RADIATION PROTECTION IN DIAGNOSTIC RADIOLOGY

Download Report

Transcript RADIATION PROTECTION IN DIAGNOSTIC RADIOLOGY

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
RADIATION PROTECTION IN
DIAGNOSTIC AND
INTERVENTIONAL RADIOLOGY
Part 15.3: Optimization of protection in
radiography
Practical exercise - HVL measurement
IAEA
International Atomic Energy Agency
Overview / Objectives
• Subject matter : HVL measurement
• Step by step procedure to be followed
to implement the considered QC test
• Interpretation of results
IAEA
15.3: Optimization of protection in radiography
2
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Part 15.3: Optimization of protection in
radiography
HVL (Filtration) Measurement
IAEA
International Atomic Energy Agency
Half Value Layer (HVL) (I)
• Possibly the most important test
• Checks whether there is sufficient filtration in
the x-ray beam to remove damaging low
energy radiation
IAEA
15.3: Optimization of protection in radiography
4
What Minimum Equipment is Needed?
• Electronic device (multi-function meter) to measure
•
•
•
•
dose
High purity aluminium layers. (Most aluminium has
high levels impurities, e.g., Cu)
Stand to hold the dosimeter, e.g., laboratory stand
Tape measure, spirit level
Roll of removable tape
IAEA
15.3: Optimization of protection in radiography
5
Tape Measure, Spirit Level
IAEA
15.3: Optimization of protection in radiography
6
Half Value Layer (II)
• Method (simplest - using a multifunction test
•
•
•
•
meter) :
Place the detector half way between the
front of the collimator and the table top.
Set 80 kVp, fixed mAs (e.g., 50 mAs)
Collimate x-ray beam to size of detector
Measure exposure three times
IAEA
15.3: Optimization of protection in radiography
7
X-Ray Tube
HVL Measurement Geometry
Collimator
Aluminium Sheets
X-Ray Beam
Detector
Table top
IAEA
15.3: Optimization of protection in radiography
8
Half Value Layer (III)
• Tape a 1 mm sheet of aluminium to the front
of the collimator and measure the dose
• Add another 1 mm sheet of aluminium,
measure again, and repeat until the dose
has fallen to below 50% of the initial,
unattenuated, value
• Remove all aluminium, and make 3 more
exposures
IAEA
15.3: Optimization of protection in radiography
9
Half Value Layer (IV)
Analysis :
• Average all the “no Al” measurements
• Plot all results on semi-log graph paper
• From the graph, find the thickness of Al required to
reduce the unattenuated dose by 50% - this is the
HVL
• The HVL must be at least that specified in local
regulations, e.g., 2.3 mm Al at 80 kVp
• Ideally, it should be at least 3.0 mm Al or higher to
assist in minimizing patient dose
IAEA
15.3: Optimization of protection in radiography
10
HVL Measurement
• Be careful of beam
hardening (semi-log plot is
not a straight line)
• Use points either side of
half initial value
• Calculate HVL :
Dose
10
(initial value = 9
50% of this = 4.5,
thus HVL = 2.6 mm Al)
1
0
1
2
3
4
mm Al
IAEA
15.3: Optimization of protection in radiography
11
Minimum HVL Values (IEC)
4.5
4
HVL (mm Al)
3.5
Acceptable
3
2.5
2
1.5
Not acceptable
1
0.5
0
50 60 70 80 90 100 110 120 130 140 150
IAEA
15.3: Optimization of protection in radiography
kVp
12
Where to Get More Information
Quality Control in Diagnostic Imaging, Gray JE,
Winkler NT, Stears J, Frank ED. Available at no
cost. http://www.diquad.com/QC%20Book.html
IAEA
15.3: Optimization of protection in radiography
13