Transcript Slide 1
Carbon Meeting Joann Rice Sacramento, CA January 22-23, 2008 “Old” versus “New” CSN Carbon “Old” CSN MetOne SASS sampler 6.7 LPM flow rate with active flow control 46.2-mm QMA Quartz filter CSN TOT analysis by RTI OC in He (900C); EC in He/O2 (920C) “New” CSN URG3000N sampler (IMPROVE-like) 22 LPM flow rate with active flow control 25-mm Pallflex Quartz filter IMPROVE_A TOR analysis by DRI OC in He (580C); EC in He/O2 (840C) SOPs posted at http://www.epa.gov/ttn/amtic/specsop.html “Old” versus “New” CSN Carbon First Look Two months (May-June 2007) of collocated old and new CSN at all Phase 1 sites Collocation was resource restricted For purposes of overlap only to minimize data loss About 10-20 collocated observations per site No artifact adjustments made to either old or new CSN data No data screening for MDL or other outliers “Old” versus “New” CSN OC “Old” versus “New” CSN OC 70 95% CI Notched Outlier Boxplot 60 95% CI Mean Diamond Outliers > 1.5 and < 3 IQR 50 Outliers > 3 IQR 40 30 20 10 0 88305 OC CSN TOT 88370 OC TOR Unadj n 88305 OC CSN TOT 88370 OC TOR Unadj Mean 522 520 95% CI 5.01 3.31 4.64 5.37 3.00 3.63 SE SD 0.19 0.16 4.25 3.68 “Old” versus “New” CSN EC “Old” versus “New” CSN EC 6 95% CI Notched Outlier Boxplot 95% CI Mean Diamond 5 Outliers > 1.5 and < 3 IQR 4 Outliers > 3 IQR 3 2 1 0 88307 EC CSN TOT 88380 EC TOR Unadj OP n 88307 EC CSN TOT 88380 EC TOR Unadj OP Mean 521 519 95% CI 0.67 0.81 0.63 0.71 0.76 0.85 SE SD 0.02 0.02 0.44 0.54 Collocated “Old” CSN OC Collocated “Old” CSN EC