Chemical analysis by EDX
Download
Report
Transcript Chemical analysis by EDX
Nathalie Siredey-Schwaller | 09/01/2014
Chemical analysis with EDX
On Supra40
How to get good quality
quantitative chemical analyses
- Understanding the involved
physical phenomena
- Choice of conditions of analyses
- Artefacts and misinterpretations
- Quantitative mapping
- Case of light elements
Basic cares : sample preparation
•
Sample should be an electric conductor
– Good electric conductivity of mounting
– Good electric junction between sample and microscope : use of silver glue. Pay
attention to the release of solvent.
– Beware of oxyde layer at the surface of sample / cleanliness of the surface.
•
Surface of the sample : analysis depth is only a few microns
– Surface should be out of contamination / clean
– A precise quantification needs a flat and horizontal surface (phirhoZ method).
• Measuring beam current : with a Cu tape put near the sample.
This allows to measure the system factor. Sample spectra could then been compared to
standards spectra.
Consequences of bad-electrical conductivity
Deviation of beam
Before recording
spectrum
After 5 min analysis
Chemical signal changes
Duane-Hunt limit should be
identical to primary electron
energy
Conditions for acquiring EDX signal
•
Position of sample
26
25.5
signal
25
intensity
(kcps) 24.5
24
EDX detector
23.5
7
8
9
10
working distance
11
Sample
WD = 8.5 to 9 mm, magnification > 300
•
No object between sample and detector
Détecteur
Echantillons
No analyses
inside the
hole
Sas d’entrée
•
Infra-red camera is « off » : it induces noise signal
How to choose voltage and beam aperture ?
• Voltage has to be chosen according to chemical elements expected
Enough energy to eject an orbital electron of the element electron reorganization and emission of characteristic X-rays : K or L, (M) spectra
The efficiency is the best for Eprimary beam = 2*Eionization
– Analysis of light elements with small voltage (better signal/background ratio)
– The highest-Z element to measure determines the voltage.
– The voltage is more specially chosen according small content amount
elements
– Massic absorption coefficients for K spectra are the better known.
– Usual values are15 keV, 20 keV
• Beam current has to be chosen according to detector
settings
– Detector with high spectral resolution only accepts small intensity signal
– Mapping needs good X-ray statistic per points, so high intensity signal
Measuring beam current / intensity of the signal
•
•
•
You have to do it in order to be able to compare your sample to standards.
Measuring System Factor SF
Conditions of measurments : on copper :
– Focus to the good working distance
– Tuning of the primary electronic beam : deviation (wooble), stigmatism.
– Check the tilt angle (must be 0°and not 70°)
– Choose the right device settings for the detector
– Calibrate. Check if spectral position of Cu-Ka peak is OK.
Peak maxima match
with color element
lines
– Once this is done, nor the focus, nor the electron beam should be changed.
Choice of device settings
•
•
•
Detector has dead time, during which X-ray signal cannot be count. For good
statistics, dead time shall not to exceed 10%
60 kcps setting = better spectral resolution, higher dead time ( less
acceptable intensity)
275 kcps setting has less spectral resolution but accepts more signal.
use of 60 kcps (40 keV) setting when peaks are close together, or overlap
use 275 kcps (20 keV) when high statistic are needed (mapping)
60 kcps
275 kcps
Conditions of spectra acquisition
•
•
Analysed area should be homogeneous
What is a point spectrum ?
– Existence of an emitting volume
rho (g/cm3)
E0 (keV)
z max (µm)
2
(Al2O3)
15
3.3
5
(Fe3O4)
15
1.3
8
(Fe)
15
0.8
20 (Pb, Au)
15
0.3
Theses phases cannot be analyzed
separately
Size of analyzed particle is about 1
micron. At 25 kV, part of the Al matrix
actually contributed to the signal.
Acquisition of a spectrum
•
Acquisition time :
–
–
–
•
higher is, better is signal / background ratio. X-ray emission
is a statistic phenomenon : relative error is proportionnal to
1/√N.. Better is 106 counts.
Small as possible if existence of C-contamination of the
sample or if sample is a bad electrical conductor.
When mapping, time spent on one point should be as less
as possible use of high beam current
Identification of elements :
–
–
–
–
An element is identified by all characteristic X-rays. Color
lines indicate the relative intensities.
All peaks should be explained
Beware of peaks overlapping
Beware of artefacts due to detector saturation : 2 photons
E1, E2 are count together as one photon E=E1+E2. Exists
in case of high signals.
