Cleaning/decontamination, disinfection, and sterilization
Download
Report
Transcript Cleaning/decontamination, disinfection, and sterilization
Cleaning/decontamination,
disinfection, and sterilization
Kumthorn Malathum, MD
Chair, ICC, Ramathibodi Hospital
Scope
Cleaning/decontamination, disinfection,
and sterilization of medical devices
Environmental cleaning
Routine: floor, bed rail, etc.
MDR bacteria
Respiratory pathogens (virus, mycobacteria)
Virulence is not
Certain enveloped (lipophilic)correlated
viruses
with
(e.g., HSV, HIV, influenza virus, and
RSV are susceptible to alcohols.
the
ease
to
be
Hepatitis B virus is an enveloped virus
that is somewhat less susceptible
but
destroyed!
is killed by 60%–70% alcohol; hepatitis
C virus also is likely killed by this
percentage of alcohol.
APIC guidelines Am J Infect
Control 1996:24;313-42
Cleaning
Removing all foreign material from objects
by using water and detergents or soaps
and washing or scrubbing the object
Must be done before any disinfection or
sterilization process
Disinfection
A process that eliminates many or all
microorganisms except spores
Done with liquid chemicals or by
pasteurization
Proper contact time and dilution of the
disinfectant must be followed
Definitions
High-level disinfection
All microorganisms except high numbers of
bacterial spores
Intermediate disinfection
M. tuberculosis, vegetative bacteria, most
viruses, and most fungi
Not necessarily kill bacterial spores
Sterilization
A process that completely eliminates or
kills all microorganisms
Classification of device, process, and products
Device
Process
Product
Classification
Examples
Critical
Implant, surgical
instrument
Sterilization
Sterilant/disinfectant
Semicritical
Flexible endoscope,
laryngoscope,
endotracheal tube
High-level
Sterilant/disinfectant
Thermometer,
hydrotherapy tank
Intermediate Hospital disinfectant
-level
with tuberculocidal
activity
Noncritical
Stethoscope,
Low-level
tabletops, bedpans
Sterilant/disinfectant
without tuberculocidal
activity
Chemical disinfection
Aldehydes: glutaraldehyde & orthophthaldehyde (OPA)
Alcohol
Biguanides: chlorhexidine
Halogen and halogen releasing agents:
chlorine, iodine
Quaternary ammonium compound
Glutaraldehyde
High-level disinfectant
Working solution pH 7.5 to 8.5, 14 to 28 days
Mode of action: cross-linking with proteins,
inhibit synthesis of DNA, RNA
2%: vegetative bacteria < 2 minutes, M.
tuberculosis, fungi, viruses < 10 minutes,
spore of Bacillus & Clostridium spp. 3 h
Use: medical equipment
Toxic: respiratory system
Glutaraldehyde
Advantages
Rapid low-temperature disinfection
OPA has greater anti-mycobacterial activity,
no activation required, less noxious, more
stable
Disadvantages
Irritating
Absorbed into plastics leads to toxicity (e.g.,
colitis)
Alcohol
Optimal conc. 60% to 80%
Not HIGH level (spores and hydrophilic
virus are not destroyed)
Use: oral and rectal thermometers, small
surfaces (multiple dose medication vials),
external surface of equipment
(stethoscope, ventilators, manual
ventilation bags)
Biguanides
Chlorhexidine (bisbiguanide)
Insoluble in water
Active against Gm +ve > Gm –ve bacteria >
yeasts & molds
Not sporicidal
Can be inactivated by nonionic surfactant
presented in soaps, hand creams, and
inorganic water contaminants (phosphate,
chlorine)
Chlorine compound
Concentration dependent:
25 ppm: mycoplasma and vegetative bacteria
(<1 ppm) within seconds
100 ppm: Bacillus subtilis spores within 5
minutes, fungus < 1 h
1000 ppm: M. tuberculosis
Household bleach 5.25% = 52,500 ppm
Iodophors
Tincture of iodine
7.5% Povidone-iodine (PVP-I), 0.7%
available iodine
Bactericidal, fungicidal, tuberculocidal, and
virucidal
Short time residual effect
Quaternary ammonium compound:
benzalkonium chloride
Associated with many outbreaks including
non-tuberculous Mycobacterium & GNR
Not recommended for use as skin and
tissue disinfectant
Use: environmental sanitation of
noncritical surfaces (floors, furniture, walls)
Problems associated with the
use of disinfectants
Ineffective cleaning
Too low concentration
