Photovac Discount Structure 2002

Download Report

Transcript Photovac Discount Structure 2002

Voyager
Gas Chromatograph
Concept
PHOTOVAC, Inc.
Voyager Basics
• Truly portable
– Only 15 pounds (6.8kg) with battery installed
– 15.4” (39cm) long x 10.6” (27cm) wide x 5.9” (15cm) high
• Multiple Column, Dual Detector Portable GC
– 3 Columns (Heavy, Middle, and Light Compounds)
– PID (Photoionization Detector) and optional ECD (Electron Capture
Detector)
• 3 Levels of Instrument Operation
– Factory programmed methods
– User modified methods
– Total VOC quick screen mode
• Intrinsically Safe (Class 1, Division 1, Groups A,B,C & D)
– The only portable GC with this classification
PHOTOVAC, Inc.
Voyager Basics
• Completely self-contained
– Internal carrier gas cylinder
– Oven
– Battery
• Internal data logger
– On-board data handling provides immediate results
– No laptop needed in field
• Comprehensive Windows software for data
download and reports
• Easy to use
– Minimal training required for point & press operation
PHOTOVAC, Inc.
Voyager Chromatography
• Carrier Gas:
– High Purity Nitrogen or Ultra Zero Air
– Carrier Life: 8 hours at 1800 psi
– Carrier Flow Control via Software (Pressure)
• Column Oven:
– Isothermal 55oC to 80oC
– Higher temperatures help clean out columns and avoid
contamination
• Detector & Injection Port Heated to Column
Temperature
• Two Sampling Methods
– Variable Volume Sample Loop
– Syringe Injection
PHOTOVAC, Inc.
Voyager Chromatography
• Benefits of the 3 column set
– Overlapping selectivities cover
• Heavy compounds (C7 to C12)
• Middle compounds (C3 to C7)
• Light compounds (C1 to C3)
• Simplified method development
• Columns do not degrade
• Vapor phase injections only
• Moderate temperatures with no ramping
PHOTOVAC, Inc.
Column Scheme: Voyager GC Assay # 1
Carrier
Gas (N2)
Column A
Column B
70%
PID
ECD
30%
Pre-Column
Column C
Make-Up
Carrier Gas
Precolumn:
4m x 0.53mm x 2.0 um SPB-35
Column A (HEAVY):
8m x 0.25mm BLANK Fused Silica
Column B (MID-RANGE): 20m x 0.32mm x 1.0 um Supelcowax10 (PEG)
Column C (LIGHT):
15m x 0.32mm x 12 um Quadrex 007-1
PHOTOVAC, Inc.
Column Scheme: Voyager GC Assay # 2 - 7
Carrier Gas
(N2 or UZ Air)
Column A
Column B
PID
Precolumn
Column C
Precolumn:
2.5m x 0.53mm x 2.0 um SPB-35
Column A (Heavy):
8m x 0.32mm BLANK Fused Silica
Column B (Mid-Range): 20m x 0.32mm x 1.5 um Supelcowax 10 (PEG)
Column C (Light):
1.2m x 1.6mm 1% SP1000 on 60/80 Carbopack B
PHOTOVAC, Inc.
Voyager Assays
• Target Compounds grouped into Assays (1-7)
• Each Assay includes
– GC Method
– Integration Settings
– Compound Library
• Assay 1 Environmental Assay
– Most comprehensive with 40 compounds
• Assay 2-7 Occupational Health and Safety
– Targeted by industry segment
PHOTOVAC, Inc.
Assay #1 Environmental
Compound List
Chloromethane
Vinyl Chloride
Bromomethane
Chloroethane
Methylene Chloride
1,1-Dichloroethylene
Acetone
Carbon Disulfide
Vinyl Acetate
t-1,2-Dichloroethylene
Methyl Ethyl Ketone
c-1,2-Dichloroethylene
Chloroform
1,1,1-Trichloroethane
Carbon Tetrachloride
Benzene
1,2-Dichloroethane
Trichloroethylene
2-Chloroethyl Vinyl Ether
1,2-Dichloropropane
Bromodichloromethane
Methyl Isobutyl Ketone
c-1,3-Dichloropropene
Toluene
t-1,3-Dichloropropene
2-Hexanone
Tetrachloroethylene
1,1,2-Trichloroethane
Dibromochloromethane
Chlorobenzene
Ethylbenzene
m-Xylene
p-Xylene
o-Xylene
Styrene
Bromoform
1,1,2,2-Tetrachloroethane
1,3-Dichlorobenzene
1,4-Dichlorobenzene
1,2-Dichlorobenzene
PHOTOVAC, Inc.
Assays #2-7: Occupational Health & Safety
Assay #2: Petroleum and Petrochemical
-
Methanol
Ethanol
MTBE
TAME
Benzene
Toluene
Ethylbenzene
m-, o-Xylenes
Naphthalene
Assay #4: ABS Rubber
- Acrylonitrile
- 1,3-Butadiene
- Styrene
Assay #5: Pulp & Paper
Assay #6: Surfactant Manufacture
-
Ethylene Oxide
Propylene Oxide
-
Hydrogen Sulfide
Methyl Mercaptan
Ethyl Mercaptan
Dimethyl Sulfide (DMS)
Dimethyl Disulfide (DMDS)
Acetone
Methyl Ethyl Ketone (MEK)
a-Pinene
Methanol
Assay #7: Latex Paint Monomers
-
Methyl Acrylate
Methyl Methacrylate
Ethyl Acrylate
n-Butyl Acrylate
iso-Butyl Acrylate
Acrylic Acid
Vinyl Acetate
Ethylene
Styrene
PHOTOVAC, Inc.
Assay Customization
• Preprogrammed assays can be customized in the
SiteChart software for specific applications
• Existing assays are modified by adding or
deleting compounds
• Full control of all hardware and software methods
development and calibration parameters
• Customized assays are then sent to Voyager from
the PC
PHOTOVAC, Inc.