The Disposable Infusion Pump Instruction

Download Report

Transcript The Disposable Infusion Pump Instruction

The Disposable Infusion Pump
Instruction
Basal + PCA Bolus Model
Characteristics
The Disposable Infusion Pump is a safe, effective and
durable micro injector for safe analgesia. It is driven by the
balanced contraction of the medical silicone rubber
reservoir, with the flow rate controlled by the Flow Rate
Controlled Tubing.
The pump is mainly used in the analgesia for all types of
post-operative local pain including childbirth, in the
therapy of chronic pains, in the analgesia for malicious
tumor as well as some other therapies that involve the selfcontrolled micro injections.
Material

The Disposable Infusion Pump contains a silicone rubber
reservoir, together with PVC tubing, ABS tubing connector.
(Warning: certain solution may be incompatible with the
materials used in the Disposable Infusion Pump. Before
use consult the instruction of the fluid or drug to be used.).

The Disposable Infusion Pump is assembled with a
standard male luer,which can connect with standard
infusion tubing.
Specification (Basal + PCA Bolus Model)
The Disposable Infusion Pump List (II)
Capacity of
Reservoir (ml)
WZ-6523C-2
100ml
WZ-6523C-3
100ml
WZ-6523C-4
100ml
WZ-6523C-4.25
100ml
WZ-6523C-5
100ml
WZ-6523C-6
100ml
WZ-6523C-7
100ml
WZ-6523C-8
100ml
WZ-6523C-9
100ml
WZ-6523C-10
100ml
WZ-6523C-13
100ml
WZ-6523C-02
275ml
WZ-6523C-04
275ml
WZ-6523C-05
275ml
WZ-6523C-09
275ml
WZ-6523C-013
275ml
WZ-6523C-016
275ml
Model
Basal Rate
(ml/hr)
1.0ml/hr
1.0ml/hr
2.0ml/hr
0.5ml/hr
1.0ml/hr
2.0ml/hr
3.0ml/hr
4.0ml/hr
5.0ml/hr
2.0ml/hr
5.0ml/hr
1.0ml/hr
2.0ml/hr
3.0ml/hr
5.0ml/hr
5.0ml/hr
8.0ml/hr
PCA Bolus
(ml/min)
0.5ml/30min
0.5ml/15min
0.5ml/15min
0.5ml/8min
1.0ml/15min
1.0ml/15min
1.0ml/15min
1.0ml/15min
1.0ml/15min
2.0ml/15min
2.0ml/15min
0.5ml/30min
0.5ml/15min
0.5ml/15min
1.0ml/15min
2.0ml/15min
2.0ml/15min
Usage: Before filling the reservoir
Pick out a pump from its packaging,
confirm that a sliding clip is fitted on the
button of PCA administration set.


Close the tubing with a clip.

Fill a syringe with a volume of
solution to be placed in the reservoir.
Usage: Filling the reservoir

Fill in the reservoir through the
filter and check valve. After filling,
remove the filter and close the
check valve with a screw cap.

Although about 0.5ml air bubble
will be remained in the reservoir
after filling, the air bubble can be
removed automaticlly through the
silicone reservior in 1-2 hours,
owing to the character of silicone
rubber material.
Usage: Quickly get rid of the air bubbles from the
tubing

Remove the clip from tubing

Open the adding port

Fit a syringe to the filter and inject
1ml solution into the adding port to
quickly eliminate the air bubbles
from the tubing

Do not forget to close the adding
port with the screw cap after the
above operation.
Usage: After filling
Remove the sliding clip from
the PCA administration set;
Attention:
This step is very important,
please do not forget it.
Otherwise, the PCA Bolus
will be out of control.
During the use, when a patient feel pain or discomfort,
he can press the button of PCA administration set, and
the pump will give the patient an additional dose of
medication according to the PCA bolus rate.
The male lure at the end of the tubing
may connect with epidural catheter
connector or other standard female lure.
 Please ensure that no air bubble existing
in the tubing before connecting catheter
with the male lure.

Attention:
Keep the reservoir and the PCA administration set on the same level when
the infusion pump starts working. Do
not hang the reservoir on a higher place.
Otherwise, the flow rate will be
increased.
Other function:


Withdrawing function
When necessary, doctors can withdraw
some solution from the reservoir with
an extractor.
Fit the extractor to a syringe and keep
that the joint is tight. Once the extractor
is fitted to check valve, the solution will
flow into the syringe immediately.
Other function:

Adding dosage function
Besides Basal infusion and the PCA bolus
infusion, if the doctors want to give the
patients extra or higher doses, they can
inject the medication from the adding port.
Attention:
1. Please keep on pressing the button of PCA
administration set when injecting from the
adding port, otherwise, the PCA
administration set will be damaged.
2. Do not forget to use a filter when injecting
3. Do not forget to close the adding port with
screw cap after injection.
Attention






The infusion pump is sterilized by E.O. gas and it is
valid for three years. Do not use after the expiration date.
Do not use if the packaging has been previously opened
or damaged. Single use only.
Read the instruction carefully before use it.
When operating it an aseptic technique should be
maintained.
Dilute the solution as required before filling it into the
Reservoir.
Do not overfill the reservoir.

The accuracy of flow rate control is within ± 10%.

The basal rate and the PCA bolus indicated on the packaging are
determined basing on the concentration of 0.9% sodium
chloride under standard room temperature. Owing to the
different concentration of solution and temperature condition as
well as other factors, the actual flow rate will be little different.
Please try to use the pump under the temperature around 2125°C.

The doctors should prescribe the dosage and use proper models
of the infusion pump for different patients.

When patients feel pain or discomfort, please press the button of
PCA administration set to manage additional doses. However,
please don't press the button if they are feeling too sleepy.
The end