Overlapping of peaks
Surface C, O
contamination
peaks
Double-photon peak
Quantitative analysis of a spectrum
Two possible methods :
• P/B ZAF method, without standards
• Phirho(Z) method, with standards. Better in case of
special conditions
– Open library with the same device settings
– Load the method of analysis
– Quantify
• Check quantitative analysis
– Built spectrum should be identical to real spectrum
– Total unnormalized massic sum should be equal to 100 %
• Quite good quality analysis if result is between 96% and 104%
• If result is less then 100% : maybe some elements have been forgotten
P/B ZAF method
• ZAF :
– Z = « atomic number » effect
Ionization cross section, ability to stop the electrons according to the material.
– A = absorption
Depends on X-ray energy, density, depth of emission
– F = fluorescence effect between elements
One photon emitted by element A is absorbed by element B, inducing fluorescence. Photon from
A is lost, one photon from B is observed.
• Intensity of X-ray characteristics peaks is compared
to intensity of background
no standards are needed
F(rz) method
• The function f(rZ) is calculated. This function describes
emission of photons X, for a given material. It depends on :
–Density r of the material
–Depth Z, from where the emission occurs
Z and A are simultaneously determined. It is then obtained curves,
describing the emitted and emergent signals, according to the
depth.
• Fluorescence between elements is a phenomenon distinct
from emission. Its calculation is similar to the previous case (P/B
ZAF).
•
Analyze with standards :
For each elemental X-ray characteristic peak, sample signal is
compared to standard signal (known composition or pure standard).
“k” is a corrective term from Z.A.F. corrections. “k” depends of amounts
of all the elements existing in the sample and is only obtained after an
iterative process.
Csample
Cs tandard
kratio
k
I sample
I s tandard
I sample
I s tandard ( pure )
Perfect match between real
spectrum and built one
Except for these energies
(C contamination)
Total sum is almost 100%
Loading of the standards
- Choice of the method : triangle near « quantify ».
- May be loaded from recorded files, may be created, may be saved.
The objects
•
One special point may be choose (beware of emitting volume !)
•
Lines may be analyzed
– Acquisition :
• Choose the number of points in the line. Distance between points should be higher than
size of emitting volume.
• Thanks to the global spectrum, determination of all the chemical elements.
• Determination of the measuring time. According to the signal intensity. At least egal to
number of points x 20 000 counts.
• Possibility of increasing width of the line
– Quantification : choose the appropriate method.
– Saving : in the project or in a file .txt, readable by Excel.
Indicates the intensity
of signal = number of
counts per second.
Use this indication to
determine the minimum
analysis time on each
point
The « hypermap » object
•
The aim is to record a mapping of a sample area. This mapping may be
subsequently analyzed.
•
Acquiring a mapping :
–
–
–
–
–
•
•
Use of a high beam current in order to reduce analysis time. Consequently, use
appropriate detector settings (130 kcps, 275 kcps)
Determination of the total points number
Determination of the total analysis time= nber of points * 20 000 counts
Select the chemical elements to display.
To begin, click on « Acquire » (parameters = right-hand triangle). To stop, click on the
same button. The mapping is realized with a lot of quick scanning. This allows to reduce
C contamination, charge effect. The mapping becomes more and more precise with time:
signal / background ratio increases.
Save the mapping : in a separate file : .rtb or .bcf, usually quite big . In this file are
recorded the full spectra of each point.
Quantitative analysis of mapping : Qmap
–
–
–
–
May be time-consuming, so to be done as a post-process.
Select the quantification method (triangle near « Qmap »).
To begin : click on « Qmap ».
Results may be displayed on maps, false color maps, or recorded on “.txt” files (1 for
each elements) readable by Excel
Qualitative mapping before quantitative analysis
•
•
•
•
This mapping was realized on Ka-Si
peak energy
This is not a Si-content mapping: at
this energy, some others peaks
overlap
Also beware, in qualitative mapping,
of variation in background signal
A
quantitative
mapping
is
mandatory
N1 = Number of counts / second
N2 = Number of points / line
Minimum = 100 points
S = Size of the image = HxV
N2 = should be chosen to be H / d
d = distance between two points (1,2
or 3 microns).
N3 = (H/d) x (V/d) is the total number
of points in the mapping
Time of the mapping, for 20 000
counts per points is (in seconds) :
N3 x 20 000 / N1
Chemical analysis including one light element
•
Preliminary considerations :
Detector should have a good spectral resolution
Characteristics X-rays of light elements have small energy. In this range, there is C contamination,
usually of high number of X-ray characteristics peaks, bad transmission of the detector window. In
order to increase signal/background ratio, primary electron energy should be small, which is
usually incompatible with the other chemical elements
•
The best method is to measure out by difference, once all other chemical elements have been
analyzed.
This needs a careful analyzing of the other elements
– PhirhoZ method with standards
– Very good standards (at least 1 million counts standards)
– Very good statistics of the spectrum (at least 1 million counts standards)
– Check on an area not containing light element that total unnormalized sum is 100±1 %.
Select « measure out by difference » inside the method