Vet ABC Troubleshooting
CHANGING THE PACK
• Service menu
Change Pack
Remove the old pack
etc. (on-screen instructions)
• Remove the Red caps from the new pack
• Install new pack ensuring arrows have disappeared
• Prime the new pack into an open vessel
• Re-install waste pipe onto new pipe
• Write on the pack the installation date
Operational Problems
•
•
•
•
Basic understanding of how the ABC works
Function of the main components
How to spot potential problems
How to carry out minor repairs
“STARTUP FAILED CHECK REAGENTS”
  Blank values high:
PLT high, pack can be contaminated, reagent too cold
  Perform concentrated cleaning or change pack
  Electromagnetic noise (X-rays)
  Isolate Vet ABC and run start-up
“STARTUP FAILED CHECK REAGENTS”
  If other parameters are OK, photometer blank failed
  Check photometer (see below)
  Pack was shaken while in transport (bubbles)
  After Transport by a carrier, wait 1-2 hours before
installation
  Tube on valve, needle, chamber or dispenser block is
loose
 Check tube connections and repair if possible
“SENSOR DEFECT OR REAGENT EMPTY”
• Pack wasn’t installed properly
• Re-seat the pack ensuring arrows have disappeared
• Valve on base of pack has collapsed inwards
• Adjust the position of the valve
• Pack is empty
• Change the pack
“SENSOR DEFECT OR REAGENT EMPTY”
• Pack is distended with excess air
• “Burp” the pack
• Red caps were not removed
• Remove the Red caps
• Waste tube is blocked or bent
• Check tube is free of obstructions
“ERROR: MOTOR PRESSURE SYRINGE”
• Pack is distended with excess air
• “Burp” the pack
• Piston is pushed out of chamber by high pressure
• Re-install the piston
• Piston is sticking
• Check movement – Service menu
Pressure Syringe.
Mechanic
“ERROR: LIQUID SYRINGE MOTOR”
• Piston or gear is damaged
• Check movement – Service menu
Syringe
Mechanic
• Dilution Block is damaged
• Check for leaks – Call Woodley Equipment.
Liquid
“START-UP FAILED HGB REFERENCE FAILURE”
• WBC Chamber is contaminated
• Run concentrated clean
• Reagent is contaminated
• Change pack
“START-UP FAILED HGB REFERENCE FAILURE”
• No reagent is available
• Prime the pack – check that reagent appears in
chambers
• Photometer is out of adjustment
• Run concentrated clean; Check photometer
adjustment – Service menu
Technician Adjustments
IF IN ANY DOUBT CALL
WOODLEY!