Contaminants unlikely to survive in
recommended use-dilution
The amount of use and costs of
antiseptics/disinfectants per bed per year
Danchaivijitr S et al. J Med Assoc Thai 2005; 88 (Suppl 10): S133-9
Places where antiseptics/disinfectants
were prepared
Danchaivijitr S et al. J Med Assoc Thai 2005; 88 (Suppl 10): S133-9
Persons who prepared
antiseptics/disinfectants used in wards
Danchaivijitr S et al. J Med Assoc Thai 2005; 88 (Suppl 10): S133-9
Microbial contamination (%)
Danchaivijitr S et al. J Med Assoc Thai 2005; 88 (Suppl 10): S133-9
Danchaivijitr S et al. J Med Assoc Thai 2005; 88 (Suppl 10): S133-9
Infection control in flexible
endoscopy
Alvarado C et al. Am J Infect Control 2000;28:138-55.
Agents recommended for high-level
disinfection of flexible endoscopes
Glutaraldehyde preparations
Peracetic acid
Orthophalaldehyde
Alvarado C et al. Am J Infect Control 2000;28:138-55.
Agents not recommended for
disinfection of flexible endoscopes
Hypochlorites
Quaternary ammonium compounds
Not sporicidal, tuberculocidal, or viricidal
against hydrophilic viruses
Phenolics
Intermediate-level disinfectants
Alvarado C et al. Am J Infect Control 2000;28:138-55.
Disinfection of a Probe Used in
Ultrasound-Guided Prostate Biopsy
Rutala WA., et al. Infect Control Hosp Epidemiol 2007; 28:916-919
Main findings
Disinfection (i.e., a reduction in bacterial
load of greater than 7 log10 CFU) could be
achieved if the needle guide was removed
from the probe
Rutala WA., et al. Infect Control Hosp Epidemiol 2007; 28:916-919
Main findings
If the needle guide was left in the probe
channel during immersion in 2%
glutaraldehyde, disinfection was not
achieved (i.e., the reduction was
approximately 1 log10 CFU)
Rutala WA., et al. Infect Control Hosp Epidemiol 2007; 28:916-919
Treatment of endoscope after
disinfection or sterilization
Rinsing
Sterile water
Alcohol rinse followed by complete drying
Only sterile water should be used for
endoscopes that pass through sterile tissues.
Alvarado C et al. Am J Infect Control 2000;28:138-55.
Treatment of endoscope after
disinfection or sterilization
Drying
Drying with alcohol and compressed air
should be done between each patient use
when tap water is used to rinse the
endoscope channels and before storage
whether tap water or sterile water is used.
Storage
Alvarado C et al. Am J Infect Control 2000;28:138-55.
Sterilization
Heat sterilization
Dry heat
Moist heat: pressure steam sterilizer
(autoclave)
Chemical
Ethylene oxide
Glutaraldehyde
Autoclave
Steam must come into direct contact with
the surface
Air must be completely removed
Downward displacement
Pre-vaccuum
Flash sterilization
Steam sterilization of patient care items for
immediate use
Not for convenience or an alternative to
purchasing additional instrument sets or to
save time
Not recommended for implantable devices
Flash sterilization
Lack of timely biologic indicators to
monitor performance, absence of
protective packaging following sterilization,
possibility for contamination of processed
items during transportation to operating
rooms, and use of minimal sterilization
cycle parameters (i.e., time, temperature,
pressure)
Parameters for flash sterilization
Ethylene oxide
Extremely penetrative
Non-corrosive
Toxic, irritant, and explosive when mixed
with air at conc. >3%
Odorless
Quality assurance for sterilization
Mechanical monitoring
Exposure time, temperature, and pressure
Ascertain that the sterilization system function
within parameters
Chemical monitoring
Does not verify sterilization
Indicate procedural errors and equipment
malfunction
ความถี่ในการตรวจสอบ
Mechanical
monitoring
ทุกครั้งที่ทาการอบ
External chemical ทุกห่อยกเว้นถ้าสามารถมองเห็นตัวบ่งชี้
indicators
ทางเคมีภายในได้ชัดเจน
Internal chemical
indicators
Biological
monitoring
ทุกห่อถ้าปฏิบัติได้
ควรใส่ในห่อขนาดใหญ่และห่อ
เครื่องมือผ่าตัด
Daily or at least weekly
AAMI ,AORN ,CDC,CSA
Definition of a chemical indicators
CI – System that reveals a change in one
or more predefined process variables
based on a chemical or physical change
resulting from exposure to a process.
