Advancements in Gas Chromatography Analyzers

Download Report

Transcript Advancements in Gas Chromatography Analyzers

Advancements in Gas
Chromatography Analyzers Keeping up with New Technology
Chuck Runkle
Gas Phase Product Specialist
ASTS June 2013
Capillary Flow Technology
-- solves difficult application
problems easily & opens up
many new (and old)
possibilities for GC & GC/MS
– Column connections (connect pre-column)
– Change MSD columns (without venting)
– Backflush (Reverse flow through column)
– Deans switch (heart cut select peaks to 2nd
column)
– Detector splitter (effluent split to two or more
detectors)
– Merge flows (2 columns to 1 MSD)
– Comprehensive 2-D GC (cut all peaks to 2nd
column)
– etc.
Purged Ultimate Union
Agilent 7890-0322 Fast Refinery Gas Analyzer
What is a Fast Refinery Gas Analyzer
• 7890 GC is designed to analyze refinery gas in under 6 min.
• The system has 5 valves and three detectors.
• The FID is for measuring hydrocarbons from C1 to C5 while C6/C6+
components are back-flushed and measured as a single peak at the
beginning of the analysis.
• The first TCD is configured with helium carrier gas to analyze permanent
gases (CO2, CO, O2, N2).
• The second TCD is configured with nitrogen carrier gas to analyze
hydrogen. This detector is mounted on the side of the instrument.
Agilent 7890-0322 Fast Refinery Gas System
Three Channel System for Analysis of Hydrogen, Permanent
Gases and Hydrocarbon Analysis
Col 2
Col 5
Back Aux
Ch 2
PCM C
Ch 1
Back Inl
7 6 5
Out
Sample
4
8
4
9
10
3
1
2
Col 4
6
Vent
LOOP
7 6 5
3
5
2
Sample
1
In
4
8
9
10
Col 1
3
2
1
In
Loop
25ul
100 ul
TCD
Loop
(221)
250 ul
Back
Ch 2
PCM B
Ch 1
Front Inl
EPC
Side
Vent
TCD
(220)
Front
860 CAP
(112)
874
Col 6
FID
(210)
3
6
Col 3
4
4
5
The design looks complicated, but it
is really just three separate channels.
Each channel by itself is very simple.
Lets break it down.
Front
Col 8
2
3
5
6
2
1
1
Col 7
Agilent Fast Refinery Gas Analyzer
Utilizes Four Types of Rotary Valves
4
- Valves are always
depicted in the OFF
position unless otherwise
noted
Out
6
LOOP
7 6 5
3
5
2
1
Sample
4
8
9
10
3
2
1
Vent
In
IN
SAMPLE
OUT
Loop
6-port GSV (Opt 701)
10-port GSV with precolumn back-flush to
vent (Opt 801)
4
4
3
5
6
2
1
6-port column isolation
valve (Opt 702)
3
5
6
2
1
6-port sequence reversal
with pre-column back-flush
to detector (Opt 708)
Agilent 7890-0322 Fast Refinery Gas System
Three Gas Sample Valve Loops in Series
Hydrogen Channel
Hydrocarbon
Channel
7 6 5
Out
Sample
4
8
4
9
10
3
1
2
Permanent Gas
Channel
6
LOOP
7 6 5
3
5
2
1
4
8
Sample
9
10
3
1
2
In
Loop
100 ul
25ul
Loop
250 ul
• There are three gas sample valves, one for each channel.
• Sample flows in through a single 1/16” tube and fills the gas sample loop
for each valve before injection.
• All three channels are injected simultaneously.
• All channels analyze in parallel after injection.
Agilent 7890-0322 Fast Refinery Gas System
Hydrogen Analysis:
Back-flush
MolSieve
Ch 2
PCM C
Ch 1
4
8
Sample
Out
9
10
Loop
100 ul
3
1
2
Vent
Hayesep Q
7 6 5
Back Aux
Sample In
Side
TCD
(221)
• The valve is turned to the OFF position after hydrogen has
eluted to the Molsieve column.
• Hydrogen is separated from helium on the MolSieve and
analyzed on the TCD
• All permanent gases and hydrocarbons are back-flushed from
the Hayesep column to vent.
Agilent 7890-0322 Fast Refinery Gas System
Permanent Gas Analysis:
Sample Load
Unibead
7 6 5
4
8
9
10
Unibead
Sample In
3
2
Vent
Sample
Out
1
