic cleaners Degradation of brass containing surgical instruments from plasma gas sterilization Impurities of autoclave steam”
52 Recommended practices for cleaning and sterilizing intraocular surgical instruments, J Cataract Refract Surg, Vol 33, June 2007
© 3M 2009. All Rights Reserved.
53 AAMI ST79 2008 Amendments - Annex N (informative)
Toxic anterior segment syndrome (TASS) and the processing of surgical instruments
Follow Mfr’s. cleaning and sterilization instructions Adequate inventory-time for processing Designated cleaning area and dedicated equipment Precleaned immediately Transport in closed containers PPE Appropriate cleaning agent & water of appropriate quality as specified by the Mfr.
Sterilization according the Mfr’s. instructions Maintenance of cleaning and sterilization equipment, boilers and water filtration systems Training © 3M 2009. All Rights Reserved.
Thank You!
© 3M 2009. All Rights Reserved.
SHARING FLASH REDUCTION “BEST PRACTICES”
55
© 3M 2009. All Rights Reserved.