CI classes
Class 1 – Process Indicators
Used to show exposure to a process. No
information about the success or failure of the
process
Class 2 – Specific Test Indicators (e.g.
BDT)
Class 3 – Single variable indicators
Respond to a single variable in the process
e.g. temperature
CI classes
Class 4 – Multivariable Indicators
Respond to two or more variables in the
process
Class 5 – Integrating Indicators (Chemical
Biological Indicators)
Respond in a way which mimics the response
of a BI if used in the same process
CI classes
Class 6 – Emulating Indicators (Cycle
Verification Indicators -Chemical Chart
Recorders)
Respond to all critical variables of the process
at levels associated with acceptable sterilizing
conditions e.g. 134 for 5 mins.
Quality assurance for sterilization
Bowie-Dick test
Performed daily, with pre-vacuum system, in
an empty chamber
Detect residual air in the sterilizer chamber
Biologic monitoring of steam
sterilization
Biologic indicators demonstrated bacterial
growth from spore strips on 15 (12%) out
of 125 cycles
Chemical indicators revealed a change of
color to black after all 125 cycles
Kelkar U et al AJIC 2004, 512-513
BI – changes
Bacillus subtilis renamed to B. atrophaeus
Bacillus stearothermophilus renamed to
Geobacillus stearothermophilus
QA for sterilization
Whenever sterilizers are used, they should
be routinely tested with biological
indicators to ensure they are working
correctly
Items that are sterilized should remain
sterile until the package is torn, wet, or
damaged. Sterility is a function of intact
packaging, not time.
ขั้นตอนการตรวจสอบด้ วยสปอร์ เทสต์
นาสปอร์เทสต์มาทาการอุ่นเชื้อในเครือ
่ งอุ่นเชื้อที่
สามารถควบคุมอุณหภูมิให้มีความเหมาะสมใน
การเจริญเติบโตได้ดี
Steam 56 1o C
EtO 37 1 o C
การอ่านผลการตรวจสอบ
ตรวจดูการเปลี่ยนแปลงสีของน้าเลี้ยงเชื้อทุก 8,
12, 24 และ 48ชม
การเปลี่ยนแปลงสีเป็นสีเหลืองแสดงว่าการทาให้
ปราศจากเชื้อไม่สมบูรณ์เชื้อโรคถูกฆ่าตายไม่
หมด
ไม่มีการเปลี่ยนแปลงสีแสดงว่าการทาให้
ปราศจากเชื้อสมบูรณ์เชื้อโรคถูกฆ่าตายหมด
การอ่ านผล
Successful
Sterilization
Process
Failed
Sterilization
Process
อ่านผล จากการเปลี่ยนแปลงสีของน้าเลี้ยงเชื้อ
ภายหลังการอุ่นเชื้อที่ 8, 12, 24, 48 ชั่วโมง ถ้ามีการ
เปลี่ยนแปลงสีเป็นเหลืองให้ทาการอ่านผลได้เลย
ปัญหาที่พบบ่อย
แช่ก่อนล้าง
แช่ของที่ควรนึ่ง หรืออบแก๊ส
การระบายอากาศในห้องที่ใช้ Glutaraldehyde
ระบบการตรวจสอบการทางานของเครื่องนึ่ง/
เครื่องอบแก๊ส
โครงสร้าง
บุคคล
สถานที่