Transcript Document

Getting Started with the EVAPro

After you remove your

EVAPro

from the box, check to make sure you have all accessories shown below.

Click here for next screen

Flow Meter Flag Fail Pass Flow Meter Ball

Quick Reference

Tester Hook Up Phase-One Phase-Two Click here for next screen

Power ON Check Oil

Power

On

Check Oil

Selector Switch Test Smoke 0.040

” 0.020

” Future

Remote Smoke Button Smoke Supply Hose Click here for next screen

Nitrogen / Shop Air Connector Fitting

12-Volt DC Power Cable Accessories Compartment

Connect nitrogen or shop air supply hose to the quick-connector shown below. The

EVAPro

comes with a built-in regulator that allows you to use nitrogen for all EVAP testing or shop air for all of your general purpose leak detection needs. Click here for next screen

Choose the proper EVAP port adaptor for the Vehicle from the accessory kit. Click here for next screen The Schrader removal tool is used for taking the core out of the EVAP service port.

Because of the volatile fumes that may be present in an evaporative emissions system, we recommend using an inert gas, such as nitrogen, for testing EVAP systems.

Do not perform tests near a source of spark or ignition

TEST SAFELY Q:

Why use Nitrogen to test EVAP system?

A:

SAE published Paper reports the hazards of using air instead of Nitrogen to test the EVAP system. http://www.sae.org/technical/papers/2007-01-1235

Q:

What is BAR requiring on the California BAR-Certified equipment?

A:

BAR says only Nitrogen can be used in BAR Certified equipment, citing “equipment requirements and safety concerns”.

Q:

Why do some other smoke machine companies say you don’t need to use Nitrogen with their machines?

A:

Most likely because their machines are not built with the licensed, patented technology to use nitrogen. Sounds like an attempt to put the best ‘

spin

’ on a product that CAN’T use Nitrogen.

Note:

All current Vacutec smoke machines contain patented technology designed to be used with Nitrogen for safer EVAP testing.

1. Remove the OBD-II service port adaptor cap. 2. With the Schrader removal tool turn clockwise and remove the core from the service port

*

.

You must remove the core before testing.

[After testing,don’t forget to reinstall the OBD-II service port Schrader core and cap].

*

If the vehicle being tested does not have an EVAP service port, then access the EVAP system at the purge valve line that leads to the charcoal canister.

Click here for next screen

Let’s do a step by step EVAP test. Remember that you must use nitrogen!

1.

Connect the

red

lead to positive on battery.

2.

Connect the

black

lead to chassis ground.

With power leads connected, the green light should be ON.

A low battery will cause the Tester’s green Power light to blink.

Click here for next screen See Trouble Codes in your Operator Manual.

For the purpose of this presentation, we will use the .040

” standard.

Power ON Check Oil

1.

2.

3.

Position the Tester’s selector switch to the .040

” calibration. Turn the Tester ON. Set the

Pass / Fail

mark by using your thumb to align the red pointer flag with the ball. Now the red flag is set to equal a .040

” leak and it is calibrated for that vehicle.

Click here for next screen Remember to calibrate for .040

” or .020” before each test.

Use scantool and close Vent Solenoid

Vent Solenoid NO - OFF To activate vent solenoid

Push YES to turn

ON

Push NO to turn

OFF TPS 2.5V

FTL 58%

If scantool doesn’t close vent solenoid, refer to Tech Tip Booklet

Click here for next screen

Phase One Power ON Check Oil

Five minutes after turning the Tester ON, it will turn off automatically.

Click here for next screen 1. Set selector switch to TEST.

2. Close EVAP vent solenoid. 3. Turn the Tester ON.

At first the ball will go to the top of the flow meter.

The ball will slowly drop until it no longer goes down.

3. Once the ball stops moving downward, check the alignment with the red flag. If the ball is below the red flag, the leak(s) are less than .040

”;

PASS

If the ball is above the red flag, the leak(s) are greater than .040

”;

FAIL

Go to

Phase Two

.

Phase Two Power ON Check Oil

1.

2.

3.

Set selector switch to SMOKE.

Turn the Tester ON.

Look for Smoke and/or dye exiting the leak(s).

Click here for next screen Five minutes after turning the Tester ON, it will turn off automatically.

Look for the smoke or the ultraviolet-activated fluorescent dye.

888-822-8832 www.vacutec.com

Click here to Exit PowerPoint