Transcript QC Tests

QC Tests
Darkroom
Film processor
V.G.Wimalasena
Principal
School of Radiography
QC tests on film darkroom




Safelight evaluation test
Evaluation of white light leakage and
processing area condition
View box quality control test
Evaluation of image duplicators
Safelight test







Load a film into a cassette in complete darkness
Expose the cassette with step wedge on it , using
appropriate exposure factors
Take the film out in the darkroom and keep it on the
darkroom bench.
Cover one half of the film with a cardboard.
Switch on the safe light and wait for 2 minutes
Process the film and observe for a defined line which
divide the image into two halves.
If there is a fog, measure the optical densities and find
the maximum fog (should be < 0.05)
Image of step
wedge
showing
safelight fog
Light leakage testing and processing
area condition





Turn on all the lights surrounding darkroom. Enter
the darkroom and switch off all the lights.
After eyes adapt to the dark, check for white light
leaks
Turn on overhead light and inspect for foreign
objects, dampness etc. on counter tops & film feed
tray
Check the film storage conditions and processing
chemical tanks.
Check the room temperature.
Recommended conditions






Film boxes should be stored vertically
Temperature 14-24 0C
Relative Humidity 30% - 60%
Chemicals must be stored in a well-ventilated area
with a temperature 5-30 0C. (for longer periods < 21
0C) and under low lighting.
Films and chemicals should not be stored together.
(radioactive isotope of potassium-40 is present in the
chemicals and can fog the films)
For films the background radiation should not exceed
7 µR/hr.
Viewbox QC






Should be cleaned weekly
Intensity should be checked once a year (weekly for
mammography)
Use a photometer to measure the brightness. ( luminance or
illuminance.)
The standard luminance is 1700 nit. For mammography 3500
nit. (Illuminance should be above 5000 lux)
The colour temperature should be 5400-10,000 0K (standard
6250 0K)
Check for uniformity of the brightness compared to the centre
(should be within + 10%).
Film Copier QC



Copy a sensitometer film and compare for
density difference( + 0.02). Should be done
weekly
Make a copy of the wire mesh image to check
the sharpness level. Should be done monthly
For digital images copy a test pattern image
and compare the densities from the similar
areas.(+ 0.02)
Processor Quality Control
1. Solution temperature
 Developer temperature should be checked
periodically through out the day. Should not
vary + 0.30C
 Fixer temperature should not vary + 30C from
that of the developer to avoid reticulation
marks.
 Use a digital thermometer with a remote probe
or a glass, alcohol- filled thermometer
2. Processing time



Change of processing time affect the image
contrast, density and the visibility of recorded
detail similar to that by the change of
temperature.
Should be maintained within + 2% to + 3%
Should be checked daily. Use a stopwatch or
digital timer
3. Replenishment rate





If the solutions are not adequately replenished the
image contrast and the optical density decreases.
Excessive replenishment has the opposite effect to
some extent.
The rate should be + 5% of the recommended values.
A film and a graduated cylinder can be used to
measure the rate.
Place the cylinder under the opening of the
replenisher line.
With the cover of the processor open, feed a 18 x 24
cm film lengthwise, collect the chemical in to the
cylinder and measure the volume. Divide the volume
by 24 to get the rate in milliliters per centimeter.
4 .Solution pH



Developer solutions function in a pH range of
10 to 11.5.
Fixer solutions should have a pH between 4
and 4.5
pH should be checked daily by using a digital
pH meter or pH paper.
5. Specific gravity & proper mixing




The chemicals should be mixed to
recommended concentrations. This is given by
the specific gravity.
Use a hydrometer to measure the specific
gravity
Developer – range from 1.07 to 1.1 and should
not vary by more than + 0.004
Fixer – range from 1.077 to 1.11 and should
not vary by more than + 0.004
6. Processor cleaning procedures


Dirty processors cannot function properly and
is the cause of breakdown.
Proper cleaning procedures should be
performed:




Daily – clean all roller racks and look for
malfunctioning parts
Monthly - drain the processing tanks, wash inside
Quarterly – drain replenishment tanks and clean
Yearly – use system cleaners to cleen the
replenisher and circulation systems
7. Processor maintenance


A proper maintenance schedule must be
maintained to ensure continued operation of
the film processor. A log of any maintenance
procedures should be kept for documentation.
Three types of maintenance



Scheduled:- observations, adjustments and
replacement performed daily, weekly, monthly
Preventive:- regular replacement of certain parts
Nonscheduled:- required when a system failure
occurs.
8. Processor monitoring


This is accomplished with the performance of
daily sensitometric tests which evaluate the
performance of processor systems and the
chemicals.
Equipment required




Sensitometer or (step wedge and x-ray machine)
Densitometer
Control chart/graph paper
Film
Processor monitoring equipment
procedure




Expose the film in the sensitometer
Process the film
Measure and record the densities.
Establish the following indicators of the processor
performance.



Base+Fog (BF) – optical density of the clear portion
Speed indicator (or mid-density point) (MD)– optical
density 1 above the base+fog and that step number
Minimum density (or lower density) (Dmin)– density & step
number closest to 0.25 above B+F.
…….Procedure continued





Maximum density (or high density) (Dmax) - the
density and the step closest to 2 above B+F density
Contrast indicator or relative density difference
(DD) = (Dmax - Dmin )
Plot these values on a control chart.
Repeat the procedure daily and measure the
densities of the same step numbers established
on the first day
Look for variations and trends.
Processor QC trouble shooting