Transcript Slide 1

Method Modifications
Mentoring Session
Technical Assistance Committee
TNI Training Disclaimer

The material presented in this session is
for informational purposes only. It is
designed to promote understanding,
consistency and clarification of technical
and accreditation requirements. It should
not be considered a change or alteration
of the Accreditation standards, the
published methods, a regulatory agency
requirement or a position of TNI.
Method Modifications
The primary objective of method
modifications is to enhance precision and
accuracy for each analysis.
All laboratories large and small:
Such as commercial and private
(including municipalities) must prove
that their method modification to
any approved standard analytical
method is valid.
What to Document
The information and supporting data,
required during the validation, must
be sufficient to allow an independent
reviewer (EPA, state or contractor) to
verify each determination and
calculation performed by the
laboratory.
Modifications
Where more than one matrix type is involved
(e.g., air, soil, water, sludge)
1.
•
2.

The laboratory must validate the method on each
matrix type.
In most cases several different samples
should be involved in the validation,
Example: Non-Potable Water several
effluent samples from different plants
should be used.
Method Modifications

Validation Criteria:
1.
Must include the documented procedure used to validate and
report the data
2.
Should explain the statistical analysis of study results.
3.
Must include all raw data results to allow for a complete
independent reviewer verification
4.
Should be checked by comparing the performance criteria
with the actual approved standard method, as written.
Quality Control


•
What is involved?
Initial Precision and Recovery
Calibration (initial and continuing)
•
Method Detection Limit (MDL),
•
Laboratory fortified blank
•
Matrix Spikes
•
Method Blanks
•
Quality Control
•
Internal standard/s
•
Surrogate standard/s
•
Tracer (for radiochemistry)
•
Control charts or other trend analyses
•
Quality Control acceptance criteria

Raw Data
• Sample numbers or other identifiers used
• Sample preparation (extraction/digestion) dates
• Analysis dates and times
• Sequence of analyses or run logs
• Sample volume
Raw Data
• Extract volume prior to each cleanup step
• Extract volume after each cleanup step
• Final extract volume prior to injection
• Digestion volume
• Titration volume
Raw Data
•
Percent solids or percent moisture
•
Dilution data - differentiating between dilution of a sample and
dilution of an extract
•
Instrument(s) and operating conditions, such as:
•
GC and/or GC/MS: including detailed information on columns used for
determination and confirmation (column length and diameter, stationary
phase, solid support, film thickness, etc.) In addition, analysis conditions
(temperature programs, flow rates, etc.) must be documented
•
•
Detectors: type, operating conditions, etc.
Chromatograms: ion current profiles, bar graph spectra, library
search results , etc
Raw Data
• Data Quantitation reports, system outputs, and
other data to link the raw data to the results
reported. (Where these data are edited manually,
explanations of all manual changes must be
included)
• Direct instrument readouts; strip charts, printer
tapes, etc., and other data to support the final
results
Raw Data
• Laboratory bench sheets and copies of all pertinent
logbook pages for all: sample preparation, cleanup, and
determinative steps of the analysis
• Dilution data - differentiating between dilution of a
sample and dilution of an extract
Example: 1
Standard Methods
2540 D.
Total Suspended
Solids
Sample Analysis:
3 (b) states to stir the
sample and pipet an
aliquot from the
midpoint of the
container
When using 1000 mL sample,
the pipet used in the method is
a 25 mL wide bore, this would
take an analyst 40 separate
measurements to deliver the
volme
Modification: Using a
graduated cylinder (1000
ml) instead of the 25mL
pipet, reduces analyst
time and handling of the
sample
Example: 2



The use of an alternate QA/QC chemical in place of the
Glucose/Glutamic Acid (GGA) for BOD/CBOD.
Laboratories that are using KHP (Potassium Hydrogen
Phthalate) for their BOD/CBOD depletion chemical.
Does the laboratory have the right to modify the method
to a degree of changing the chemistry?
References




40 CFR 136.6 Method modifications and analytical
requirements.
Flexibility to Modify CWA Method Nov 20, 2007
EPA: Solutions to Analytical Chemistry Problems
with Clean Water Act Methods 2007 (Pumpkin Book)
(EPA 821-B-98-002 EPA Approval of Alternate Test
Procedures for Organic and Inorganic Analytes in
Wastewater and Drinking Water March 1999)