Loop
Back Inl
250 ul
Back
Ch 2
PCM B
Ch 1
TCD
(220)
MolSieve
4
3
5
6
2
1
Agilent 7890-0322 Fast Refinery Gas System
Hydrocarbon Analysis
4
3
5
6
2
LOOP
1
25ul
Front Inl
EPC
Front
860 CAP
(112)
Front
2M 5um DB-1
874
FID
(210)
4
3
5
6
2
1
25M 8um
HP-AL/S
Agilent Fast Refinery Gas Analyzer
3 Channels of Data in under 6 minutes
FID: Hydrocarbons
TCD1: Permanent
Gases
TCD2: Hydrogen
Refinery Gas TCD
Universal Capillary Nat Gas Solution
Value to Customers
Features of the Universal Nat Gas Solution
• Can be configured with both LSV and GSV for both liquid and gas injections
• The FID channel is configured with a 6-port backflush valve and a DB-1 60m x
0.25mm x 1.0um
– This channel can be configured for either a C6+ backflush or an extended
analysis simply through a method change.
• TCD channel is equipped with a Deans Switch configured with a PLOT Q capillary
column and a Molsieve column. The columns from the Deans Switch are teed into a
single TCD to give a more compact analysis.
– This channel can be configured to either separate oxygen and nitrogen or to
report a N2/O2 composite, simply through a method change.
• H2S can be analyzed on the PLOT Q TCD channel.
– If the user needs high level H2S the system could be ordered with Hastelloy
valves and Nickel tubing.
Simplified Natural Gas O2/N2
Composite
Natural Gas C6 Plus with O2 & N2
Natural Gas Extended with O2 & N2
Universal Capillary Nat Gas Solution Review
Value to Customers
• We currently have 15 natural gas analyzers
• These systems all revolve around a matrix of 5 variables
– Extended Analysis
– Hydrogen Capacity
– O2/N2 separation
– High H2S
– Liquefied Samples
• We have the ability to configure one instrument based on CFT that can be
configured around 4 of the 5 nat gas variables. This system will not analyze for
hydrogen.
• We do not see a big need for hydrogen analysis in the nat gas industry and if a
customer needs this, they can order a refinery gas system.
Universal Capillary Nat Gas Solution
Order No : Universal NGA Item :
INJECTION
FLOW
SOURCE
OVEN LEFT SIDE
PORT
VALVE COMPARTMENT
BLOCK 1
BLOCK 2
5/852/861
= #87X
4/XXX
S
1/800/502/504
P
W
DETECTOR
Sample In
IN
SAMPLE
S/N :
Date : 2/5/2012
Valve System
C
Sample Out
7 6 5
1.0 mL
Loop
4
8
9
10
3
1
2
0.25 mL
Loop
Col 2
Back
R
Col 1
7890-0020 CAP
(112)
EPC
1
P
S
2
0.53mm FSOT
transfer line
888
Back
Col 3
3
3
Front
860 CAP
(112)
EPC
TCD
(220)
Col 4
1
Front Inl
G3184-60065
Back Inl
Front
4
874
5
6
3
FID
2
(210)
1
Front Aux SAMPLE
S
W
1
AUX 2
3
OUT
P
Col 5
C
0.25mm x 5m FSOT
transfer line
6/850/861
Options
2x112, 210, 220, 301, 305, 306, 502, 504, 708, 752, 800, 850, 852
860, 2x861, 5x872, 874, 888, 19095P-QO4PT, 19095P-MS0E,
160-2255-10, 123-1036, 160-2535-5
3/XXX
Remarks
SP1 options
7890-0990,
7890-0020,
7890-0999
2/708/5x872
Col 1: HP-PLOT Q PT, 19095P-QO4PT, 30m x 0.53mm x 40um
Col 2: HP-PLOT Molsieve, 19095P-MS0E, 30m x 0.53mm x 30 um
Col 3: Restrictor, FST, 160-2255-10, (cut to 1.5m), 5m x 0.250mm
Col 4: DB-1, 123-1036, (2m cut), 60m x 0.25mm x 1.0 um
Col 5: DB-1, 123-1036, (58m cut), 60m x 0.25mm x 1.0 um
Universal Capillary Nat Gas Solution
Example Chromatogram from TCD Channel
Universal Capillary Nat Gas Solution Review
Example Chromatogram from FID Channel 79% Hexanes
LPG Injection, Extended Analysis
MS with Fixed Gases
MS with Fixed Gases Single TCD
In Summery
Capillary Flow Technology
-- solves difficult application
problems easily & opens up
many new (and old)
possibilities for GC & GC/MS